Coated VICRYL J490GRecalls & reported problems
Ethicon, LLC · Undyed Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (350)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Coated VICRYL J490G absorbable surgical suture is used to close tissue. The 67 reports mostly describe the suture material separating or breaking, often detaching from the needle mid-procedure or fraying so it cannot be used; no deaths or injuries were reported. No recalls name this device. Reports in the last 12 months fell to 0 from 15 in the prior12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 67
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 67 adverse event reports that list this exact device (DI 10705031040441), received between Mar 3, 2021 and Apr 25, 2025 — classified as 67 malfunctions.
The most frequently reported specific device problem is “Material Separation” (53 reports).
3 other recalls cover devices cleared under the same FDA submission (K022269); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation53
- Break9
- Material Frayed4
- Component Misassembled1
Patient problems
- No Clinical Signs, Symptoms or Conditions67
Recent adverse event reports
The 25 most recent of 67 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 25, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IN MID SURGERY, THE NEEDLE DETACHED FROM THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21908742 ↗MalfunctionFeb 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN ANIMAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THEY HAD JUST OPENED A NEW BOX OF SUTURE FOR A DENTAL APPOINTMENT A WEEK TO TWO WEEKS AGO. UNFORTUNATELY, THE SUTURES SEEMED TO BREAK OFF FROM THE NEEDLE AND THEN BECAME UNUSABLE. THEY BROKE 7 OUT OF THE 12 SUTURES IN THIS BOX. SHE ONLY HAD 3 SUTURES LEFT OUT OF THIS BOX SO THEY WERE ONLY ABLE TO USE 2 SUTURES SUCCESSFULLY. THE PATIENT DID NOT HAVE ANY ADVERSE REACTION TO THE SUTURE. THE DATE WE NOTICED WAS (B)(6) 2025 WE BROKE 6 THAT DAY AND WE BROKE ONE SUTURE ON (B)(6) 2025. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431963 ↗MalfunctionFeb 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN ANIMAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THEY HAD JUST OPENED A NEW BOX OF SUTURE FOR A DENTAL APPOINTMENT A WEEK TO TWO WEEKS AGO. UNFORTUNATELY, THE SUTURES SEEMED TO BREAK OFF FROM THE NEEDLE AND THEN BECAME UNUSABLE. THEY BROKE 7 OUT OF THE 12 SUTURES IN THIS BOX. SHE ONLY HAD 3 SUTURES LEFT OUT OF THIS BOX SO THEY WERE ONLY ABLE TO USE 2 SUTURES SUCCESSFULLY. THE PATIENT DID NOT HAVE ANY ADVERSE REACTION TO THE SUTURE. THE DATE WE NOTICED WAS (B)(6) 2025 WE BROKE 6 THAT DAY AND WE BROKE ONE SUTURE ON (B)(6) 2025. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431962 ↗MalfunctionFeb 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN ANIMAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THEY HAD JUST OPENED A NEW BOX OF SUTURE FOR A DENTAL APPOINTMENT A WEEK TO TWO WEEKS AGO. UNFORTUNATELY THE SUTURES SEEMED TO BREAK OFF FROM THE NEEDLE AND THEN BECAME UNUSABLE. THEY BROKE 7 OUT OF THE 12 SUTURES IN THIS BOX. SHE ONLY HAD 3 SUTURES LEFT OUT OF THIS BOX SO THEY WERE ONLY ABLE TO USE 2 SUTURES SUCCESSFULLY. THE PATIENT DID NOT HAVE ANY ADVERSE REACTION TO THE SUTURE. THE DATE WE NOTICED WAS (B)(6) 2025 WE BROKE 6 THAT DAY AND WE BROKE ONE SUTURE ON (B)(6) 2025. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431960 ↗MalfunctionFeb 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN ANIMAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THEY HAD JUST OPENED A NEW BOX OF SUTURE FOR A DENTAL APPOINTMENT A WEEK TO TWO WEEKS AGO. UNFORTUNATELY THE SUTURES SEEMED TO BREAK OFF FROM THE NEEDLE AND THEN BECAME UNUSABLE. THEY BROKE 7 OUT OF THE 12 SUTURES IN THIS BOX. SHE ONLY HAD 3 SUTURES LEFT OUT OF THIS BOX SO THEY WERE ONLY ABLE TO USE 2 SUTURES SUCCESSFULLY. THE PATIENT DID NOT HAVE ANY ADVERSE REACTION TO THE SUTURE. THE DATE WE NOTICED WAS 1-14-25 WE BROKE 6 THAT DAY AND WE BROKE ONE SUTURE ON 1-21-25. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431959 ↗MalfunctionFeb 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN ANIMAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THEY HAD JUST OPENED A NEW BOX OF SUTURE FOR A DENTAL APPOINTMENT A WEEK TO TWO WEEKS AGO. UNFORTUNATELY THE SUTURES SEEMED TO BREAK OFF FROM THE NEEDLE AND THEN BECAME UNUSABLE. THEY BROKE 7 OUT OF THE 12 SUTURES IN THIS BOX. SHE ONLY HAD 3 SUTURES LEFT OUT OF THIS BOX SO THEY WERE ONLY ABLE TO USE 2 SUTURES SUCCESSFULLY. THE PATIENT DID NOT HAVE ANY ADVERSE REACTION TO THE SUTURE. THE DATE WE NOTICED WAS 1-14-25 WE BROKE 6 THAT DAY AND WE BROKE ONE SUTURE ON (B)(6)2025. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431958 ↗MalfunctionFeb 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN ANIMAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THEY HAD JUST OPENED A NEW BOX OF SUTURE FOR A DENTAL APPOINTMENT A WEEK TO TWO WEEKS AGO. UNFORTUNATELY, THE SUTURES SEEMED TO BREAK OFF FROM THE NEEDLE AND THEN BECAME UNUSABLE. THEY BROKE 7 OUT OF THE 12 SUTURES IN THIS BOX. SHE ONLY HAD 3 SUTURES LEFT OUT OF THIS BOX SO THEY WERE ONLY ABLE TO USE 2 SUTURES SUCCESSFULLY. THE PATIENT DID NOT HAVE ANY ADVERSE REACTION TO THE SUTURE. THE DATE WE NOTICED WAS (B)(6)2025 WE BROKE 6 THAT DAY AND WE BROKE ONE SUTURE ON (B)(6)2025. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431957 ↗MalfunctionFeb 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN ANIMAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THEY HAD JUST OPENED A NEW BOX OF SUTURE FOR A DENTAL APPOINTMENT A WEEK TO TWO WEEKS AGO. UNFORTUNATELY THE SUTURES SEEMED TO BREAK OFF FROM THE NEEDLE AND THEN BECAME UNUSABLE. THEY BROKE 7 OUT OF THE 12 SUTURES IN THIS BOX. SHE ONLY HAD 3 SUTURES LEFT OUT OF THIS BOX SO THEY WERE ONLY ABLE TO USE 2 SUTURES SUCCESSFULLY. THE PATIENT DID NOT HAVE ANY ADVERSE REACTION TO THE SUTURE. THE DATE WE NOTICED WAS 1-14-25 WE BROKE 6 THAT DAY AND WE BROKE ONE SUTURE ON 1-21-25. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431956 ↗MalfunctionFeb 12, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP BY THE SURGEON THAT WHILE DEVICE WAS IN USE, THE SUTURE POPPED OFF FROM THE NEEDLE. THIS RESULTED IN THE NEED TO GRAB ANOTHER SUTURE. THIS IS THE SAME SUTURE CODE AND LOT NUMBER FROM PREVIOUS CUSTOMER COMPLAINT. ONE DEVICE RETURNING. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21370342 ↗MalfunctionJan 9, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2024 AND SUTURE WAS USED. A SUTURE HAD POPPED OFF DURING USE ON THE PATIENT. THE NURSE WAS NOT ON SITE TO RECOVER SUTURE, BUT THEY WILL BAG THE SUTURE AND NEEDLE IF IT HAPPENS AGAIN. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21116717 ↗MalfunctionJan 9, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. A SUTURE HAD POPPED OFF DURING USE ON THE PATIENT. THE NURSE WAS NOT ON SITE TO RECOVER SUTURE, BUT THEY WILL BAG THE SUTURE AND NEEDLE IF IT HAPPENS AGAIN. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21116460 ↗MalfunctionJan 9, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. A SUTURE HAD POPPED OFF DURING USE ON THE PATIENT. THE NURSE WAS NOT ON SITE TO RECOVER SUTURE, BUT THEY WILL BAG THE SUTURE AND NEEDLE IF IT HAPPENS AGAIN. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21116117 ↗MalfunctionNov 15, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A MOHS CLOSURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. USED ON THE FACE, THE DOCTOR WAS USING THE SUTURE AND THE SUTURE KEPT SLIPPING OUT OF THE NEEDLE. HAD A SMALL DELAY IN CASE BUT WERE ABLE TO COMPLETE BY USING A SUTURE FROM ANOTHER LOT AND BOX. NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20694776 ↗MalfunctionNov 15, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A MOHS CLOSURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. USED ON THE FACE, THE DOCTOR WAS USING THE SUTURE AND THE SUTURE KEPT SLIPPING OUT OF THE NEEDLE. HAD A SMALL DELAY IN CASE BUT WERE ABLE TO COMPLETE BY USING A SUTURE FROM ANOTHER LOT AND BOX. NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20694720 ↗MalfunctionNov 15, 