Coated VICRYL J547GRecalls & reported problems

Ethicon, LLC · Violet Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant

All models of Coated VICRYL (350)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705031041110), received between Sep 25, 2019 and Mar 21, 2025 — classified as 5 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

3 other recalls cover devices cleared under the same FDA submission (K022269); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Adverse Event Without Identified Device or Use Problem5
  • Material Frayed1
  • Material Separation1

Patient problems

  • Discomfort5
  • Inflammation5
  • No Consequences Or Impact To Patient4
  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 21, 2025Break; Material Frayed; Break; Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, VICRYL 8-0 IS PASSED AND THE SUTURE FRAYS AND BREAKS EASILY, SO ANOTHER SUTURE MUST BE PASSED. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21662559 ↗
  • MalfunctionJan 29, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, STITCHES ARE PASSED AND THE NEEDLE IS FRAYED AND WHEN THE KNOT IS TIGHTENED THE SUTURE IS TORN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21265756 ↗
  • MalfunctionJan 29, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, STITCHES ARE PASSED AND THE NEEDLE IS FRAYED AND WHEN THE KNOT IS TIGHTENED THE SUTURE IS TORN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21265748 ↗
  • InjuryDec 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A STRABISMUS SURGERY ON (B)(6) 2020 AND THE SUTURE WAS USED. AFTER THE SURGERY, THE PATIENT EXPERIENCED DISCOMFORT INCLUDING YELLOWING OF THE EYEBALLS. THERE WAS NO SUTURE DEFICIENCY NOTED BEFORE OR DURING USE. POST OP EXAMINATION SHOWED INCREASED INFLAMMATORY RESPONSE. VIGAMOX OPHTHALMIC DROPS, LOTEMAX OPHTHALMIC GEL AND/OR PREDNISONE FORTE 1% OPHTHALMIC DROPS WERE USED TO TREAT THE INFLAMMATORY RESPONSE. THERE WAS NO SURGICAL INTERVENTION PERFORMED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Discomfort; Inflammation

    MAUDE report 10946444 ↗
  • InjuryDec 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A STRABISMUS SURGERY ON (B)(6) 2020 AND THE SUTURE WAS USED. AFTER THE SURGERY, THE PATIENT EXPERIENCED DISCOMFORT INCLUDING YELLOWING OF THE EYEBALLS. THERE WAS NO SUTURE DEFICIENCY NOTED BEFORE OR DURING USE. POST OP EXAMINATION SHOWED INCREASED INFLAMMATORY RESPONSE. VIGAMOX OPHTHALMIC DROPS, LOTEMAX OPHTHALMIC GEL AND/OR PREDNISONE FORTE 1% OPHTHALMIC DROPS WERE USED TO TREAT THE INFLAMMATORY RESPONSE. THERE WAS NO SURGICAL INTERVENTION PERFORMED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Discomfort; Inflammation

    MAUDE report 10946393 ↗
  • InjuryDec 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A PTERYGIUM SURGERY ON (B)(6) 2020 AND THE SUTURE WAS USED. AFTER THE SURGERY, THE PATIENT EXPERIENCED DISCOMFORT INCLUDING YELLOWING OF THE EYEBALLS. THERE WAS NO SUTURE DEFICIENCY NOTED BEFORE OR DURING USE. POST OP EXAMINATION SHOWED INCREASED INFLAMMATORY RESPONSE. VIGAMOX OPHTHALMIC DROPS, LOTEMAX OPHT EYEHALMIC GEL AND/OR PREDNISONE FORTE 1% OPHTHALMIC DROPS WERE USED TO TREAT THE INFLAMMATORY RESPONSE. THERE WAS NO SURGICAL INTERVENTION PERFORMED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Discomfort; Inflammation

    MAUDE report 10946308 ↗
  • InjuryDec 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A STRABISMUS SURGERY ON (B)(6) 2020 AND THE SUTURE WAS USED. AFTER THE SURGERY, THE PATIENT EXPERIENCED DISCOMFORT INCLUDING YELLOWING OF THE EYEBALLS. THERE WAS NO SUTURE DEFICIENCY NOTED BEFORE OR DURING USE. POST OP EXAMINATION SHOWED INCREASED INFLAMMATORY RESPONSE. VIGAMOX OPHTHALMIC DROPS, LOTEMAX OPHTHALMIC GEL AND/OR PREDNISONE FORTE 1% OPHTHALMIC DROPS WERE USED TO TREAT THE INFLAMMATORY RESPONSE. THERE WAS NO SURGICAL INTERVENTION PERFORMED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Discomfort; Inflammation

    MAUDE report 10946303 ↗
  • InjuryDec 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A STRABISMUS SURGERY ON (B)(6) 2020 AND THE SUTURE WAS USED. AFTER THE SURGERY, THE PATIENT EXPERIENCED DISCOMFORT INCLUDING YELLOWING OF THE EYEBALLS. THERE WAS NO SUTURE DEFICIENCY NOTED BEFORE OR DURING USE. POST OP EXAMINATION SHOWED INCREASED INFLAMMATORY RESPONSE. VIGAMOX OPHTHALMIC DROPS, LOTEMAX OPHTHALMIC GEL AND/OR PREDNISONE FORTE 1% OPHTHALMIC DROPS WERE USED TO TREAT THE INFLAMMATORY RESPONSE. THERE WAS NO SURGICAL INTERVENTION PERFORMED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Discomfort; Inflammation

    MAUDE report 10945603 ↗
  • MalfunctionSep 14, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2020; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE DETACHED FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10531002 ↗
  • MalfunctionSep 25, 2019Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. IT IS UNKNOWN IF ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9119107 ↗
  • MalfunctionSep 25, 2019Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. IT IS UNKNOWN IF ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9119106 ↗
  • MalfunctionSep 25, 2019Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. IT IS UNKNOWN IF ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9119100 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K022269. One submission often covers many models, so check the product and lot details in each record.

  • Z-2144-2024 · May 10, 2024 · Open, Classified
    VICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in cardiovascular and
  • Z-0989-2020 · Jan 3, 2020 · Terminated
    Coated VICRYL (Polyglactin 910) Suture, 18" (45cm) Product Code J497H. The device is a synthetic absorbable sterile surgical suture.
  • Z-1087-06 · May 4, 2006 · Terminated
    Coated Vicryl (polyglactin 910) suture. Needle Type CTXB Suture size: 1. Vicryl suture is a synthetic absorbable sterile surgical suture comprised of a copolymer made from 90% gyc

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page