PROLENE M8703Recalls & reported problems
Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant
Quick summary
This Ethicon Prolene blue polypropylene surgical suture is used to close tissue during operations. Most of the 70 reports describe the suture strand snapping or separating, often during coronary artery bypass procedures, and sometimes delaying the operation; no deaths and no injuries were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 70
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 10705031045781), received between Apr 3, 2019 and Sep 29, 2026 — classified as 70 malfunctions.
The most frequently reported specific device problem is “Material Separation” (35 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation35
- Break33
- Appropriate Device Problem Term/Code Not Available1
- Component Misassembled1
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions64
- No Consequences Or Impact To Patient6
Recent adverse event reports
The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934185 ↗MalfunctionMay 29, 2026Break; Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFT PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP THAT DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE SUTURE BROKE TWICE DURING SUTURING. NO PATIENT CONSEQUENCES WERE REPORTED. THE DEVICE IS NOT BEING RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25330760 ↗MalfunctionMar 2, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE NEEDLE PULLED OFF. DURING THE PROCEDURE, THE SURGEON TRIED TO TIE THE SUTURE AFTER PASSAGE AND SIX DIFFERENT STRANDS BROKE DURING TYING. IT DELAYED THE PROCEDURE AROUND 15-20 MINUTES AND AFFECTED THE QUALITY OF THE ANASTOMOSIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24490378 ↗MalfunctionMar 2, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE NEEDLE PULLED OFF. DURING THE PROCEDURE, THE SURGEON TRIED TO TIE THE SUTURE AFTER PASSAGE AND SIX DIFFERENT STRANDS BROKE DURING TYING. IT DELAYED THE PROCEDURE AROUND 15-20 MINUTES AND AFFECTED THE QUALITY OF THE ANASTOMOSIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24490305 ↗MalfunctionMar 2, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE NEEDLE PULLED OFF. DURING THE PROCEDURE, THE SURGEON TRIED TO TIE THE SUTURE AFTER PASSAGE AND SIX DIFFERENT STRANDS BROKE DURING TYING. IT DELAYED THE PROCEDURE AROUND 15-20 MINUTES AND AFFECTED THE QUALITY OF THE ANASTOMOSIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24490275 ↗MalfunctionMar 2, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE NEEDLE PULLED OFF. DURING THE PROCEDURE, THE SURGEON TRIED TO TIE THE SUTURE AFTER PASSAGE AND SIX DIFFERENT STRANDS BROKE DURING TYING. IT DELAYED THE PROCEDURE AROUND 15-20 MINUTES AND AFFECTED THE QUALITY OF THE ANASTOMOSIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24490202 ↗MalfunctionMar 2, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE NEEDLE PULLED OFF. DURING THE PROCEDURE, THE SURGEON TRIED TO TIE THE SUTURE AFTER PASSAGE AND SIX DIFFERENT STRANDS BROKE DURING TYING. IT DELAYED THE PROCEDURE AROUND 15-20 MINUTES AND AFFECTED THE QUALITY OF THE ANASTOMOSIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24490192 ↗MalfunctionMar 2, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE NEEDLE PULLED OFF. DURING THE PROCEDURE, THE SURGEON TRIED TO TIE THE SUTURE AFTER PASSAGE AND SIX DIFFERENT STRANDS BROKE DURING TYING. IT DELAYED THE PROCEDURE AROUND 15-20 MINUTES AND AFFECTED THE QUALITY OF THE ANASTOMOSIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24490176 ↗MalfunctionDec 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED TO THE DL BY THE SALES REP THAT THE 7-0 PROLINE BV-1 MULTIPASS NEEDLE THAT SUTURES KEPT BREAKING DURING SURGERY. THE 5-0 PROLINE C-1 NEEDLE ALSO CONTINUED TO BREAK. THERE WAS NO ADVERSE IMPACT ON PATIENT/USER REPORTED. DEVICES ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23704725 ↗MalfunctionDec 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED TO THE DL BY THE SALES REP THAT THE 7-0 PROLINE BV-1 MULTIPASS NEEDLE THAT SUTURES KEPT BREAKING DURING SURGERY. THE 5-0 PROLINE C-1 NEEDLE ALSO CONTINUED TO BREAK. THERE WAS NO ADVERSE IMPACT ON PATIENT/USER REPORTED. DEVICES ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23704716 ↗MalfunctionDec 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED TO THE DL BY THE SALES REP THAT THE 7-0 PROLINE BV-1 MULTIPASS NEEDLE THAT SUTURES KEPT BREAKING DURING SURGERY. THE 5-0 PROLINE C-1 NEEDLE ALSO CONTINUED TO BREAK. THERE WAS NO ADVERSE IMPACT ON PATIENT/USER REPORTED. DEVICES ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23704711 ↗MalfunctionDec 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED TO THE DL BY THE SALES REP THAT THE 7-0 PROLINE BV-1 MULTIPASS NEEDLE THAT SUTURES KEPT BREAKING DURING SURGERY. THE 5-0 PROLINE C-1 NEEDLE ALSO CONTINUED TO BREAK. THERE WAS NO ADVERSE IMPACT ON PATIENT/USER REPORTED. DEVICES ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23704688 ↗MalfunctionDec 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED TO THE DL BY THE SALES REP THAT THE 7-0 PROLINE BV-1 MULTIPASS NEEDLE THAT SUTURES KEPT BREAKING DURING SURGERY. THE 5-0 PROLINE C-1 NEEDLE ALSO CONTINUED TO BREAK. THERE WAS NO ADVERSE IMPACT ON PATIENT/USER REPORTED. DEVICES ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23704684 ↗MalfunctionOct 29, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED TO THE DL BY THE SALES REP THAT THE 7-0 PROLINE BV-1 MULTIPASS NEEDLE THAT SUTURES KEPT BREAKING DURING SURGERY. THE 5-0 PROLINE C-1 NEEDLE ALSO CONTINUED TO BREAK. THERE WAS NO ADVERSE IMPACT ON PATIENT/USER REPORTED. DEVICES ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23413890 ↗MalfunctionJul 8, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SURGEON STATED THAT THE NEEDLE WILL FELL OUT THE SWAGE WITHOUT FORCE. ADVERSE IMPACT PATIENT/USER: NO. DEVICE IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22437269 ↗MalfunctionMay 6, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DL BY SALES REP THAT NEEDLE FALLS OFF OF THE SUTURE. NO ADVERSE IMPACT PATIENT/USER WAS REPORTED. DEVICE IS AVAILABLE FOR RETURN. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21971006 ↗MalfunctionApr 28, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SURGEON WHO STATED HE HAD A 7-0 DISINTEGRATE ON A DISTAL. HE HAD A COMPLAINT ON A 8-0 MONTHS AGO. HE¿S SINCE BEEN USING 7-0 AND HE NOTIFIED ME YESTERDAY OF HIS SIMILAR EXPERIENCE WITH THE 7-0. HIS COLLEAGUES WERE NOT USING THE 8-0 AT THE TIME. THEY ARE CURRENTLY USING 7-0. I HAVE NOT HAD ANY FEEDBACK FROM HIS COLLEAGUES. THERE WERE NO A PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21918569 ↗MalfunctionApr 28, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SURGEON WHO STATED HE HAD A 7-0 DISINTEGRATE ON A DISTAL. SURGEON IS WILLING TO HAVE A RAPID RESPONSE CALL TO DISCUSS. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE COMPLAINT IS INTRA OP. THE SURGEON HAS STATED PROLENE IS DEGRADING AS HE IS TYING HIS KNOTS ON HIS DISTAL ANASTOMOSIS. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21918488 ↗MalfunctionJan 7, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SURGEONS SAID THE SUTURES WERE BREAKING FAR TOO EASILY. THEY SWITCHED TO SINGLE SUTURES AND THE SUTURES DID NOT CONTINUE TO BREAK. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21095271 ↗MalfunctionJan 7, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE 2ND SURGEON HAD ISSUES WITH THE MULTIPACK OF THE SUTURE BREAKING, SWITCHED TO INDIVIDUAL SUTURE AND HAD NO MORE ISSUES W/ BREAKING, THEY PULLED 3 BOXES OF THIS LOT THEY HAD, DUE TO THE COMPLAINTS SURGEONS WON'T USE THEM FOR FEAR OF BREAKING AND CAUSING CONSEQUENCES TO THE PATIENT. THERE WERE NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21095232 ↗MalfunctionJan 6, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, SURGEONS AND NURSING TEAM REQUESTED TO PULL THE REMAINING LOTS OF 102TPM - 7-0 M8703 MULTIPACK SUTURES DUE TO 2 PRIOR INCIDENTS OF SUTURES BREAKING ON THEM. (CARDIAC SURGEONS) THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21084802 ↗MalfunctionJan 6, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, SURGEONS AND NURSING TEAM REQUESTED TO PULL THE REMAINING LOTS OF 102TPM - 7-0 M8703 MULTIPACK SUTURES DUE TO 2 PRIOR INCIDENTS OF SUTURES BREAKING ON THEM. (CARDIAC SURGEONS) THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21084792 ↗MalfunctionJan 6, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, SURGEONS AND NURSING TEAM REQUESTED TO PULL THE REMAINING LOTS OF 102TPM - 7-0 M8703 MULTIPACK SUTURES DUE TO 2 PRIOR INCIDENTS OF SUTURES BREAKING ON THEM. (CARDIAC SURGEONS) THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21084752 ↗MalfunctionNov 27, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. WHILE SOWING DISTAL ANASTOMOSIS, THE NEEDLE KEPT POPPING OFF THE SUTURE. THEY USED A NEW SUTURE FROM A DIFFERENT LOT NUMBER TO COMPLETE THE CASE WITHOUT PATIENT HARM. THE NEEDLE DID NOT FALL INTO THE PATIENT. THE NEEDLE STAYED IN THE NEEDLE HOLDER. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20792635 ↗MalfunctionNov 27, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. WHILE SOWING DISTAL ANASTOMOSIS, THE NEEDLE KEPT POPPING OFF THE SUTURE. THEY USED A NEW SUTURE FROM A DIFFERENT LOT NUMBER TO COMPLETE THE CASE WITHOUT PATIENT HARM. THE NEEDLE DID NOT FALL INTO THE PATIENT. THE NEEDLE STAYED IN THE NEEDLE HOLDER. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20791781 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2455-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
- Z-2454-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2462-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
- Z-2457-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
- Z-2452-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2463-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
- Z-2456-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
- Z-2449-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
- Z-2448-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.