Skip to content

PROLENE 8807HRecalls & reported problems

Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Prolene 8807H is a blue polypropylene nonabsorbable surgical suture. The 35 reports mostly describe the thread separating from the needle or breaking during vascular surgery, with a few mentioning bleeding. No deaths and no injuries were reported, and no recalls name this device. Reports fell from 11 in the earlier 12-month period to 1 in the most recent.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 10705031020719), received between Jun 19, 2019 and Jul 17, 2025 — classified as 34 malfunctions.

The most frequently reported specific device problem is “Material Separation” (30 reports).

21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 7 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 5 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation30
  • Break11

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • No Consequences Or Impact To Patient8
  • Hemorrhage/Blood Loss/Bleeding2

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 17, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING AN ANASTOMOSIS, THE NEEDLE PULLED OFF UPON THE PASSAGE OF THE ARTERIAL WALL. A COMPETITION DEVICE WAS USED FOR THE ANASTOMOSIS. THERE WERE NO ADVERSE PATIENT CONSEQUENCES FOR THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22531805 ↗
  • MalfunctionJul 17, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING AN ANASTOMOSIS, THE NEEDLE PULLED OFF UPON THE PASSAGE OF THE ARTERIAL WALL. A COMPETITION DEVICE WAS USED FOR THE ANASTOMOSIS. THERE WERE NO ADVERSE PATIENT CONSEQUENCES FOR THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22531804 ↗
  • MalfunctionJul 17, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING AN ANASTOMOSIS, THE NEEDLE PULLED OFF UPON THE PASSAGE OF THE ARTERIAL WALL. A COMPETITION DEVICE WAS USED FOR THE ANASTOMOSIS. THERE WERE NO ADVERSE PATIENT CONSEQUENCES FOR THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22531803 ↗
  • MalfunctionApr 25, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, QUALITY ISSUE ON SUTURES, DETECTED DURING VASCULAR SURGERY: THE NEEDLE BECOMES LOOSE FROM THE THREAD ON SEVERAL UNITS. CONSEQUENCES: CONTINUOUS STITCH ON A PROSTHESIS - BLEEDING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21910147 ↗
  • MalfunctionApr 25, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, QUALITY ISSUE ON SUTURES, DETECTED DURING VASCULAR SURGERY: THE NEEDLE BECOMES LOOSE FROM THE THREAD ON SEVERAL UNITS. CONSEQUENCES: CONTINOUS STITCH ON A PROSTHESIS - BLEEDING . NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21910146 ↗
  • MalfunctionMar 18, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MANAGEMENT OF BILATERAL FEMORAL THROMBECTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLES OF THE PROLENE 6-0 SUTURE THREADS PULLED OFF. NO CONSEQUENCES FOR THE PATIENT. RISK OF BLEEDING AND RISK OF NEEDLE LOSS IN THE PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21633730 ↗
  • MalfunctionJan 31, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6)2024 AND SUTURE WAS USED. CRIMPING ISSUE THAT GAVE WAY DURING THE PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21285693 ↗
  • MalfunctionJan 31, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID SURGERY IN (B)(6) 2025 AND SUTURE WAS USED. THE NODE GAVE WAY ON CAROTID SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21285357 ↗
  • MalfunctionJan 6, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE HAS PULLED OFF THE SUTURE. ANOTHER DEVICE WAS USED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21085412 ↗
  • MalfunctionJan 6, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN (B)(6)2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE HAS PULLED OFF THE SUTURE. ANOTHER DEVICE WAS USED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21085411 ↗
  • MalfunctionJan 6, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE HAS PULLED OFF THE SUTURE. ANOTHER DEVICE WAS USED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21085408 ↗
  • MalfunctionApr 17, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE PULLED OFF DURING USE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19129316 ↗
  • MalfunctionMar 19, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A CORONARY ARTERY BYPASS SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. THE SURGEON USED A PROLENE WIRE, BUT ON SEVERAL OCCASIONS, THE THREAD BROKE AT THE SLIGHTEST PRESSURE (EVEN MINIMAL) AND THE NEEDLE PULLED OFF. AS A RESULT, THE STAFF HAD TO CHANGE THE BOX OF THREADS, WITH THE SAME REFERENCE NUMBER BUT UNKNOWN LOT NUMBER. THIS ISSUE CAUSED A LOSS OF SERVICE TIME. THERE WERE NO CLINICAL CONSEQUENCES FOR THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18934253 ↗
  • MalfunctionMar 19, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A CORONARY ARTERY BYPASS SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. THE SURGEON USED A PROLENE WIRE, BUT ON SEVERAL OCCASIONS, THE THREAD BROKE AT THE SLIGHTEST PRESSURE (EVEN MINIMAL) AND THE NEEDLE PULLED OFF. AS A RESULT, THE STAFF HAD TO CHANGE THE BOX OF THREADS, WITH THE SAME REFERENCE NUMBER BUT UNKNOWN LOT NUMBER. THIS ISSUE CAUSED A LOSS OF SERVICE TIME. THERE WERE NO CLINICAL CONSEQUENCES FOR THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18934252 ↗
  • MalfunctionMar 13, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A THROMBOENDOARTERECTOMY (TEA) OF THE RIGHT CAROTID ARTERY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SURGEON WANTING TO ANASTOMOSE THE CAROTID ARTERY, USED THE PROLENE WIRE. BUT ON SEVERAL OCCASIONS, THE THREADS BROKE AT THE SLIGHTEST PRESSURE (EVEN MINIMAL) AND THE NEEDLE PULLED OFF. AS A RESULT, THE STAFF HAD TO CHANGE THE BOX OF WIRES, WITH THE SAME REFERENCE NUMBER BUT UNKNOWN LOT NUMBER. THIS ISSUE CAUSED A LOSS OF TIME. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18894249 ↗
  • MalfunctionMar 13, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A THROMBOENDOARTERECTOMY (TEA) OF THE RIGHT CAROTID ARTERY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SURGEON WANTING TO ANASTOMOSE THE CAROTID ARTERY, USED THE PROLENE WIRE. BUT ON SEVERAL OCCASIONS, THE THREADS BROKE AT THE SLIGHTEST PRESSURE (EVEN MINIMAL) AND THE NEEDLE PULLED OFF. AS A RESULT, THE STAFF HAD TO CHANGE THE BOX OF WIRES, WITH THE SAME REFERENCE NUMBER BUT UNKNOWN LOT NUMBER. THIS ISSUE CAUSED A LOSS OF TIME. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18894248 ↗
  • MalfunctionMar 13, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A THROMBOENDOARTERECTOMY (TEA) OF THE RIGHT CAROTID ARTERY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SURGEON WANTING TO ANASTOMOSE THE CAROTID ARTERY, USED THE PROLENE WIRE. BUT ON SEVERAL OCCASIONS, THE THREADS BROKE AT THE SLIGHTEST PRESSURE (EVEN MINIMAL) AND THE NEEDLE PULLED OFF. AS A RESULT, THE STAFF HAD TO CHANGE THE BOX OF WIRES, WITH THE SAME REFERENCE NUMBER BUT UNKNOWN LOT NUMBER. THIS ISSUE CAUSED A LOSS OF TIME. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18894240 ↗
  • MalfunctionMar 13, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A THROMBOENDOARTERECTOMY (TEA) OF THE RIGHT CAROTID ARTERY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SURGEON WANTING TO ANASTOMOSE THE CAROTID ARTERY, USED THE PROLENE WIRE. BUT ON SEVERAL OCCASIONS, THE THREADS BROKE AT THE SLIGHTEST PRESSURE (EVEN MINIMAL) AND THE NEEDLE PULLED OFF. AS A RESULT, THE STAFF HAD TO CHANGE THE BOX OF WIRES, WITH THE SAME REFERENCE NUMBER BUT UNKNOWN LOT NUMBER. THIS ISSUE CAUSED A LOSS OF TIME. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18894238 ↗
  • MalfunctionMar 13, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A THROMBOENDOARTERECTOMY (TEA) OF THE RIGHT CAROTID ARTERY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SURGEON WANTING TO ANASTOMOSE THE CAROTID ARTERY, USED THE PROLENE WIRE. BUT ON SEVERAL OCCASIONS, THE THREADS BROKE AT THE SLIGHTEST PRESSURE (EVEN MINIMAL) AND THE NEEDLE PULLED OFF. AS A RESULT, THE STAFF HAD TO CHANGE THE BOX OF WIRES, WITH THE SAME REFERENCE NUMBER BUT UNKNOWN LOT NUMBER. THIS ISSUE CAUSED A LOSS OF TIME. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18894235 ↗
  • MalfunctionJul 7, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE TEARS OFF THE THREAD. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14944880 ↗
  • MalfunctionDec 22, 2021Material Separation; Material Separation

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2021 SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE PULLED OFF OF THE THREAD. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13059411 ↗
  • MalfunctionSep 23, 2021Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF OF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12519587 ↗
  • MalfunctionSep 23, 2021Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF OF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12519586 ↗
  • MalfunctionSep 23, 2021Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF OF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12519580 ↗
  • MalfunctionMay 26, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) OF 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE SPLIT AND TORE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11889125 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.

  • Z-1515-2025 · Mar 7, 2025 · Open, Classified
    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
  • Z-1096-2025 · Dec 20, 2024 · Open, Classified
    PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
  • Z-2466-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE 122CM M1.5, D10189
  • Z-2463-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
  • Z-2461-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
  • Z-2450-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
  • Z-2460-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
  • Z-2457-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
  • Z-2453-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
  • Z-2449-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
  • Z-2455-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
  • Z-2454-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
  • Z-2464-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
  • Z-2459-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
  • Z-2462-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 3-0 BLU, D9844

…and 6 more.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →

Sources and updates