MONOCRYL MCP490GRecalls & reported problems

Ethicon, LLC · Antibacterial Undyed Monofilament (Poliglecaprone 25), Absorbable Surgical Suture · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10705031046542), received between May 1, 2019 and May 6, 2026 — classified as 3 injuries, 11 malfunctions.

The most frequently reported specific device problem is “Material Separation” (8 reports).

2 other recalls cover devices cleared under the same FDA submission (K050845, K201996); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation8
  • Break3
  • Material Integrity Problem3
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Discomfort3
  • Skin Inflammation/ Irritation3
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN FACE PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE OF SUTURING A WOUND ON THE FACE, WHILE THE DOCTOR WAS DOING THE CLOSING OF THE SKIN, THE NEEDLE WAS DETACHED FROM THE THREAD OF THE SUTURE AS IT PASSED THE POINT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25094459 ↗
  • MalfunctionApr 17, 2026Material Frayed; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE SUTURE PEELED WHEN THE NEEDLE PASSED, CAUSING THE THREAD TO LOOSE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24913764 ↗
  • MalfunctionApr 17, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE SUTURE HEAD BROKE AS THE NEEDLE PASSED THROUGH, CAUSING THE THREAD TO COME LOOSE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912699 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. CASE LOGGED IN RELATION TO THE SECOND INCIDENT MENTIONED IN (B)(4). SHE COMMENTED THAT THIS HAD ALSO HAPPENED SOME MONTHS PRIOR, ALBEIT ON A 3/0 MONOCRYL. SHE INFORMED ME THAT THE NEEDLE TIP WAS NOT RECOVERED AND THAT THE PATIENT HAD BEEN SENT FOR A CT SCAN - WHICH SUGGESTED THAT THE NEEDLE TIP WAS IN-SITU. I ASKED HER TO RETAIN THE SUTURE AND NEEDLE. HOWEVER, THESE HAD ALREADY BEEN DISPOSED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24714901 ↗
  • MalfunctionJan 22, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, SUTURE THREADS WERE BREAKING CONSISTENTLY. NO PATIENT CONSEQUENCES REPORTED WHEN THIS COMPLAINT EVENT OCCURRED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24146673 ↗
  • MalfunctionMar 25, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN APPENDECTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT IS REPORTED THAT WHEN THE PATIENT IS SUTURED, THE THREAD IS RELEASED, CAUSING THE NEEDLE TO UNTIE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21691294 ↗
  • InjuryOct 17, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A DE QUERVAIN'S SYNDROME RELEASE PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. POST-OP, THE PATIENT HAD DELAYED ABSORPTION AND DID NOT ABSORB WELL. THE PATIENT EXPERIENCED IRRITATION AT THE SUTURE SITE AND FELT THAT THERE WERE SUTURE REMNANTS UNDER HER SKIN EVEN AFTER 3 MONTHS POST-OP. ON (B)(6) 2024, THE DOCTOR REVISED THE SCAR WITH A DIFFERENT SUTURE AND REMOVED ALL THE REMNANT SUTURES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Discomfort; Skin Inflammation/ Irritation

    MAUDE report 20472360 ↗
  • InjuryOct 17, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A DE QUERVAIN'S SYNDROME RELEASE PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. POST-OP, THE PATIENT HAD DELAYED ABSORPTION AND DID NOT ABSORB WELL. THE PATIENT EXPERIENCED IRRITATION AT THE SUTURE SITE. THE SKIN WAS RAISED. ABOUT 1 MONTH AFTER THE OPERATION ON (B)(6) 2024, THE DOCTOR REMOVED THE SUTURE, PROVIDED WOUND CARE ADVICE, AND PUT A DRESSING ON. THE PATIENT IS CURRENTLY DOING FINE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Discomfort; Skin Inflammation/ Irritation

    MAUDE report 20472339 ↗
  • InjuryOct 17, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A DE QUERVAIN'S SYNDROME RELEASE PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. POST-OP, THE PATIENT HAD DELAYED ABSORPTION AND DID NOT ABSORB WELL. THE PATIENT EXPERIENCED IRRITATION AT THE SUTURE SITE. THE SKIN WAS RAISED. AFTER POST-OP 3 WEEKS, THE PATIENT HAD IRRITATION AT SUTURE SITE. IT WAS STATED THAT THE REACTION WAS NOTED ON (B)(6) 2024. THE DOCTOR REMOVED THE SUTURE, PROVIDED WOUND CARE ADVICE, AND PUT A DRESSING ON. THE PATIENT IS CURRENTLY DOING FINE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Discomfort; Skin Inflammation/ Irritation

    MAUDE report 20472297 ↗
  • MalfunctionAug 13, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2024 AND SUTURE WAS USED. THE SECOND SUTURE IS PASSED TO THE TABLE, THE NEEDLE BECOMES DETACHED FROM THE THREAD WHEN PASSING THROUGH THE TISSUE FOR THE FIRST TIME. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19977602 ↗
  • MalfunctionFeb 22, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE'S NEEDLE AND LINE IS SEPARATE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13579428 ↗
  • MalfunctionFeb 17, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE'S NEEDLE AND LINE IS SEPARATE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13554563 ↗
  • MalfunctionSep 11, 2020Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2020 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10521162 ↗
  • MalfunctionMay 1, 2019Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A GRAFT PROCEDURE IN THE RIGHT LEG ON AN UNKNOWN DATE IN 2019 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE STAYED IN THE NEEDLE HOLDER, AND THE SUTURE DISASSEMBLED FROM THE NEEDLE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8569145 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K050845, K201996. One submission often covers many models, so check the product and lot details in each record.

  • Z-2149-2024 · May 10, 2024 · Open, Classified
    MONOCRYL Plus Antibacterial (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated. PRODUCT CO
  • Z-1312-2009 · Feb 2, 2009 · Terminated
    MONOCRYL Plus Antibacterial (poliglecaprone 25) Suture; Ethicon, Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and/o

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →