Skip to content

PROLENE DC588Recalls & reported problems

Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705031036604), received between Aug 20, 2025 and Aug 20, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (1 report).

21 other recalls cover devices cleared under the same FDA submission (K133356, K811872); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Delivered as Unsterile Product1
  • Difficult to Open or Remove Packaging Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2025Delivered as Unsterile Product; Difficult to Open or Remove Packaging Material; Delivered as Unsterile Product; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN CARDIOVASCULAR PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN CARDIOVASCULAR SURGERY, THE CONTENTS WERE PRESSED AGAINST THE OUTER (THE STERILE PACKAGE) SEALING AND COULD NOT TAKE OUT WHEN TRIED TO TAKE IT OUT FROM THE PACKAGE DURING SURGERY. THE PRODUCT WAS NOT USED BECAUSE THERE WAS A POSSIBILITY THAT STERILIZATION FAILURE OCCURRED. THIS EVENT WAS FOUND BEFORE USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22853160 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133356, K811872. One submission often covers many models, so check the product and lot details in each record.

  • Z-1515-2025 · Mar 7, 2025 · Open, Classified
    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
  • Z-1096-2025 · Dec 20, 2024 · Open, Classified
    PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
  • Z-2462-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
  • Z-2458-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
  • Z-2460-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
  • Z-2457-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
  • Z-2465-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
  • Z-2448-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
  • Z-2453-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
  • Z-2450-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
  • Z-2463-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
  • Z-2452-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
  • Z-2464-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
  • Z-2459-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
  • Z-2461-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746

…and 6 more.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page