ETHILON 9003GRecalls & reported problems
Ethicon, LLC · Black Monofilament Nylon, Nonabsorbable Surgical Suture · Class 2 · Implant
All models of ETHILON (133)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10705031022355), received between Feb 20, 2019 and Jan 29, 2026 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
8 other recalls cover devices cleared under the same FDA submission (K946173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Material Separation3
- Device Contaminated During Manufacture or Shipping1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 29, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CATARACT REMOVAL AND INSERTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. KEPT DISINTEGRATING OR BREAKING WITH LITTLE TO NO FORCE. DOES NOT HOLD A SUTURE. DISCOVERED DURING PATIENT USE OPEN ANOTHER UNIT. NO ADVERSE EVENT. THERE WERE NO SIGNIFICANT DELAYS. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24208434 ↗MalfunctionJan 29, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CATARACT REMOVAL AND INSERTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. KEPT DISINTEGRATING OR BREAKING WITH LITTLE TO NO FORCE. DOES NOT HOLD A SUTURE. DISCOVERED DURING PATIENT USE OPEN ANOTHER UNIT. NO ADVERSE EVENT. THERE WERE NO SIGNIFICANT DELAYS. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24208433 ↗MalfunctionMay 7, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. WHEN THEY OPENED THE SUTURE BOX THERE IS A RED FOREIGN MATTER IN THE BOX. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21976805 ↗MalfunctionJan 14, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2024 AND SUTURE WAS USED. THE SUTURE SNAPPED IMMEDIATELY AND BROKE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21152333 ↗MalfunctionJan 14, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE SNAPPED IMMEDIATELY AND BROKE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21152300 ↗MalfunctionAug 18, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. WHEN IT WAS USED, THE SUTURE WAS LOOSE AND COULD NOT BE HANDLED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12340225 ↗MalfunctionJul 27, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE DETACHED FROM THE SUTURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12230159 ↗MalfunctionFeb 20, 2019Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN OPHTHALMIC OPERATION ON AN UNKNOWN DATE AND SUTURE WAS USED. THE SUTURE WAS DETACHED FROM THE NEEDLE EASILY WHEN A SCRUB NURSE HOLDING THE NEEDLE IN USUAL WAY. THE LOT NUMBER IS UNKNOWN. FURTHER DETAILS ARE NOT PROVIDED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8353696 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K946173. One submission often covers many models, so check the product and lot details in each record.
- Z-1516-2025 · Mar 7, 2025 · Open, ClassifiedETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
- Z-1097-2025 · Dec 20, 2024 · Open, ClassifiedPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
- Z-2486-2020 · Jun 10, 2020 · Open, ClassifiedPERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
- Z-1841-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
- Z-1842-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
- Z-0747-2019 · Dec 18, 2018 · TerminatedPERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
- Z-1839-2015 · May 8, 2015 · TerminatedEthicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
- Z-0492-2015 · Oct 17, 2014 · TerminatedEthicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE devices
- ETHILON 697GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE223 reports
- ETHILON 698GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE119 reports
- ETHILON 1696GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE118 reports
- Monosof SN-3695Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls3 reports
- McKesson S931BXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls2 reports
- McKesson S661GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls2 reports
- Surgilon 88861919-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls2 reports
- McKesson S697GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Synthetic, Polyamide2 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page