Coated VICRYL J345HRecalls & reported problems
Johnson & Johnson International Inc. · Violet Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (50)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Coated VICRYL J345H is a violet braided absorbable surgical suture made by Johnson & Johnson. The 47 reports mostly describe the thread breaking or pulling apart from the needle during surgery, and sometimes not dissolving as expected after healing, along with inflammation, pain, wound reopening and fluid drainage. No deaths were reported; 11 reports involved injuries. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 47
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 47 adverse event reports that list this exact device (DI 10705031039094), received between Oct 2, 2018 and Aug 24, 2026 — classified as 11 injuries, 36 malfunctions.
The most frequently reported specific device problem is “Break” (20 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break20
- Material Separation12
- Adverse Event Without Identified Device or Use Problem8
- Material Integrity Problem7
Patient problems
- No Clinical Signs, Symptoms or Conditions24
- Inflammation12
- Pain9
- Wound Dehiscence7
- Pocket Erosion6
- Fluid Discharge5
- Impaired Healing5
- Skin Inflammation/ Irritation5
Recent adverse event reports
The 25 most recent of 47 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 24, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50410611 ↗MalfunctionDec 16, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. THE NEEDLE AND SUTURE WAS DETACHED WHILE SUTURING. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23818421 ↗MalfunctionDec 12, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. WHILE SUTURING, THE NEEDLE AND SUTURE WAS DETACHED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23790635 ↗MalfunctionJul 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, DURING CESAREAN SECTION SURGERY, A THREAD NEEDLE WAS DETACHED IN THE TISSUE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22559745 ↗MalfunctionApr 29, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A LOWER SEGMENT CESAREAN SECTION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHILE SUTURING THE FASCIA. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21925451 ↗MalfunctionMar 14, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A THREE LOWER UTERINE CESAREAN SECTIONS+LEFT OVARIAN CYSTECTOMY PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. AT 09:20, A PREGNANT WOMAN WITH A HISTORY OF CESAREAN SECTION WAS HOSPITALIZED FOR PLANNED DELIVERY. SURGERY ON THE FOLLOWING DAY FROM 10:38 TO 12:16. DURING THE THREE LOWER UTERINE CESAREAN SECTIONS+LEFT OVARIAN CYSTECTOMY SURGERY, THE SUTURE WAS BROKEN DURING THE SUTURING OF THE PERITONEUM, CAUSING BLEEDING AT THE SUTURE SITE. THE BLEEDING STOPPED AFTER THE SUTURING WAS REPEATED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21610048 ↗MalfunctionMar 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. DURING THE SURGERY, THE SUTURE WAS PULLED APART WHEN IT WAS USED TO SEW THE PERITONEUM. CHANGED TO ANOTHER ONE TO CONTINUE SUTURING. THIS INCREASED THE PATIENT'S MEDICAL EXPENSES AND INCREASED THE OPERATION TIME. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: Insufficient Information
MAUDE report 21587532 ↗MalfunctionJan 27, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. POST-OP, NONABSORPTION OF SUTURE POST-OP. THIS CASE IS FROM HEALTH AUTHORITY. THE SUTURE WAS FOUND NOT ABSORBED AND NEEDED TO BE REMOVED AGAIN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 21246194 ↗MalfunctionJan 7, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE SUTURING PROCESS, THE NEEDLE BROKE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21095592 ↗MalfunctionDec 13, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN EPISIOTOMY PROCEDURE ON (B)(6)2024 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY DURING PASSAGE THROUGH TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. THE 1 ACTUAL SAMPLE WAS DISCARDED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20922267 ↗MalfunctionOct 9, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE PATIENT UNDERWENT SURGERY AT 3:00 PM. AFTER THE DELIVERY OF THE FETUS, THE DOCTOR SUTURED IT IN TWO LAYERS WITH SUTURE. BEFORE SUTURING, THE CUTTING EDGE, ESPECIALLY THE CORNERS ON BOTH SIDES, WAS CHECKED FOR TEARING AND ELONGATION. THE FIRST LAYER IS CONTINUOUSLY SUTURED WITHOUT PENETRATING THE ENDOMETRIAL LAYER. THE SECOND LAYER OF CONTINUOUS MATTRESS EMBEDDING SUTURE WAS PERFORMED WITH A NEEDLE INSERTION DEPTH OF 2/3 OF THE CUTTING EDGE. WHEN SUTURING THE SECOND LAYER OF TISSUE, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED, AND THE UTERINE WOUND WAS STILL BLEEDING. THE DOCTOR IMMEDIATELY REMOVED THE SUTURE, REPLACED IT WITH A NEW ONE, AND SUTURED THE PATIENT AGAIN. THE HEMOSTASIS WAS SUCCESSFUL AND THE SURGERY WENT SMOOTHLY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20413326 ↗MalfunctionSep 24, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT FULL-TERM PREGNANCY ON (B)(6) 2024 AND SUTURE WAS USED. AT 9:30, THE PATIENT UNDERWENT SURGERY UNDER SPINAL ANESTHESIA AND WAS SUCCESSFULLY DELIVERED. THE DOCTOR USED SUTURE TO SUTURE THE SKIN, BUT THE SUTURE BROKE. THE DOCTOR CHECKED THAT THE SKIN NEEDLE WAS INTACT. IMMEDIATELY REPLACE THE SKIN NEEDLE WITH A NEW ONE FOR SUTURING THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20296486 ↗InjuryApr 16, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A VAGINAL DELIVERY ON (B)(6) 2024 AND SUTURE WAS USED. THE PARTURIENT DELIVERED A MATURE LIVE FEMALE INFANT THROUGH A PERINEAL INCISION. THE DOCTOR SUTURED THE PERINEAL INCISION. FOUND POOR ABSORPTION AND HEALING OF THE SUTURE, REMOVED THE SUTURE, DISINFECTED IT, AND SUTURED IT AGAIN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Impaired Healing
MAUDE report 19121389 ↗MalfunctionDec 21, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ENDOSCOPIC OVARIAN LESION RESECTION PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. AT 10:30 AM, DURING THE USE OF SUTURE FOR OVARIAN RECONSTRUCTION, THE SUTURE BROKE. THE DOCTOR CHECKED THAT THERE WAS NO OBVIOUS DAMAGE AT THE SUTURED INCISION. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18373972 ↗MalfunctionDec 20, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THE SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18366763 ↗MalfunctionDec 11, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT C-SECTION PROCEDURE ON 22 OCT 2023 AND SUTURE WAS USED. THE PATIENT WAS ADMITTED TO THE HOSPITAL FOR SURGERY DUE TO FULL TERM PREGNANCY, AND THE DELIVERY WAS SMOOTH. DURING THE PROCEDURE, THE SUTURE BROKE WHILE THE DOCTOR WAS SUTURING THE SKIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED BUT UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18300785 ↗MalfunctionNov 14, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18132178 ↗MalfunctionSep 14, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED WHEN SURGEON USED SUTURE TO SUTURE THE PATIENT'S WOUND. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17749659 ↗InjuryJul 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE FOR A PERINEAL LACERATION ON (B)(6) 2023 AND SUTURE WAS USED. ON THE 6TH DAY AFTER PERINEAL LACERATION SURGERY, THE PATIENT COMPLAINED OF PERINEAL PAIN AND DEHISCENCE. THE PATIENT HAD INFLAMMATION. THE PATIENT'S VULVAR WOUND WAS RED, SWOLLEN, TINGLING, WITH EXPOSED SUTURE, AND DID NOT HEAL. THE DOCTOR CUT OFF THE SURFACE SUTURE AND CLEANED IT WITH POTASSIUM PERMANGANATE SOLUTION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Inflammation; Itching Sensation; Pain; Pocket Erosion; Skin Inflammation/ Irritation; Swelling/ Edema; Wound Dehiscence
MAUDE report 17405818 ↗InjuryJul 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON (B)(6) 2023 DUE TO UMBILICAL CORD PROLAPSE AND OBSTRUCTIVE DELIVERY AND SUTURE WAS USED ON THE SURGICAL INCISION. ON THE 6TH DAY AFTER THE SURGERY, THE PATIENT'S SURGICAL INCISION APPEARED EXUDATION. THE PATIENT HAD POOR WOUND HEALING AND WOUND SECRETION. THE DOCTOR GAVE THE EXUDATE BACTERIA NEGATIVE CULTURE. AFTER THE SUTURE WAS REMOVED, THE SURGICAL INCISION DID NOT HEAL WELL. THE PATIENT WAS GIVEN DEBRIDEMENT AND DRESSING CHANGE, AND STERILE GAUZE WAS GIVEN TO COVER THE SURGICAL INCISION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Fluid Discharge; Impaired Healing
