PROLENE 8726HRecalls & reported problems
Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant
Quick summary
This Ethicon Prolene 8726H is a blue polypropylene surgical suture used to close tissue. The 48 reports mostly describe the thread breaking during surgery or coming apart from the needle, with 47 malfunctions, 1 death and no injuries. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 47
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 47 adverse event reports that list this exact device (DI 10705031019805), received between Oct 15, 2019 and Jul 2, 2026 — classified as 1 death, 46 malfunctions.
The most frequently reported specific device problem is “Break” (28 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break28
- Material Separation18
- Material Integrity Problem2
- Component Misassembled1
Patient problems
- No Clinical Signs, Symptoms or Conditions44
- No Consequences Or Impact To Patient4
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 25 most recent of 47 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 2, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ANASTOMOSIS PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. SEVERAL 6-0 SUTURES KEPT BREAKING DURING ANASTOMOSIS WITH NO CAUSE. SCRUB NURSE/SURGEON SUSPECT THAT SUTURES WERE FAULTY AS IT WOULD BREAK VERY EASILY. SUTURES WERE NOT EXPIRED HOWEVER NOTICED ALL THE FAULTY SUTURES (3 SUTURE PACKS) WERE FROM SAME. REMOVED THESE SUTURES FROM OPERATING ROOM PYXIS. NEW PACKS OPENED AND USED WITH DIFFERENT LOT NUMBER. NO ISSUES NOTED WITH THESE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25697723 ↗MalfunctionJul 2, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ANASTOMOSIS PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. SEVERAL 6-0 SUTURES KEPT BREAKING DURING ANASTOMOSIS WITH NO CAUSE. SCRUB NURSE/SURGEON SUSPECT THAT SUTURES WERE FAULTY AS IT WOULD BREAK VERY EASILY. SUTURES WERE NOT EXPIRED HOWEVER NOTICED ALL THE FAULTY SUTURES (3 SUTURE PACKS) WERE FROM SAME. REMOVED THESE SUTURES FROM OPERATING ROOM PYXIS. NEW PACKS OPENED AND USED WITH DIFFERENT LOT NUMBER. NO ISSUES NOTED WITH THESE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25697645 ↗MalfunctionJun 23, 2026Material Separation; Material Integrity Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN VASCULAR PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING A VASCULAR PROCEDURE, MULTIPLE SUTURES FROM THE SAME PACKAGE DETACHED FROM THE NEEDLE. THEY DON'T KNOW HOW MANY SUTURES WERE OPEN OR USED. THE USER INDICATED THAT WHILE PASSING THE SUTURE, IT DID NOT FEEL NORMAL AND REQUIRED MORE FORCE THAN EXPECTED. THE SUTURE WAS DESCRIBED AS FEELING DIFFERENT COMPARED TO TYPICAL USE, AND UPON PULLING THROUGH, IT SNAPPED. NO PATIENT CONSEQUENCES WERE REPORTED. THE DEVICE IS EXPECTED TO BE RETURNED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25595607 ↗MalfunctionJun 23, 2026Material Separation; Material Integrity Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN VASCULAR PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING A VASCULAR PROCEDURE, MULTIPLE SUTURES FROM THE SAME PACKAGE DETACHED FROM THE NEEDLE. THEY DON'T KNOW HOW MANY SUTURES WERE OPEN OR USED. THE USER INDICATED THAT WHILE PASSING THE SUTURE, IT DID NOT FEEL NORMAL AND REQUIRED MORE FORCE THAN EXPECTED. THE SUTURE WAS DESCRIBED AS FEELING DIFFERENT COMPARED TO TYPICAL USE, AND UPON PULLING THROUGH, IT SNAPPED. NO PATIENT CONSEQUENCES WERE REPORTED. THE DEVICE IS EXPECTED TO BE RETURNED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25595605 ↗MalfunctionOct 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP BY THE ACCOUNT CONTACT THAT, DURING UNSPECIFIED PROCEDURE, THE SUTURE WAS BREAKING. THEY PULLED ANOTHER ONE AND IT WORKED FINE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE IS AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23293860 ↗MalfunctionAug 5, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ENDARECTOMY (RT) PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT THE SURGEON STATED THAT THE SUTURE BREAKS RANDOMLY. THE SURGEON WAS WORRIED THAT HE MIGHT LOSE THE NEEDLE. STERILE SAMPLES ARE AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22718884 ↗MalfunctionAug 5, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ENDARECTOMY (RT) PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT THE SURGEON STATED THAT THE SUTURE BREAKS RANDOMLY. THE SURGEON WAS WORRIED THAT HE MIGHT LOSE THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. STERILE SAMPLES ARE AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22718866 ↗MalfunctionJan 14, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP BY THE SURGEON DURING THE CASE THESE SUTURES WERE SNAPPING IN HALF AND SEEMED TO BE BRITTLE. DEVICES WERE RETURNING. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21152668 ↗MalfunctionJan 14, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP BY THE SURGEON DURING THE CASE THESE SUTURES WERE SNAPPING IN HALF AND SEEMED TO BE BRITTLE. DEVICES WERE RETURNING. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21152667 ↗MalfunctionDec 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG AND AORTIC VALVE REPLACEMENT ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE SNAPPED TWICE IN THE MIDDLE WHEN THEY WERE TRYING TO TIE THE HEART STRING INTO THE HEART. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20897215 ↗MalfunctionDec 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG AND AORTIC VALVE REPLACEMENT ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE SNAPPED IN THE MIDDLE WHEN THEY WERE TRYING TO TIE THE HEART STRING INTO THE HEART. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20896668 ↗MalfunctionNov 18, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFTING PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING CONTINUOUS SUTURING, THE SUTURE BREAKAGE OCCURRED AT A DIFFERENT POINT FROM THE LIGATURE POINT DURING LIGATION. IT WAS NOT A CONTROL RELEASE NEEDLE. IT WAS USED ON BLOOD VESSEL. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20709037 ↗MalfunctionMar 5, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WOULD BREAK WHEN THE DOCTOR WAS TYING IT DURING A VESICULAR/CARDIOVASCULAR PROCEDURE. THEY OPENED A DIFFERENT LOT NUMBER TO COMPLETE THE CASE WITHOUT PATIENT HARM. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18838492 ↗MalfunctionOct 11, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN (B)(6) 2023 AND SUTURE WAS USED. WHILE DOING SURGERY THE SUTURE NEEDLES HAD POPPED. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17916015 ↗MalfunctionOct 11, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN (B)(6) 2023 AND SUTURE WAS USED. WHILE DOING SURGERY THE SUTURE NEEDLES HAD POPPED. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17916012 ↗MalfunctionSep 28, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A GASTRIC BYPASS PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE SUTURE BROKE EASILY. IT HAPPENED AT LEAST 2 TIMES WHERE THE SUTURE WOULD POP OFF WHILE SUTURING, INTRA-OPERATIVELY. SWITCHED BOXES AND THE PROBLEM STOPPED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17834805 ↗MalfunctionSep 28, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A GASTRIC BYPASS PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE SUTURE BROKE EASILY. IT HAPPENED AT LEAST 2 TIMES WHERE THE SUTURE WOULD POP OFF WHILE SUTURING, INTRA-OPERATIVELY. SWITCHED BOXES AND THE PROBLEM STOPPED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17834462 ↗MalfunctionJan 12, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE THREAD WAS DETACHED FROM THE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16151040 ↗MalfunctionNov 10, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. TWO SUTURES OUT OF THREE BROKE DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15776717 ↗MalfunctionNov 10, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. TWO SUTURES OUT OF THREE BROKE DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15776097 ↗MalfunctionSep 16, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE THREAD POPPED OFF THE NEEDLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15436991 ↗MalfunctionSep 16, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE THREAD POPPED OFF THE NEEDLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15436989 ↗MalfunctionSep 14, 2022Component Misassembled; Component Misassembled
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. AT FIRST APPEARANCE, TWO NEEDLES WERE VISIBLE AS USUAL, BUT WHEN THE NEEDLE HOLDER GRASPED THE NEEDLE AND PULLED OUT THE SUTURE, ONE NEEDLE WITH A SUTURE CAME OUT FROM THE INSIDE. THERE WAS ONE DOUBLE-ARMED SUTURE AND ONE SINGLE-ARMED SUTURE IN THE PACKAGE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15422809 ↗MalfunctionSep 2, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE NEEDLE WAS EASILY DETACHED FROM THE SUTURE. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15352330 ↗MalfunctionSep 2, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE NEEDLE WAS EASILY DETACHED FROM THE SUTURE. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15352311 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2460-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
- Z-2461-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
- Z-2462-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
- Z-2449-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
- Z-2463-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
- Z-2455-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2454-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
- Z-2450-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2448-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
- Z-2452-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.