Coated VICRYL J602HRecalls & reported problems
Ethicon, LLC · Violet Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (350)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Coated VICRYL J602H suture from ETHICON is a violet braided absorbable surgical stitch used to close tissue. The 76 reports mostly describe the thread separating or fraying, or the attached needle breaking during surgery, once leaving a needle fragment that had to be retrieved, plus two injuries. No deaths. Most reports mention no lasting effect on the patient, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 76
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 76 adverse event reports that list this exact device (DI 10705031041615), received between Jan 10, 2018 and Feb 18, 2026 — classified as 2 injuries, 74 malfunctions.
The most frequently reported specific device problem is “Material Separation” (36 reports).
3 other recalls cover devices cleared under the same FDA submission (K022269); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation36
- Break29
- Material Frayed4
- Adverse Event Without Identified Device or Use Problem2
- Device Contaminated During Manufacture or Shipping2
- Appropriate Device Problem Term/Code Not Available1
- Defective Component1
- Device Markings/Labelling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions58
- No Consequences Or Impact To Patient16
- Abdominal Pain1
- Hernia1
- Not Applicable1
- Obstruction/Occlusion1
- Wound Dehiscence1
Recent adverse event reports
The 25 most recent of 76 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 18, 2026Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE SUTURE WAS FRAYED DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24382871 ↗MalfunctionFeb 18, 2026Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE SUTURE WAS FRAYED DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24382855 ↗MalfunctionFeb 2, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A COLECTOMY PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING LAPAROSCOPIC COLECTOMY, WHEN THE DOCTOR PASSED THROUGH THE NEEDLE INTO SKIN AND ATTEMPTED TO LIFT IT, THE SWAGE PART OF THE NEEDLE BROKE. THE SAME AMOUNT OF FORCE AS USUAL WAS APPLIED, BUT THE NEEDLE BROKE. HOWEVER, THE NEEDLE HAD BEEN HELD BY A NURSE. THE PRODUCT WAS USED ON THE SKIN. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24236954 ↗MalfunctionDec 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING SURGERY, THERE HAD BEEN A MIDDLE PART OF NEEDLE BROKEN AND A NEEDLE FELL. NO PIECES WERE LEFT INSIDE THE PATIENT, BUT THE DETAILS ARE UNKNOWN AT THIS TIME, SO IT IS BEING CONFIRMED. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23708283 ↗MalfunctionAug 25, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS FOUND THAT THERE WAS A PIECE OF CARDBOARD MIXED IN WITH THE TRANSPARENT PACKAGE, AND THE CUSTOMER WOULD LIKE TO REFRAIN FROM SHIPPING IT. THIS EVENT WAS FOUND BEFORE USE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22888804 ↗MalfunctionJul 28, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. NEEDLE BREAKAGE OCCURRED DURING THE OPERATION. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22637629 ↗MalfunctionMay 13, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING A LAPAROSCOPIC SURGERY, THE NEEDLE AT THE SWAGE PART WAS BROKEN DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THE TIP PART ALSO HAS BEEN RETRIEVED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22010920 ↗MalfunctionJan 17, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING A LAPAROSCOPIC LEFT OVARIAN TUMOR RESECTION, DURING WOUND CLOSURE, THE NEEDLE WAS BROKEN AT THE MIDDLE PART WHEN IT PASSED THROUGH THE TISSUE AND SUTURE THE FASCIA AT THE PORT (12MM) AS USUAL. THE DOCTOR THOUGHT THE TIP OF NEEDLE AT THE MIDDLE, NOT THE SWAGE PART (THE ONE WITH THE NEEDLE) WENT TO THE PATIENT¿S BODY BUT THE NEEDLE WAS FOUND FROM DEEP INSIDE THE FASCIA. THE PRODUCT WAS USED ON THE FASCIA. IT WAS NOT A CONTROL RELEASE NEEDLE. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21184008 ↗MalfunctionSep 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN SURGERY, IT WAS NOT A CONTROL RELEASE NEEDLE. THE NEEDLE TIP WAS BROKEN. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20268482 ↗MalfunctionSep 10, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN APPENDECTOMY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. ON (B)(6) 2024, AT APPROXIMATELY 12:24 PM, THE PHARMACY SATELLITE TEAM AT OT REPORTED AN INCIDENT WHERE THE SUTURE DETACHED FROM THE NEEDLE BODY WHEN ATTEMPTING TO USE THE SUTURE DURING A VLA OPERATION BY THE OPERATOR. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20181113 ↗MalfunctionJul 3, 2024Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE WAS BROKEN. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19664713 ↗MalfunctionJun 13, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC APPENDECTOMY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, FOREIGN MATTER LIKE COTTON FOUND ON THE NEEDLE TIP. THE NEW PRODUCT WAS USED. THE PRODUCT WAS USED AT THE PLACE TO STAND THE PORT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19532380 ↗MalfunctionFeb 22, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING AN UNKNOWN RESPIRATORY SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE DURING USE. THE SURGEON STOPPED USING THE PRODUCT. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18759912 ↗MalfunctionFeb 2, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN THE SURGEON WAS ABOUT TO USE THE PRODUCT, THE SUTURE WAS DETACHED FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18636591 ↗MalfunctionOct 25, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN SURGERY, THE NEEDLE WAS DETACHED FROM THE SUTURE DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18004759 ↗MalfunctionOct 25, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18004245 ↗MalfunctionSep 19, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN OB-GYN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE SUTURE WAS DETACHED FROM THE NEEDLE WHEN THE SURGEON APPLIED THE FIRST STITCH ON TISSUE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17774541 ↗MalfunctionJul 24, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN DURING USE MANY TIMES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17382071 ↗MalfunctionJul 24, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON 04/07/2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN DURING USE MANY TIMES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17382069 ↗MalfunctionJul 24, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN DURING USE MANY TIMES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17382063 ↗MalfunctionJul 24, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN DURING USE MANY TIMES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17382058 ↗MalfunctionJul 24, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN DURING USE MANY TIMES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17382048 ↗MalfunctionJun 22, 2023Defective Component; Defective Component
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, AFTER OPENING THE PACKAGE AND BEFORE USE, IT WAS FOUND THAT THE SUTURE WAS PARTIALLY THICKENED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17181520 ↗MalfunctionJun 14, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN OB-GYN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE OB-GYN SURGERY, WHEN THE SURGEON APPLIED THE PRODUCT ON FASCIAL TISSUE, THE NEEDLE WAS DETACHED FROM THE SUTURE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17126071 ↗MalfunctionMay 10, 2023Material Frayed; Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, BEFORE USE THE SURGEON HELD THE PRODUCT WITH FORCEPS, HE REALIZED THAT THE SUTURE WAS UNTIED. ON(B)(4) 2023 ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT THE SUTURE WAS FRAYED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16908596 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K022269. One submission often covers many models, so check the product and lot details in each record.
- Z-2144-2024 · May 10, 2024 · Open, ClassifiedVICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in cardiovascular and
- Z-0989-2020 · Jan 3, 2020 · TerminatedCoated VICRYL (Polyglactin 910) Suture, 18" (45cm) Product Code J497H. The device is a synthetic absorbable sterile surgical suture.
- Z-1087-06 · May 4, 2006 · TerminatedCoated Vicryl (polyglactin 910) suture. Needle Type CTXB Suture size: 1. Vicryl suture is a synthetic absorbable sterile surgical suture comprised of a copolymer made from 90% gyc
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page