Coated VICRYL J519HRecalls & reported problems
Johnson & Johnson International Inc. · Violet Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (50)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 10705031040892), received between May 13, 2019 and May 2, 2023 — classified as 1 injury, 17 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break11
- Material Separation7
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- No Consequences Or Impact To Patient2
- Abscess1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
InjuryMay 2, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE NEEDLE GOT BROKEN DURING THE CLOSURE OF THE APONEUROSIS. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Abscess; No Clinical Signs, Symptoms or Conditions
MAUDE report 16853108 ↗MalfunctionSep 29, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE GOT DETACHED FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12550193 ↗MalfunctionSep 29, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE GOT DETACHED FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12550191 ↗MalfunctionSep 29, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE GOT DETACHED FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12550188 ↗MalfunctionSep 29, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE GOT DETACHED FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12550187 ↗MalfunctionSep 29, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE GOT DETACHED FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12550186 ↗MalfunctionSep 29, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE GOT DETACHED FROM THE NEEDLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12550184 ↗MalfunctionAug 9, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12297330 ↗MalfunctionAug 9, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12297314 ↗MalfunctionAug 9, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12297304 ↗MalfunctionAug 9, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12297286 ↗MalfunctionJul 9, 2021Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON (B)(6) 2021 AND THE SUTURE WAS USED FOR KNEE CAPSULE CLOSURE. DURING SURGERY, THE SUTURE BROKE. ANOTHER LIKE SUTURE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12147663 ↗MalfunctionJul 9, 2021Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON (B)(6) 2021 AND THE SUTURE WAS USED FOR KNEE CAPSULE CLOSURE. DURING SURGERY, THE SUTURE BROKE. ANOTHER LIKE SUTURE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12147658 ↗MalfunctionJul 9, 2021Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON 6/15/2021 AND THE SUTURE WAS USED FOR KNEE CAPSULE CLOSURE. DURING SURGERY, THE SUTURE BROKE. ANOTHER LIKE SUTURE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12147655 ↗MalfunctionJul 9, 2021Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON (B)(6) 2021 AND THE SUTURE WAS USED FOR KNEE CAPSULE CLOSURE. DURING SURGERY, THE SUTURE BROKE. ANOTHER LIKE SUTURE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12147651 ↗MalfunctionJul 9, 2021Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY ON (B)(6) 2021 AND THE SUTURE WAS USED FOR KNEE CAPSULE CLOSURE. DURING SURGERY, THE SUTURE BROKE. ANOTHER LIKE SUTURE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12147642 ↗MalfunctionJul 17, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT SURGERY IN 2019 AND THE SUTURE WAS USED. DURING CLOSING THE JOINT KNEE CAPSULE, THE SUTURE BROKE. THE PROCEDURE WAS NOT DELAYED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WILL BE REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8801135 ↗MalfunctionMay 13, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE CAME OFF THE END OF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8605053 ↗
FDA clearance and classification
The FDA regulates product code GAM (suture, absorbable, synthetic, polyglycolic acid) under 21 CFR 878.4493 as a Class II device. 21 CFR 878.4493 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Coated VICRYL J519H
Has the Coated VICRYL J519H been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705031040892), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Coated VICRYL J519H?
The FDA's MAUDE database lists 18 adverse event reports for this device, received between May 13, 2019 and May 2, 2023. In those reports, the most common specific device problems are “Break” (11) and “Material Separation” (7); the most common patient problem is “Abscess” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Coated VICRYL J519H?
None of the reports are classified as deaths, 1 as an injury and 17 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Coated VICRYL J519H MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the Coated VICRYL J519H, and is it still sold?
The FDA's device database lists Johnson & Johnson International Inc. as the company responsible for it (brand Coated VICRYL, model J519H). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,085 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID502 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death402 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID329 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page