PROLENE D9833Recalls & reported problems
Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10705031035126), received between Jul 11, 2023 and May 15, 2026 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (11 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Markings/Labelling Problem11
- Break6
- Material Frayed5
Patient problems
- No Clinical Signs, Symptoms or Conditions17
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2026Break; Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS FRAYED AND THE DOCTOR FELT THAT THE SUTURE WAS BROKEN EASILY DURING SURGERY. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25190786 ↗MalfunctionMay 15, 2026Break; Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS FRAYED AND THE DOCTOR FELT THAT THE SUTURE WAS BROKEN EASILY DURING SURGERY. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25190754 ↗MalfunctionMay 15, 2026Break; Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS FRAYED AND THE DOCTOR FELT THAT THE SUTURE WAS BROKEN EASILY DURING SURGERY. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25190750 ↗MalfunctionMay 15, 2026Break; Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS FRAYED AND THE DOCTOR FELT THAT THE SUTURE WAS BROKEN EASILY DURING SURGERY. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25190717 ↗MalfunctionMay 1, 2026Break; Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS FRAYED AND THE DOCTOR FELT THAT THE SUTURE WAS BROKEN EASILY DURING SURGERY. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25057525 ↗MalfunctionMay 29, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE REP THAT TWO SUTURES BROKE WHILE SUTURING UNKNOWN TISSUE. A THIRD LIKE DEVICE FROM A NEW BOX WAS USED TO CONTINUE WITH THE PROCEDURE. UNKNOWN NUMBER OF STERILE SAMPLES RETURNING. EVENT DEVICES MAY BE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22107363 ↗MalfunctionAug 29, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BENTALL PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE SUTURE THICKNESS WRITTEN ON THE SURFACE IN THE PACKAGE (5-0) WAS DIFFERENT FROM THE DESCRIPTION INSIDE THE PACKAGE (6-0). NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17644593 ↗MalfunctionAug 29, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BENTALL PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. WHE SUTURE THICKNESS WRITTEN ON THE SURFACE IN THE PACKAGE (5-0) WAS DIFFERENT FROM THE DESCRIPTION INSIDE THE PACKAGE (6-0). NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17644588 ↗MalfunctionAug 29, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BENTALL PROCEDURE ON 8/7/2023 AND SUTURE WAS USED. THE SUTURE THICKNESS WRITTEN ON THE SURFACE IN THE PACKAGE (5-0) WAS DIFFERENT FROM THE DESCRIPTION INSIDE THE PACKAGE (6-0). NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17644583 ↗MalfunctionAug 16, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON (B)(6)2023 AND SUTURE WAS USED. ALTHOUGH IT WAS 5-0, WHAT WAS PRINTED ON THE PACKAGE WAS THE NOTATION 6-0. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER EVENT DETAILS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17557204 ↗MalfunctionAug 16, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. ALTHOUGH IT WAS 5-0, WHAT WAS PRINTED ON THE PACKAGE WAS THE NOTATION 6-0. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER EVENT DETAILS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17557199 ↗MalfunctionAug 16, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. ALTHOUGH IT WAS 5-0, WHAT WAS PRINTED ON THE PACKAGE WAS THE NOTATION 6-0. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER EVENT DETAILS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17557197 ↗MalfunctionAug 16, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. ALTHOUGH IT WAS 5-0, WHAT WAS PRINTED ON THE PACKAGE WAS THE NOTATION 6-0. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER EVENT DETAILS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17557193 ↗MalfunctionAug 16, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. ALTHOUGH IT WAS 5-0, WHAT WAS PRINTED ON THE PACKAGE WAS THE NOTATION 6-0. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER EVENT DETAILS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17557183 ↗MalfunctionAug 16, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. ALTHOUGH IT WAS 5-0, WHAT WAS PRINTED ON THE PACKAGE WAS THE NOTATION 6-0. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER EVENT DETAILS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17557181 ↗MalfunctionAug 2, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE UNKNOWN SURGERY, THERE WAS A PRINTING ERROR ON THE PACKAGE. ALTHOUGH IT WAS 5-0, WHAT WAS PRINTED INSIDE THE OPENED PACKAGE WAS THE NOTATION 6-0. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17443353 ↗MalfunctionJul 11, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, AFTER OPENING THE PACKAGE, IT WAS FOUND THAT THERE WAS A PRINTING ERROR ON THE PACKAGE. ALTHOUGH THE PRODUCT WAS 5-0, WHAT WAS PRINTED WAS 6-0. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17302681 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
- Z-2456-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
- Z-2450-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2463-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
- Z-2451-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
- Z-2448-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
- Z-2462-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
- Z-2459-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
- Z-2455-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page