ETHICON TPW20Recalls & reported problems

Ethicon, LLC · Temporary Cardiac Pacing Wire · Class 2

Quick summary

This Ethicon temporary cardiac pacing wire is used to keep the heart beating at a set rate during surgery. Most of the 39 reports describe the wire's outer coating splitting apart or the wire breaking, sometimes mid-operation with the chest open; a few mention the product arriving unsterile. None of the reports describe a death or an injury. There have been 10 reports in the past 12 months, up from 2 in the 12 months before that. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
39
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 10705031050877), received between May 8, 2019 and Mar 30, 2026 — classified as 39 malfunctions.

The most frequently reported specific device problem is “Material Separation” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 7 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 8 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation26
  • Break11
  • Delivered as Unsterile Product2

Patient problems

  • No Clinical Signs, Symptoms or Conditions38
  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT THE PRODUCT WAS SPLITTING. CV PACK THE SUTURE BOOTS WERE BREAKING/SPLITTING IN HALF, UNSAFE WHILE WORKING IN AN OPEN CHEST CAVITY. PACING WIRES- WERE SPLITTING. THE SCRUB TECH REMOVED THE DEFECTIVE ITEMS, ENSURED THE PATIENT WASN'T HARMED, DISCUSSED WITH THE SURGEON AND ESCALATED TO TEAM LEADER. DEVICES ARE NOT AVAILABLE FOR RETURN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24729544 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT THE PRODUCT WAS SPLITTING. CV PACK THE SUTURE BOOTS WERE BREAKING/SPLITTING IN HALF, UNSAFE WHILE WORKING IN AN OPEN CHEST CAVITY. PACING WIRES- WERE SPLITTING. THE SCRUB TECH REMOVED THE DEFECTIVE ITEMS, ENSURED THE PATIENT WASN'T HARMED, DISCUSSED WITH THE SURGEON AND ESCALATED TO TEAM LEADER. DEVICES ARE NOT AVAILABLE FOR RETURN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24728036 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT THE PRODUCT WAS SPLITTING. CV PACK THE SUTURE BOOTS WERE BREAKING/SPLITTING IN HALF, UNSAFE WHILE WORKING IN AN OPEN CHEST CAVITY. PACING WIRES- WERE SPLITTING. THE SCRUB TECH REMOVED THE DEFECTIVE ITEMS, ENSURED THE PATIENT WASN'T HARMED, DISCUSSED WITH THE SURGEON AND ESCALATED TO TEAM LEADER. DEVICES ARE NOT AVAILABLE FOR RETURN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24727929 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT THE PRODUCT WAS SPLITTING. CV PACK THE SUTURE BOOTS WERE BREAKING/SPLITTING IN HALF, UNSAFE WHILE WORKING IN AN OPEN CHEST CAVITY. PACING WIRES- WERE SPLITTING. THE SCRUB TECH REMOVED THE DEFECTIVE ITEMS, ENSURED THE PATIENT WASN'T HARMED, DISCUSSED WITH THE SURGEON AND ESCALATED TO TEAM LEADER. DEVICES ARE NOT AVAILABLE FOR RETURN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24727294 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT THE PRODUCT WAS SPLITTING. CV PACK THE SUTURE BOOTS WERE BREAKING/SPLITTING IN HALF, UNSAFE WHILE WORKING IN AN OPEN CHEST CAVITY. PACING WIRES- WERE SPLITTING. THE SCRUB TECH REMOVED THE DEFECTIVE ITEMS, ENSURED THE PATIENT WASN'T HARMED, DISCUSSED WITH THE SURGEON AND ESCALATED TO TEAM LEADER. DEVICES ARE NOT AVAILABLE FOR RETURN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24727278 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT THE PRODUCT WAS SPLITTING. CV PACK THE SUTURE BOOTS WERE BREAKING/SPLITTING IN HALF, UNSAFE WHILE WORKING IN AN OPEN CHEST CAVITY. PACING WIRES- WERE SPLITTING. THE SCRUB TECH REMOVED THE DEFECTIVE ITEMS, ENSURED THE PATIENT WASN'T HARMED, DISCUSSED WITH THE SURGEON AND ESCALATED TO TEAM LEADER. DEVICES ARE NOT AVAILABLE FOR RETURN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24727245 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT THE PRODUCT WAS SPLITTING. CV PACK THE SUTURE BOOTS WERE BREAKING/SPLITTING IN HALF, UNSAFE WHILE WORKING IN AN OPEN CHEST CAVITY. PACING WIRES- WERE SPLITTING. THE SCRUB TECH REMOVED THE DEFECTIVE ITEMS, ENSURED THE PATIENT WASN'T HARMED, DISCUSSED WITH THE SURGEON AND ESCALATED TO TEAM LEADER. DEVICES ARE NOT AVAILABLE FOR RETURN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24727240 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT THE PRODUCT WAS SPLITTING. CV PACK THE SUTURE BOOTS WERE BREAKING/SPLITTING IN HALF, UNSAFE WHILE WORKING IN AN OPEN CHEST CAVITY. PACING WIRES- WERE SPLITTING. THE SCRUB TECH REMOVED THE DEFECTIVE ITEMS, ENSURED THE PATIENT WASN'T HARMED, DISCUSSED WITH THE SURGEON AND ESCALATED TO TEAM LEADER. DEVICES ARE NOT AVAILABLE FOR RETURN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24727107 ↗
  • MalfunctionJan 20, 2026Delivered as Unsterile Product

