Coated VICRYL J346HRecalls & reported problems
Johnson & Johnson Medical GmbH · Violet Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Johnson & Johnson Medical coated VICRYL absorbable suture has 93 adverse event reports. They mostly describe needles breaking or suture separating from the needle during surgery, sometimes leaving fragments needing imaging to check; 2 reports mention foreign bodies in patients and 2 wound dehiscence. There were no deaths and 1 injury. No recalls name this device. Reports fell from 9 to 2 between the prior and recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 93
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 93 adverse event reports that list this exact device (DI 10705031039117), received between Apr 29, 2019 and Jan 22, 2026 — classified as 1 injury, 92 malfunctions.
The most frequently reported specific device problem is “Material Separation” (50 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation50
- Break44
- Material Deformation1
- Material Frayed1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions64
- No Consequences Or Impact To Patient24
- Foreign Body In Patient2
- Insufficient Information2
- Wound Dehiscence2
Recent adverse event reports
The 25 most recent of 93 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 22, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE ON TWO DIFFERENT CASES. THE SURGEON WAS CLOSING THE UTERINE CAVITY. THE NEEDLE TIP BROKE OFF WHILE SUTURING AND WE WERE UNABLE TO LOCATE THE NEEDLE TIP. THEY HAD TO PERFORM AN ABDOMINAL X-RAY TO CONFIRM THERE WAS NO FOREIGN BODY. THE SECOND TIME THE NEEDLE TIP BROKE OFF, THE NEEDLE WAS PLACED IN THE NEEDLE COUNTER AND PROCEEDED TO BREAK OFF. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24147875 ↗MalfunctionJan 22, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE ON TWO DIFFERENT CASES. THE SURGEON WAS CLOSING THE UTERINE CAVITY. THE NEEDLE TIP BROKE OFF WHILE SUTURING AND WE WERE UNABLE TO LOCATE THE NEEDLE TIP. THEY HAD TO PERFORM AN ABDOMINAL X-RAY TO CONFIRM THERE WAS NO FOREIGN BODY. THE SECOND TIME THE NEEDLE TIP BROKE OFF, THE NEEDLE WAS PLACED IN THE NEEDLE COUNTER AND PROCEEDED TO BREAK OFF. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24147874 ↗MalfunctionJul 30, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAESAREAN SECTION PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE THREAD BECAME SEPARATED FROM THE NEEDLE. THERE WAS NO HARM ON PATIENT. HAPPENED DURING REPAIR OF THE UTERUS. THE SUTURE WAS REPLACED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22663758 ↗MalfunctionJul 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. WHEN CLOSING THE FASCIA ON THE PATIENT THE NEEDLE WAS BLUNT INCISION CLOSURE WAS STOP THE NEEDLE POINT BROKE OFF. XRAY WAS NEGATIVE FOR ANY FOREIGN MATERIALS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22513136 ↗MalfunctionJul 3, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, NEEDLE POPPED OFF DURING SUTURING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22396576 ↗MalfunctionApr 24, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21904099 ↗MalfunctionApr 2, 2025Break; Break
IT WAS REPORTED AN ANIMAL UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, PROCEDURE THE SUTURE WOULD START TO DISSOLVE WITH IN DAYS CREATING WOUND DEHISCE. ANIMAL CURRENTLY DOING WELL. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Wound Dehiscence
MAUDE report 21754582 ↗MalfunctionApr 2, 2025Break; Break
IT WAS REPORTED A ANIMAL UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, PROCEDURE THE SUTURE WOULD START TO DISSOLVE WITH IN DAYS CREATING WOUND DEHISCE. ANIMAL CURRENTLY DOING WELL. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Wound Dehiscence
MAUDE report 21753327 ↗MalfunctionFeb 25, 2025Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN OPEN GYNECOLOGICAL PROCEDURE ON AN UNKNOWN DATE IN 2025 AND SUTURE WAS USED. THE NEEDLE HAD DETACHED DIRECTLY FROM THE SUTURE WITHOUT STRONG TRACTION WHEN PULLING THROUGH THE TISSUE. THE OPERATION WAS CONTINUED AND COMPLETED AS PLANNED USING SUTURES FROM A DIFFERENT BATCH. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED. NO USED SUTURES WERE RETAINED, BUT DISCARDED DIRECTLY INTRA-OPERATIVELY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21457810 ↗MalfunctionFeb 25, 2025Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN OPEN GYNECOLOGICAL PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. ONCE THE NEEDLE BROKE WITHOUT EXCESSIVE FORCE; THE BROKEN PIECE COULD BE RECOVERED. THE OPERATION WAS CONTINUED AND COMPLETED AS PLANNED USING SUTURES FROM A DIFFERENT BATCH. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED. NO USED SUTURES WERE RETAINED, BUT DISCARDED DIRECTLY INTRA-OPERATIVELY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21457681 ↗MalfunctionFeb 25, 2025Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN OPEN GYNECOLOGICAL PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE NEEDLE HAD DETACHED DIRECTLY FROM THE SUTURE WITHOUT STRONG TRACTION WHEN PULLING THROUGH THE TISSUE. THE OPERATION WAS CONTINUED AND COMPLETED AS PLANNED USING SUTURES FROM A DIFFERENT BATCH. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED. NO USED SUTURES WERE RETAINED, BUT DISCARDED DIRECTLY INTRA-OPERATIVELY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21457680 ↗MalfunctionOct 22, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP IS REPORTING FROM THE CUSTOMER THAT THE NEEDLE HAD CAME OFF OF THE SUTURE TWICE. NO PATIENT HARM REPORTED AND NO ADDITIONAL INFORMATION GIVEN TO REP. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20504511 ↗MalfunctionOct 22, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP IS REPORTING FROM THE CUSTOMER THAT THE NEEDLE HAD CAME OFF OF THE SUTURE TWICE. NO PATIENT HARM REPORTED AND NO ADDITIONAL INFORMATION GIVEN TO REP. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20504510 ↗MalfunctionSep 19, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. SURGEON SUTURING WOUND BED CLOSED AND SECOND NEEDLE BROKE DURING USE AND LODGED INTO PATIENT'S TISSUE; WAS SEARCHED FOR AND RECOVERED FROM PATIENT'S BACK. ALL PIECES ACCOUNTED FOR AND NONE LEFT IN PATIENT. ALL PIECES ACCOUNTED FOR AND REMOVED FROM STERILE FIELD. NO OTHER DETAILS PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20263883 ↗MalfunctionSep 19, 2024Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. SURGEON SUTURING WOUND BED CLOSED AND NEEDLE BROKE DURING USE. FIRST SUTURE NEEDLE FELL TO FLOOR WHEN BROKEN AND WAS RECOVERED. ALL PIECES ACCOUNTED FOR AND NONE LEFT IN PATIENT. ALL PIECES ACCOUNTED FOR AND REMOVED FROM STERILE FIELD. NO OTHER DETAILS PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20263882 ↗MalfunctionAug 26, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES HAVE NEEDLE "POPPING OFF" OF SUTURE TOO EASILY FOR 2 DIFFERENT SURGEONS IN THE BIRTHING CENTER. ALMOST LIKE IT WAS A CODE OF OUR "POP OFF NEEDLES". THERE WAS NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20065644 ↗MalfunctionAug 26, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES POPPED OFF NEEDLE INTRA OP DURING UTERINE CLOSURE FROM C SECTION.THERE WAS NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20065621 ↗MalfunctionAug 22, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. IT APPEARS THAT THE THREAD COMES OFF THE NEEDLE EASILY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20026850 ↗MalfunctionAug 22, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. IT APPEARS THAT THE THREAD COMES OFF THE NEEDLE EASILY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20026681 ↗MalfunctionAug 21, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. IT APPEARS THAT THE THREAD COMES OFF THE NEEDLE EASILY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20018391 ↗MalfunctionAug 21, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. IT APPEARS THAT THE THREAD COMES OFF THE NEEDLE EASILY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20017954 ↗MalfunctionAug 21, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. IT APPEARS THAT THE THREAD COMES OFF THE NEEDLE EASILY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20016260 ↗MalfunctionAug 21, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A C-SECTION PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT APPEARS THAT THE THREAD COMES OFF THE NEEDLE EASILY. THIS HAS HAPPENED REPEATEDLY SINCE LAST WEEK 6-8 TIMES. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20015575 ↗MalfunctionAug 21, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. IT APPEARS THAT THE THREAD COMES OFF THE NEEDLE EASILY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20013462 ↗MalfunctionAug 21, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. IT APPEARS THAT THE THREAD COMES OFF THE NEEDLE EASILY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20013284 ↗
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page