Coated VICRYL J386HRecalls & reported problems
Ethicon, LLC · Violet Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (350)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Coated VICRYL J386H is a violet braided absorbable synthetic suture used to close surgical wounds. The 35 reports mostly describe the suture or its needle breaking during surgery, and in several cases the broken needle fell into the patient's body but was retrieved. No deaths and no injuries were reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 35
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 10705031039490), received between Oct 29, 2020 and May 29, 2025 — classified as 35 malfunctions.
The most frequently reported specific device problem is “Break” (34 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break34
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions35
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 29, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNSPECIFIED PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE NEEDLE BROKE WHILE THEY WERE PERFORMING THE PROCEDURE. THE CUSTOMER CONFIRMED THAT THEY HAD USED A BRAND-NEW BOX. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22110675 ↗MalfunctionMay 29, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNSPECIFIED PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE NEEDLE BROKE WHILE THEY WERE PERFORMING THE PROCEDURE. THE CUSTOMER CONFIRMED THAT THEY HAD USED A BRAND-NEW BOX. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22110673 ↗MalfunctionMar 29, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. THE BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19002914 ↗MalfunctionMar 29, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. THE BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19002913 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968303 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968302 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968299 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968296 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968289 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968287 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968285 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968282 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968280 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968278 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968273 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968271 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968264 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968262 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024. MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY, BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. NO PATIENT CONSEQUENCE WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968128 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024. MULTIPLE INCIDENTS OCCURRED DURING ORTHOPEDIC SURGERIES AND GS PROCEDURES. BROKEN NEEDLES FELL INTO THE PATIENT'S BODY, BUT WERE ALL TAKEN OUT WITHOUT ANY OBVIOUS NEGATIVE IMPACT. NO PATIENT CONSEQUENCE WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968126 ↗MalfunctionFeb 9, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE DURING SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18675856 ↗MalfunctionFeb 9, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE DURING SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18675852 ↗MalfunctionFeb 9, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE DURING SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18675851 ↗MalfunctionFeb 6, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE HAS BEEN BREAKING, THEY HAVE HAD THREE BREAK WITHIN THE LAST TWO DAYS. THEY HAVE USED THIS BRAND FOR THE LAST SEVERAL YEARS WITH NO ISSUES BUT THIS IS OBVIOUSLY A SAFETY CONCERN FOR THEM. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18651843 ↗MalfunctionFeb 6, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE HAS BEEN BREAKING, THEY HAVE HAD THREE BREAK WITHIN THE LAST TWO DAYS. THEY HAVE USED THIS BRAND FOR THE LAST SEVERAL YEARS WITH NO ISSUES BUT THIS IS OBVIOUSLY A SAFETY CONCERN FOR THEM. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18651842 ↗
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page