NUROLON N104TRecalls & reported problems

Ethicon, LLC · Class 2 · Implant

All models of NUROLON (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705031047259), received between Sep 30, 2022 and Nov 27, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

9 other recalls cover devices cleared under the same FDA submission (K946173); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 27, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT VENTRICULO PERITONEAL SHUNT ON (B)(6) 2024 AND A SUTURE WAS USED. DURING THE INTERRUPTING SUTURE FOR CLOSING HEAD, THE SUTURE WAS USED FOR SUTURING THE GALEA, BUT THE NEEDLE WAS TOO LARGE TO FIT IN THE FINAL STITCH, SO ANOTHER SUTURE WAS USED. WHEN THE NEEDLE-SUTURE WAS APPLIED TO THE GALEA WHICH IS IN FRONT OF THE SURGEON AND THE OPPOSITE SIDE WAS SUTURED, THE SWAGE PART WAS GRASPED WITH A NEEDLE HOLDER AND THE NEEDLE WAS DEEPLY APPLIED TO THE GALEA AND IT WAS RELEASED, THE SWAGE PART WAS BROKEN AND FROM THE NEEDLE TIP TO THE BODY PART WAS TEMPORARILY BURIED IN THE TISSUE OF THE HEAD SKIN AND LOST SIGHT OF IT. THE SIDE OF THE SUTURE WAS REMOVED, BUT THE NEEDLE BODY COULD NOT BE FOUND WHERE IT HAD GONE, AND IT WAS SEARCHED FOR AND FOUND BY CT, SO IT REMOVED ALL OF THEM FROM THE PATIENT WITHOUT ADDITIONAL INCISION AS IT WAS. SINCE NO MISSING NEEDLE WAS FOUND, THE WOUND WAS CLOSED. THERE WAS NOTHING UNUSUAL AFTER THAT, AND THE BROKEN NEEDLE WAS REMOVED DURING THE SURGERY, SO THERE IS NO ADDITIONAL POSTOPERATIVE PROCEDURE OR RE-OPERATION. THE NURSE SAID THAT THE BROKEN NEEDLE¿S TOTAL LENGTH WAS CONFIRMED AND SHOULD NOT BE MISSING, BUT WANTED US

    Patient problems: Foreign Body In Patient

    MAUDE report 20791564 ↗
  • MalfunctionSep 30, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, POOR SUTURE QUALITY WHEN IT WAS TIED THE THREAD, NEEDLE BREAKS. BAD STRESS FACTOR. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15513931 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K946173. One submission often covers many models, so check the product and lot details in each record.

  • Z-0153-2027 · Sep 8, 2026 · Open, Classified
    PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H
  • Z-1516-2025 · Mar 7, 2025 · Open, Classified
    ETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
  • Z-1097-2025 · Dec 20, 2024 · Open, Classified
    PERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
  • Z-2486-2020 · Jun 10, 2020 · Open, Classified
    PERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
  • Z-1841-2019 · Apr 19, 2019 · Terminated
    ETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue
  • Z-1842-2019 · Apr 19, 2019 · Terminated
    ETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue
  • Z-0747-2019 · Dec 18, 2018 · Terminated
    PERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approx
  • Z-1839-2015 · May 8, 2015 · Terminated
    Ethicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
  • Z-0492-2015 · Oct 17, 2014 · Terminated
    Ethicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page