PROLENE 8747HRecalls & reported problems
Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 20
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 10705031020122), received between Jun 14, 2022 and Apr 16, 2026 — classified as 20 malfunctions.
The most frequently reported specific device problem is “Break” (18 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break18
- Material Separation2
Patient problems
- No Clinical Signs, Symptoms or Conditions20
Recent adverse event reports
The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 16, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE SUTURE UNDERGONE SUTURE BREAKAGE DURING THE CARDIAC SURGERY WHILE DOING DISTAL ANASTOMOSIS. THE SUTURE GOT DETACHED FROM SWAGE POINT. DURING THE SURGERY 2 FOILS AFFECTED AND HOSPITAL HAS REQUESTED TO REPLACE THE REMAINING 13 FOILS ON PRIORITY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. THERE WERE NO SURGICAL DELAYS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24895805 ↗MalfunctionApr 16, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE SUTURE UNDERGONE SUTURE BREAKAGE DURING THE CARDIAC SURGERY WHILE DOING DISTAL ANASTOMOSIS. THE SUTURE GOT DETACHED FROM SWAGE POINT. DURING THE SURGERY 2 FOILS AFFECTED AND HOSPITAL HAS REQUESTED TO REPLACE THE REMAINING 13 FOILS ON PRIORITY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. THERE WERE NO SURGICAL DELAYS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24895803 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22026627 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22026552 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22026543 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22026467 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22026340 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22026279 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22026211 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22026191 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22025996 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22025995 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22025829 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22025818 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22025816 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22025676 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT OVER THE PAST COUPLE WEEKS DURING FIVE HEART PROCEDURES TWO TO THREE SUTURES WOULD BREAK WHILE SUTURING THE ANASTOMOSIS. ANOTHER LIKE DEVICE WAS USED TO CONTINUE WITH THE PROCEDURE TILL COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. EIGHT STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22025675 ↗MalfunctionSep 20, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A CABG PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE NEEDLE IS VERY EASILY POPPING OFF OF THE SUTURE. NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17782903 ↗MalfunctionSep 20, 2023Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A CABG PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE NEEDLE IS VERY EASILY POPPING OFF OF THE SUTURE. NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17782878 ↗MalfunctionJun 14, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14684340 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2459-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
- Z-2448-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
- Z-2456-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
- Z-2460-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2450-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
- Z-2451-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
- Z-2461-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
- Z-2449-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
- Z-2463-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page