Coated VICRYL J718TRecalls & reported problems
Ethicon, LLC · Violet Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (350)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705031042520), received between Mar 15, 2022 and Mar 15, 2022 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Material Separation” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Separation12
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) "222" AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770894 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) "222" AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770880 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) "222" AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770874 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770872 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770871 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770867 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770859 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770854 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770843 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770828 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770807 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) "222" AND SUTURE WAS USED. DURING THE PROCEDURE, MULTIPLE STRANDS OF SUTURE THE NEEDLE POPPED OFF PREMATURELY. NEEDLE WOULD BARELY MAKE IT THROUGH THE TISSUE BEFORE THE NEEDLE WOULD POP OFF. MULTIPLE LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE AS THE PREMATURE POP OFF CONTINUED TO HAPPEN THROUGHOUT THE CASE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770802 ↗
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page