MAUDE report 17405437 ↗MalfunctionJul 17, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT SURGERY FOR THE DELIVERY OF THE PATIENT'S SECOND PREGNANCY WITH SUTURING OF PERINEAL LACERATION ON AN UNKNOWN DATE AND SUTURE WAS USED. NINE DAYS AFTER THE SURGERY, THE PATIENT COMPLAINED OF PERINEAL PAIN AND RUPTURE. THE PATIENT'S VULVAR WOUND IS RED, SWOLLEN, STINGING, WITH EXPOSED SUTURE, AND DOES NOT HEAL. THE PATIENT HAD POST OP INFLAMMATION. CUT OFF THE SURFACE SUTURE ACCORDING TO THE DOCTOR'S INSTRUCTIONS AND CLEAN IT WITH POTASSIUM PERMANGANATE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Erythema; Impaired Healing; Inflammation; Pain; Pocket Erosion; Skin Inflammation/ Irritation; Swelling/ Edema
MAUDE report 17332888 ↗MalfunctionJul 6, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A GYNECOLOGICAL PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, WHILE SUTURING THE PATIENT¿S PERINEUM, THE DOCTOR NOTICED THE TIP OF THE SUTURE APPEARS TO BE MISSING. SHE BELIEVES IT BROKE OFF- AS THE INITIAL RUNNING STITCH WENT IN AS NORMAL, BUT THE FIRST OF HER BACK STITCH WOULD NOT PUNCTURE THE SKIN. NO FURTHER SUTURE WAS DONE WITH THIS ATRAUMATIC NEEDLE. A REPLACEMENT SUTURE WAS GIVEN WHICH WAS THE SAME LOT TO DOCTOR AND IT WAS NOT DEFECTIVE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17273893 ↗InjuryApr 27, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LATERAL EPISIOTOMY PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. POST-OP, ON THE 26TH DAY AFTER THE PROCEDURE, THE PATIENT COMPLAINED OF PERINEAL PAIN AND RUPTURE. THE PATIENT HAD INFLAMMATION. THE PATIENT'S EXTERNAL GENITALIA WOUND WAS RED, SWOLLEN, STINGING, EXUDING, WITH EXPOSED SUTURE, AND DOES NOT HEAL. THE DOCTOR CUT OFF THE SURFACE SUTURE OF THE EXTERNAL GENITALIA, SUTURED IT AGAIN, AND WASHED IT WITH POTASSIUM PERMANGANATE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Fluid Discharge; Inflammation; Pain; Pocket Erosion; Skin Inflammation/ Irritation; Swelling/ Edema; Wound Dehiscence
MAUDE report 16825994 ↗MalfunctionJan 23, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A WOUND CLOSURE PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. POST-OP, THE DOCTOR USED SUTURE TO SEW THE PATIENT'S WOUND DUE TO PERINEAL LACERATION, AND THE PERINEAL SUTURE WAS SUCCESSFULLY HOSPITALIZED WITHOUT DISCOMFORT. IN THE PUERPERIUM, THERE WAS SUTURE REACTION, THE SUTURE WAS NOT ABSORBED, AND THE WOUND WAS TINGLING, SO THE DOCTOR REMOVED THE SUTURE AND THE PAIN WAS RELIEVED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Inflammation; Insufficient Information; Local Reaction; Pocket Erosion
MAUDE report 16230901 ↗MalfunctionDec 29, 2022Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. AFTER THE SURGERY, THE PATIENT SUFFERED FROM UN-ABSORPTION OF SUTURE, SLIGHT EXUDATION OF ABDOMINAL WOUND, AND SLIGHTLY RED SKIN. TWO DAYS AFTER THE DOCTOR REMOVED THE SUTURE BY HAND, THE PATIENTS WOUND HEALED NATURALLY AND FINALLY TURNED TO WOUND HEALING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: Fluid Discharge; Inflammation; Skin Inflammation/ Irritation
MAUDE report 16068017 ↗
FDA clearance and classification
The FDA regulates product code GAM (suture, absorbable, synthetic, polyglycolic acid) under 21 CFR 878.4493 as a Class II device. 21 CFR 878.4493 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Coated VICRYL J345H
Has the Coated VICRYL J345H been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705031039094), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Coated VICRYL J345H?
The FDA's MAUDE database lists 47 adverse event reports for this device, received between Oct 2, 2018 and Aug 24, 2026. In those reports, the most common specific device problems are “Break” (20), “Material Separation” (12) and “Material Integrity Problem” (7); the most common patient problems are “Inflammation” (12), “Pain” (9) and “Wound Dehiscence” (7). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Coated VICRYL J345H?
None of the reports are classified as deaths, 11 as injuries and 36 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Coated VICRYL J345H MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the Coated VICRYL J345H, and is it still sold?
The FDA's device database lists Johnson & Johnson International Inc. as the company responsible for it (brand Coated VICRYL, model J345H). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,085 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID502 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death402 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID329 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page