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE DEVICE PACKAGE WAS FOUND UNSEALED WHEN UNPACKING AND COUNTING. ONE END OF THE PACKAGE IS NOT SEALED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24119570 ↗
  • MalfunctionDec 26, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. WHEN PREPARING FOR USE BEFORE SURGERY, IT WAS FOUND THAT THE SUTURE IS PRESSED ONTO THE PACKAGING EDGE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23898583 ↗
  • MalfunctionJun 30, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. ENCLOSED PLEASE FIND DEFECT PHOTO. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22359513 ↗
  • MalfunctionJun 26, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. ENCLOSED PLEASE FIND DEFECT PHOTO. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22338569 ↗
  • MalfunctionJun 27, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE ON AN UNKNOWN DATE AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USING FOR MONITORING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630645 ↗
  • MalfunctionJun 27, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A STOP JUMPING AND BUILD A BRIDGE PROCEDURE ON AN UNKNOWN DATE AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USING FOR MONITORING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630634 ↗
  • MalfunctionJun 27, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TRICUSPID VALVE RECONSTRUCTION SURGERY PROCEDURE ON AN UNKNOWN DATE AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USING FOR MONITORING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630620 ↗
  • MalfunctionJun 27, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MITRAL VALVE REPLACEMENT PROCEDURE ON AN UNKNOWN DATE AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USING FOR MONITORING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630612 ↗
  • MalfunctionJun 27, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MITRAL VALVE REPLACEMENT PROCEDURE ON AN UNKNOWN DATE AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USING FOR MONITORING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630611 ↗
  • MalfunctionJun 27, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CONTINUOUSLY JUMPING AND BUILDING BRIDGES PROCEDURE ON AN UNKNOWN DATE AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USING FOR MONITORING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630580 ↗
  • MalfunctionJun 27, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CONTINUOUSLY JUMPING AND BUILDING BRIDGES PROCEDURE ON AN UNKNOWN DATE AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USING FOR MONITORING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630578 ↗
  • MalfunctionJul 28, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USE FOR MONITORING. CHANGED TO ANOTHER TWO TO CONTINUE THE CARDIAC SURGERY, AND THE SAME PROBLEM HAPPENED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17417688 ↗
  • MalfunctionJul 28, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USE FOR MONITORING. CHANGED TO ANOTHER TWO TO CONTINUE THE CARDIAC SURGERY, AND THE SAME PROBLEM HAPPENED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17417673 ↗
  • MalfunctionJul 28, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USE FOR MONITORING. CHANGED TO ANOTHER TWO TO CONTINUE THE CARDIAC SURGERY, AND THE SAME PROBLEM HAPPENED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17417646 ↗
  • MalfunctionJul 17, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BYPASS PROCEDURE ON (B)(6) 2023 AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17331833 ↗
  • MalfunctionJul 17, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BYPASS PROCEDURE ON (B)(6) 2023 AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17331830 ↗
  • MalfunctionJul 17, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BYPASS PROCEDURE ON (B)(6) 2023 AND TEMPORARY PACING WIRE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE (STRAIGHT NEEDLE) OCCURRED DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17331825 ↗

Other ELECTRODE, PACEMAKER, TEMPORARY devices

All ELECTRODE, PACEMAKER, TEMPORARY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.