PROLENE M8726Recalls & reported problems
Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 10705031045927), received between May 15, 2019 and Aug 13, 2026 — classified as 22 malfunctions.
The most frequently reported specific device problem is “Material Separation” (18 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation18
- Break4
- Material Frayed2
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- No Consequences Or Impact To Patient8
Recent adverse event reports
The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50237743 ↗MalfunctionAug 13, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50237742 ↗MalfunctionAug 13, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50237740 ↗MalfunctionMar 26, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A PEDIATRIC CARDIAC PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE SALES REP THAT DURING PEDIATRIC CARDIAC SURGERY THE NEEDLE WAS DETACHING FROM THE SUTURE. THERE WERE NO PATIENT ADVERSE EVENT. PRODUCT IS AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24701790 ↗MalfunctionMar 26, 2026Material Separation
IT WAS REPORTED TO THE SALES REP THAT DURING PEDIATRIC CARDIAC SURGERY THE NEEDLE WAS DETACHING FROM THE SUTURE. THERE WERE NO PATIENT ADVERSE EVENT. PRODUCT IS AVAILABLE FOR RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24701789 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE WHEN SUTURING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22029264 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE WHEN SUTURING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22029263 ↗MalfunctionApr 23, 2024Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT A CABG (CORONARY ARTERY BYPASS GRAFT) PROCEDURE ON AN UNKNOWN DATE IN (B)(6) 2024 AND SUTURE WAS USED. EVERY TIME THEY THREW THE NEEDLE, THE NEEDLE WAS POPPING OFF THE SUTURE. NO PATIENT HARM WAS REPORTED, AND THE CASE WAS COMPLETED WITH A SIMILAR DEVICE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19162456 ↗MalfunctionApr 23, 2024Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT A CABG (CORONARY ARTERY BYPASS GRAFT) PROCEDURE ON AN UNKNOWN DATE IN (B)(6) 2024 AND SUTURE WAS USED. EVERY TIME THEY THREW THE NEEDLE, THE NEEDLE WAS POPPING OFF THE SUTURE. NO PATIENT HARM WAS REPORTED, AND THE CASE WAS COMPLETED WITH A SIMILAR DEVICE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19162455 ↗MalfunctionFeb 15, 2023Break; Material Frayed; Break; Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. THE DOCTOR WENT TO USE THE SUTURE AND NOTED THAT THE 6-0 SUTURE WAS SPLIT AND UNRAVELING. THIS STITCH WAS REMOVED AND A NEW LIKE SUTURE WAS USED TO CONTINUE SEWING THE SAPHENOUS VEIN ON THE HEART. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16379738 ↗MalfunctionDec 29, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. THE SWEDGE WAS DETACHING FROM TWO SUTURES. A NEW SUTURE WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16075425 ↗MalfunctionMar 9, 2022Material Separation; Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A PEDIATRIC CARDIOVASCULAR PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING PROCEDURE THE SUTURES WERE 'FALLING OFF.' NO OTHER INFORMATION IS AVAILABLE AT THIS TIME. NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13714445 ↗MalfunctionDec 28, 2020Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN HEART PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF TOO EASILY. ANOTHER LIKE DEVICE OF A SEPARATE LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11076344 ↗MalfunctionDec 28, 2020Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN HEART PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF TOO EASILY. ANOTHER LIKE DEVICE OF A SEPARATE LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11076341 ↗MalfunctionSep 25, 2020Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARB SURGERY ON (B)(6) 2020 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF OF THE SUTURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10587078 ↗MalfunctionDec 12, 2019Break; Material Frayed; Break; Material Frayed
IT WAS REPORTED THAT THE PATIENT UNDERWENT CABG PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED TO SUTURE ANASTOMOSIS. DURING THE PROCEDURE, THE SUTURE MAY HAVE BROKEN AND SUTURE MAY HAVE UNRAVELED. A SUTURE THAT UNRAVELED WAS AROUND THE AORTA AND THE SURGEON SPOTTED THIS BEFORE CLOSING. THE PROCEDURE WAS COMPLETED WITH NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9461177 ↗MalfunctionJun 17, 2019Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT A VASCULAR OR HEART PROCEDURE ON UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE IS POPPING OFF VERY FAST. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE AND DIFFERENT LOTS. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8705864 ↗MalfunctionJun 17, 2019Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT A VASCULAR OR HEART PROCEDURE ON UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE IS POPPING OFF VERY FAST. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE AND DIFFERENT LOTS. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8705862 ↗MalfunctionJun 17, 2019Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT A VASCULAR OR HEART PROCEDURE ON UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE IS POPPING OFF VERY FAST. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE AND DIFFERENT LOTS. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8705861 ↗MalfunctionJun 17, 2019Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT A VASCULAR OR HEART PROCEDURE ON UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE IS POPPING OFF VERY FAST. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE AND DIFFERENT LOTS. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8705858 ↗MalfunctionJun 17, 2019Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT A VASCULAR OR HEART PROCEDURE ON UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE IS POPPING OFF VERY FAST. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE AND DIFFERENT LOTS. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8705857 ↗MalfunctionMay 15, 2019Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A HEART CASE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE NEEDLE UNEXPECTEDLY POPPED OFF FROM THE SUTURE. THE CASE WAS COMPLETED WITH ANOTHER SUTURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8614324 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2454-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
- Z-2461-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
- Z-2459-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
- Z-2450-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2452-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
- Z-2460-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
- Z-2448-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
- Z-2449-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2463-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.