Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705031045958), received between May 26, 2020 and Apr 2, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Material Separation3
Patient problems
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 2, 2024Material Separation; Material Separation
IT WAS REPORTED THAT PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFT (CABG) PROCEDURE ON AND UNKNOWN DATE IN 2024, AND SUTURE WAS USED ON THE VESSELS. THE NEEDLE ON THE 7-0 PROLENE POPPED OFF DURING VESSEL ANASTOMOSIS. A NEW PACKET OF THE SAME SUTURE FROM THE SAME BOX WAS OPENED AND IT OCCURRED AGAIN DURING THE ANASTOMOSIS. THE HCP OPENED ANOTHER BOX OF THE SAME LOT NUMBER AND EXPERIENCED THE SAME THING OFF OF THE SURGICAL FIELD WHEN TESTED. THERE WAS NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19019634 ↗MalfunctionApr 2, 2024Material Separation; Material Separation
IT WAS REPORTED THAT PATIENT UNDERWENT A CORONARY ARTERY BYPASS GRAFT (CABG) PROCEDURE ON AND UNKNOWN DATE IN 2024, AND SUTURE WAS USED ON THE VESSELS. THE NEEDLE ON THE 7-0 PROLENE POPPED OFF DURING VESSEL ANASTOMOSIS. A NEW PACKET OF THE SAME SUTURE FROM THE SAME BOX WAS OPENED AND IT OCCURRED AGAIN DURING THE ANASTOMOSIS. THE HCP OPENED ANOTHER BOX OF THE SAME LOT NUMBER AND EXPERIENCED THE SAME THING OFF OF THE SURGICAL FIELD WHEN TESTED. THERE WAS NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19019631 ↗MalfunctionNov 2, 2020Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON (B)(6) 2020; AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10773805 ↗MalfunctionMay 26, 2020Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A THORACOTOMY ON (B)(6) 2020 AND SUTURE WAS USED. DURING THE PROCEDURE, THE DEVICE KEPT BREAKING. SWITCHED TO ANOTHER LOT TO COMPLETE THE CASE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10086570 ↗MalfunctionMay 26, 2020Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A THORACOTOMY ON (B)(6) 2020 AND SUTURE WAS USED. DURING THE PROCEDURE, THE DEVICE KEPT BREAKING. SWITCHED TO ANOTHER LOT TO COMPLETE THE CASE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10086569 ↗MalfunctionMay 26, 2020Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A THORACOTOMY ON (B)(6) 2020 AND SUTURE WAS USED. DURING THE PROCEDURE, THE DEVICE KEPT BREAKING. SWITCHED TO ANOTHER LOT TO COMPLETE THE CASE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10086568 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2454-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2449-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
- Z-2457-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2461-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
- Z-2460-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
- Z-2456-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
- Z-2448-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
- Z-2452-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
- Z-2462-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page