2024Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A MOHS CLOSURE ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. USED ON THE FACE, THE DOCTOR WAS USING THE SUTURE AND THE SUTURE KEPT SLIPPING OUT OF THE NEEDLE. HAD A SMALL DELAY IN CASE BUT WERE ABLE TO COMPLETE BY USING A SUTURE FROM ANOTHER LOT AND BOX. NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20694699 ↗MalfunctionJun 25, 2024Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE FALLS OFF THE NEEDLE WHEN THEY GO TO USE IT, THAT IT IS NOT SWAGED ON FULLY. THEY HAVE TRIED 5 PACKS AND EACH TIME THE SUTURE JUST FALLS RIGHT OFF FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19609841 ↗MalfunctionJun 25, 2024Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE FALLS OFF THE NEEDLE WHEN THEY GO TO USE IT, THAT IT IS NOT SWAGED ON FULLY. THEY HAVE TRIED 5 PACKS AND EACH TIME THE SUTURE JUST FALLS RIGHT OFF FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19609323 ↗MalfunctionJun 25, 2024Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE FALLS OFF THE NEEDLE WHEN THEY GO TO USE IT, THAT IT IS NOT SWAGED ON FULLY. THEY HAVE TRIED 5 PACKS AND EACH TIME THE SUTURE JUST FALLS RIGHT OFF FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19609109 ↗MalfunctionJun 25, 2024Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE FALLS OFF THE NEEDLE WHEN THEY GO TO USE IT, THAT IT IS NOT SWAGED ON FULLY. THEY HAVE TRIED 5 PACKS AND EACH TIME THE SUTURE JUST FALLS RIGHT OFF FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19609015 ↗MalfunctionJun 25, 2024Material Separation; Material Separation
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE FALLS OFF THE NEEDLE WHEN THEY GO TO USE IT, THAT IT IS NOT SWAGED ON FULLY. THEY HAVE TRIED 5 PACKS AND EACH TIME THE SUTURE JUST FALLS RIGHT OFF FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19608965 ↗MalfunctionApr 22, 2024Material Frayed; Material Frayed
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN ANIMAL SURGERY ON AN UNKNOWN DATE; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS COMING UNDONE WHILE SUTURING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19156040 ↗MalfunctionApr 22, 2024Material Frayed; Material Frayed
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN ANIMAL SURGERY ON AN UNKNOWN DATE; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS COMING UNDONE WHILE SUTURING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19154918 ↗MalfunctionApr 22, 2024Material Frayed; Material Frayed
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN ANIMAL SURGERY ON AN UNKNOWN DATE; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS COMING UNDONE WHILE SUTURING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19154908 ↗MalfunctionApr 22, 2024Material Frayed; Material Frayed
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN ANIMAL SURGERY ON AN UNKNOWN DATE; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS COMING UNDONE WHILE SUTURING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19154907 ↗MalfunctionApr 19, 2024Material Separation; Material Separation
IT WAS REPORTED THAT PATIENT UNDERWENT A GENERAL SURGERY PROCEDURE ON AN UNKNOWN DATE IN 2024, AND SUTURE WAS USED. THE SWAGED VICRYL CODE IS PERFORMING LIKE A CONTROL-RELEASE; THE NEEDLE IS POPPING OFF. NO PATIENT HARM REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19144019 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K022269. One submission often covers many models, so check the product and lot details in each record.
- Z-2144-2024 · May 10, 2024 · Open, ClassifiedVICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in cardiovascular and
- Z-0989-2020 · Jan 3, 2020 · TerminatedCoated VICRYL (Polyglactin 910) Suture, 18" (45cm) Product Code J497H. The device is a synthetic absorbable sterile surgical suture.
- Z-1087-06 · May 4, 2006 · TerminatedCoated Vicryl (polyglactin 910) suture. Needle Type CTXB Suture size: 1. Vicryl suture is a synthetic absorbable sterile surgical suture comprised of a copolymer made from 90% gyc
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page