Coated VICRYL J453HRecalls & reported problems

Johnson & Johnson International Inc. · Violet Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant

All models of Coated VICRYL (50)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705031040021), received between Jun 17, 2019 and Jan 29, 2025 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Break” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break10
  • Material Separation4
  • Material Twisted/Bent2

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient6

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN PROCEDURE, THE SUTURE IS COMING OFF THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21263173 ↗
  • MalfunctionJan 13, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN PROCEDURE, THE SUTURE IS COMING OFF THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21139417 ↗
  • MalfunctionDec 17, 2024Break; Material Separation; Material Twisted/Bent; Break; Material Separation; Material Twisted/Bent

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP IS REPORTING THAT THE SUTURE IS BREAKING AT TIMES ALONG WITH THE NEEDLES POPPING OFF AND BENDING WHEN PASSING THROUGH. THEY HAD TO KEEP USING MORE UNTIL IT WORKED TO COMPLETE THE CASE. NEEDLES DID NOT FALL INTO THE PATIENT AND THERE WAS NO PATIENT HARM REPORTED.. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20951474 ↗
  • MalfunctionDec 17, 2024Break; Material Separation; Material Twisted/Bent; Break; Material Separation; Material Twisted/Bent

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP IS REPORTING THAT THE SUTURE IS BREAKING AT TIMES ALONG WITH THE NEEDLES POPPING OFF AND BENDING WHEN PASSING THROUGH. THEY HAD TO KEEP USING MORE UNTIL IT WORKED TO COMPLETE THE CASE. NEEDLES DID NOT FALL INTO THE PATIENT AND THERE WAS NO PATIENT HARM REPORTED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20951466 ↗
  • MalfunctionDec 1, 2023Break; Break

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT A SPAY PROCEDURE ON 30-OCT-2023 AND SUTURE WAS USED. THE NEEDLE BROKE OFF WHEN SUTURING INSIDE THE PET. THE PROCEDURE WAS EXTENDED BY AN UNKNOWN LENGTH OF TIME. THERE WERE NO ADVERSE CONSEQUENCES. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18248458 ↗
  • MalfunctionDec 1, 2023Break; Break

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT A SPAY PROCEDURE ON 30-OCT-2023 AND SUTURE WAS USED. THE NEEDLE BROKE OFF WHEN SUTURING INSIDE THE PET. THE PROCEDURE WAS EXTENDED BY AN UNKNOWN LENGTH OF TIME. THERE WERE NO ADVERSE CONSEQUENCES. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18248431 ↗
  • MalfunctionJun 17, 2019Break; Break

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, WHILE THE VETERINARIAN WAS TYING OFF PEDICLES, THE VETERINARIAN COULDN'T APPLY TENSION TO THE SUTURE WITHOUT THE SUTURE BREAKING. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SUTURE. THERE WERE NO CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8704063 ↗
  • MalfunctionJun 17, 2019Break; Break

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, WHILE THE VETERINARIAN WAS TYING OFF PEDICLES, THE VETERINARIAN COULDN'T APPLY TENSION TO THE SUTURE WITHOUT THE SUTURE BREAKING. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SUTURE. THERE WERE NO CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8704061 ↗
  • MalfunctionJun 17, 2019Break; Break

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, WHILE THE VETERINARIAN WAS TYING OFF PEDICLES, THE VETERINARIAN COULDN'T APPLY TENSION TO THE SUTURE WITHOUT THE SUTURE BREAKING. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SUTURE. THERE WERE NO CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8704060 ↗
  • MalfunctionJun 17, 2019Break; Break

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, WHILE THE VETERINARIAN WAS TYING OFF PEDICLES, THE VETERINARIAN COULDN'T APPLY TENSION TO THE SUTURE WITHOUT THE SUTURE BREAKING. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SUTURE. THERE WERE NO CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8704059 ↗
  • MalfunctionJun 17, 2019Break; Break

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, WHILE THE VETERINARIAN WAS TYING OFF PEDICLES, THE VETERINARIAN COULDN'T APPLY TENSION TO THE SUTURE WITHOUT THE SUTURE BREAKING. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SUTURE. THERE WERE NO CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8704057 ↗
  • MalfunctionJun 17, 2019Break; Break

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, WHILE THE VETERINARIAN WAS TYING OFF PEDICLES, THE VETERINARIAN COULDN'T APPLY TENSION TO THE SUTURE WITHOUT THE SUTURE BREAKING. THE PROCEDURE WAS COMPLETED USING ADDITIONAL SUTURE. THERE WERE NO CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8704053 ↗

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page