Coated VICRYL J555GRecalls & reported problems

Ethicon, LLC · Violet Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant

All models of Coated VICRYL (350)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Coated VICRYL J555G is a violet braided absorbable surgical suture used to close wounds. The 38 reports mostly describe the thread fraying, breaking or separating during surgery, and needles pulling off the suture mid-procedure. Reports also mention lumps and inflammation at the wound site after surgery. There are 18 injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
18
Recalls naming it
0

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 10705031041219), received between Feb 23, 2021 and Sep 25, 2024 — classified as 18 injuries, 20 malfunctions.

The most frequently reported specific device problem is “Material Frayed” (19 reports).

3 other recalls cover devices cleared under the same FDA submission (K022269); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 12 reportsFeb 2024: 4 reportsMar 2024: 5 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 5 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Frayed19
  • Adverse Event Without Identified Device or Use Problem18
  • Defective Component12
  • Break8
  • Material Separation5

Patient problems

  • No Clinical Signs, Symptoms or Conditions20
  • Granuloma14
  • Inflammation6
  • Hypersensitivity/Allergic reaction2

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE PULLED OFF THE SUTURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20306879 ↗
  • MalfunctionSep 25, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE PULLED OFF THE SUTURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20306875 ↗
  • MalfunctionSep 25, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE PULLED OFF THE SUTURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20306874 ↗
  • MalfunctionSep 25, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE PULLED OFF THE SUTURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20306872 ↗
  • MalfunctionSep 25, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE PULLED OFF THE SUTURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20306871 ↗
  • InjuryMar 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN EYE SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED AN UNCOMFORTABLE REACTION AFTER USAGE OF THE SUTURE THAT OCCURRED AROUND THE WOUND CLOSURE AREA. THE PATIENT WAS REFERRED TO A CORNEA SPECIALIST AND TOOK STEROIDS TO EASE INFLAMMATION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hypersensitivity/Allergic reaction; Inflammation

    MAUDE report 18935414 ↗
  • InjuryMar 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN EYE SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED AN UNCOMFORTABLE REACTION AFTER USAGE OF THE SUTURE THAT OCCURRED AROUND THE WOUND CLOSURE AREA. THE PATIENT WAS REFERRED TO A CORNEA SPECIALIST AND TOOK STEROIDS TO EASE INFLAMMATION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hypersensitivity/Allergic reaction; Inflammation

    MAUDE report 18935412 ↗
  • MalfunctionMar 5, 2024Material Frayed; Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS FOUND TO BE FRAYING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED BUT UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18841652 ↗
  • MalfunctionMar 1, 2024Material Frayed; Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING SURGERY, SUTURE WAS FOUND TO BE FRAYING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED BUT UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18819055 ↗
  • MalfunctionMar 1, 2024Material Frayed; Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING SURGERY, SUTURE WAS FOUND TO BE FRAYING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED BUT UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18819001 ↗
  • InjuryFeb 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGICAL PROCEDURE ON THE EYE AND SUTURE WAS USED. POST-OP, THE PATIENT PRESENTED WITH INFLAMMATION. THE SURGEON PRESCRIBED STEROIDS TO CONTROL THE INFLAMMATION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation

    MAUDE report 18714809 ↗
  • InjuryFeb 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGICAL PROCEDURE ON THE EYE AND SUTURE WAS USED. POST-OP, THE PATIENT PRESENTED WITH INFLAMMATION. THE SURGEON PRESCRIBED STEROIDS TO CONTROL THE INFLAMMATION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation

    MAUDE report 18714801 ↗
  • InjuryFeb 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGICAL PROCEDURE ON THE EYE AND SUTURE WAS USED. POST-OP, THE PATIENT PRESENTED WITH INFLAMMATION. THE SURGEON PRESCRIBED STEROIDS TO CONTROL THE INFLAMMATION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma; Inflammation

    MAUDE report 18714741 ↗
  • InjuryFeb 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGICAL PROCEDURE ON THE EYE AND SUTURE WAS USED. POST-OP, THE PATIENT PRESENTED WITH INFLAMMATION. THE SURGEON PRESCRIBED STEROIDS TO CONTROL THE INFLAMMATION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma; Inflammation

    MAUDE report 18714700 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma

    MAUDE report 18548007 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma

    MAUDE report 18548006 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma

    MAUDE report 18548005 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma

    MAUDE report 18548001 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma

    MAUDE report 18547999 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma

    MAUDE report 18547998 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma

    MAUDE report 18547991 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma

    MAUDE report 18547989 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma

    MAUDE report 18547973 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE.

    Patient problems: Granuloma

    MAUDE report 18547971 ↗
  • InjuryJan 19, 2024Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem; Material Frayed; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STRABISMUS SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE CLINICIAN OPINES THAT IN THE LAST 3 YEARS HE HAS EXPERIENCED A PROGRESSIVE DETERIORATION IN THE QUALITY OF SUTURE THREAD. THE PATIENT EXPERIENCED FORMATION OF GRANULOMAS RELATED TO CONJUNCTIVAL FLAPS. THE SURGEON ALSO STATED THAT WHEN THE THREAD GETS WET IT LOSES ITS HOMOGENEITY OF WEAVE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Granuloma

    MAUDE report 18547967 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K022269. One submission often covers many models, so check the product and lot details in each record.

  • Z-2144-2024 · May 10, 2024 · Open, Classified
    VICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in cardiovascular and
  • Z-0989-2020 · Jan 3, 2020 · Terminated
    Coated VICRYL (Polyglactin 910) Suture, 18" (45cm) Product Code J497H. The device is a synthetic absorbable sterile surgical suture.
  • Z-1087-06 · May 4, 2006 · Terminated
    Coated Vicryl (polyglactin 910) suture. Needle Type CTXB Suture size: 1. Vicryl suture is a synthetic absorbable sterile surgical suture comprised of a copolymer made from 90% gyc

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices