Coated VICRYL VCP423HRecalls & reported problems

Ethicon Inc. · Antibacterial Undyed Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant

All models of Coated VICRYL (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 10705031053113), received between Jan 16, 2020 and Mar 11, 2026 — classified as 5 injuries, 14 malfunctions.

The most frequently reported specific device problem is “Failure to Degrade” (7 reports).

2 other recalls cover devices cleared under the same FDA submission (K032420, K132580, K181652); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 10 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Degrade7
  • Adverse Event Without Identified Device or Use Problem5
  • Material Separation5
  • Break2

Patient problems

  • Inflammation11
  • Pocket Erosion11
  • Wound Dehiscence5
  • No Consequences Or Impact To Patient4
  • Impaired Healing3
  • No Clinical Signs, Symptoms or Conditions3
  • Erythema2
  • Fluid Discharge2

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 11, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PATIENT EXPERIENCED INFLAMMATION AND REJECTION OF THE WIRE. SUTURE USED ON PTG/PTH/UVC LCA. SUTURED PLAN: SUBCUTANEOUS IN SEPARATE POINTS. RESORPTION PROBLEM WITH REPEATED INFLAMMATION : ABOUT 5 TO 10% OF PATIENTS. THE SURGEON HAS NOT CHANGED ANYTHING IN HIS PRACTICE EXCEPT THE SWITCH TO SUTURES MORE. THIS DOES NOT HAPPEN SYSTEMATICALLY BUT IT NOW CONCERNS SEVERAL CASES. IT HAS HAPPENED SEVERAL TIMES SINCE 3 OR 4 MONTHS SO IT IS DIFFICULT TO ESTABLISH WHICH BOX THE THREAD BELONGED TO. ESPECIALLY SINCE IT IS A THREAD THAT IS REGULARLY CONSUMED ON THE ESTABLISHMENT BY SEVERAL SURGEONS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Erythema; Fluid Discharge; Impaired Healing; Inflammation; Swelling/ Edema; Unspecified Infection; Wound Dehiscence

    MAUDE report 24576046 ↗
  • InjuryMar 11, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PATIENT EXPERIENCED INFLAMMATION AND REJECTION OF THE WIRE. SUTURE USED ON PTG/PTH/UVC LCA SUTURED PLAN: SUBCUTANEOUS IN SEPARATE POINTS RESORPTION PROBLEM WITH REPEATED INFLAMMATION: ABOUT 5 TO 10% OF PATIENTS THE SURGEON HAS NOT CHANGED ANYTHING IN HIS PRACTICE EXCEPT THE SWITCH TO SUTURES MORE THIS DOES NOT HAPPEN SYSTEMATICALLY BUT IT NOW CONCERNS SEVERAL CASES. IT HAS HAPPENED SEVERAL TIMES SINCE 3 OR 4 MONTHS SO IT IS DIFFICULT TO ESTABLISH WHICH BOX THE THREAD BELONGED TO. ESPECIALLY SINCE IT IS A THREAD THAT IS REGULARLY CONSUMED ON THE ESTABLISHMENT BY SEVERAL SURGEONS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Impaired Healing; Pocket Erosion; Wound Dehiscence

    MAUDE report 24571200 ↗
  • InjuryMar 11, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PATIENT EXPERIENCED INFLAMMATION AND REJECTION OF THE WIRE. SUTURE USED ON PTG/PTH/UVC LCA, SUTURED PLAN: SUBCUTANEOUS IN SEPARATE POINTS, RESORPTION PROBLEM WITH REPEATED INFLAMMATION : ABOUT 5 TO 10% OF PATIENTS, THE SURGEON HAS NOT CHANGED ANYTHING IN HIS PRACTICE EXCEPT THE SWITCH TO SUTURES MORE, THIS DOES NOT HAPPEN SYSTEMATICALLY BUT IT NOW CONCERNS SEVERAL CASES. IT HAS HAPPENED SEVERAL TIMES SINCE 3 OR 4 MONTHS SO IT IS DIFFICULT TO ESTABLISH WHICH BOX THE THREAD BELONGED TO. ESPECIALLY SINCE IT IS A THREAD THAT IS REGULARLY CONSUMED ON THE ESTABLISHMENT BY SEVERAL SURGEONS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Erythema; Fluid Discharge; Impaired Healing; Inflammation; Pocket Erosion; Swelling/ Edema; Unspecified Infection; Wound Dehiscence

    MAUDE report 24571196 ↗
  • MalfunctionMar 2, 2026Failure to Degrade

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS RESULTS IN INFLAMMATION AND REJECTION OF THE WIRE. SUTURE USED ON PTG/PTH/UVC LCA. SUBCUTANEOUS IN SEPARATE POINTS. RESORPTION PROBLEM WITH REPEATED INFLAMMATION: ABOUT 5 TO 10% OF PATIENTS. THE SURGEON HAS NOT CHANGED ANYTHING IN HIS PRACTICE EXCEPT THE SWITCH TO SUTURES MORE. THIS DOES NOT HAPPEN SYSTEMATICALLY BUT IT NOW CONCERNS SEVERAL CASES. IT HAS HAPPENED SEVERAL TIMES SINCE 3 OR 4 MONTHS SO IT IS DIFFICULT TO ESTABLISH WHICH BOX THE THREAD BELONGED TO. ESPECIALLY SINCE IT IS A THREAD THAT IS REGULARLY CONSUMED ON THE ESTABLISHMENT BY SEVERAL SURGEONS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation; Pocket Erosion

    MAUDE report 24484930 ↗
  • MalfunctionMar 2, 2026Failure to Degrade

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS RESULTS IN INFLAMMATION AND REJECTION OF THE WIRE. SUTURE USED ON PTG/PTH/UVC LCA. SUBCUTANEOUS IN SEPARATE POINTS. RESORPTION PROBLEM WITH REPEATED INFLAMMATION: ABOUT 5 TO 10% OF PATIENTS. THE SURGEON HAS NOT CHANGED ANYTHING IN HIS PRACTICE EXCEPT THE SWITCH TO SUTURES MORE. THIS DOES NOT HAPPEN SYSTEMATICALLY BUT IT NOW CONCERNS SEVERAL CASES. IT HAS HAPPENED SEVERAL TIMES SINCE 3 OR 4 MONTHS SO IT IS DIFFICULT TO ESTABLISH WHICH BOX THE THREAD BELONGED TO. ESPECIALLY SINCE IT IS A THREAD THAT IS REGULARLY CONSUMED ON THE ESTABLISHMENT BY SEVERAL SURGEONS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation; Pocket Erosion

    MAUDE report 24484876 ↗
  • MalfunctionMar 2, 2026Failure to Degrade

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS RESULTS IN INFLAMMATION AND REJECTION OF THE WIRE. SUTURE USED ON PTG/PTH/UVC LCA. SUBCUTANEOUS IN SEPARATE POINTS. RESORPTION PROBLEM WITH REPEATED INFLAMMATION: ABOUT 5 TO 10% OF PATIENTS. THE SURGEON HAS NOT CHANGED ANYTHING IN HIS PRACTICE EXCEPT THE SWITCH TO SUTURES MORE. THIS DOES NOT HAPPEN SYSTEMATICALLY BUT IT NOW CONCERNS SEVERAL CASES. IT HAS HAPPENED SEVERAL TIMES SINCE 3 OR 4 MONTHS SO IT IS DIFFICULT TO ESTABLISH WHICH BOX THE THREAD BELONGED TO. ESPECIALLY SINCE IT IS A THREAD THAT IS REGULARLY CONSUMED ON THE ESTABLISHMENT BY SEVERAL SURGEONS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation; Pocket Erosion

    MAUDE report 24484768 ↗
  • MalfunctionMar 2, 2026Failure to Degrade

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS RESULTS IN INFLAMMATION AND REJECTION OF THE WIRE. SUTURE USED ON PTG/PTH/UVC LCA. SUBCUTANEOUS IN SEPARATE POINTS. RESORPTION PROBLEM WITH REPEATED INFLAMMATION: ABOUT 5 TO 10% OF PATIENTS. THE SURGEON HAS NOT CHANGED ANYTHING IN HIS PRACTICE EXCEPT THE SWITCH TO SUTURES MORE. THIS DOES NOT HAPPEN SYSTEMATICALLY BUT IT NOW CONCERNS SEVERAL CASES. IT HAS HAPPENED SEVERAL TIMES SINCE 3 OR 4 MONTHS SO IT IS DIFFICULT TO ESTABLISH WHICH BOX THE THREAD BELONGED TO. ESPECIALLY SINCE IT IS A THREAD THAT IS REGULARLY CONSUMED ON THE ESTABLISHMENT BY SEVERAL SURGEONS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation; Pocket Erosion

    MAUDE report 24484764 ↗
  • MalfunctionMar 2, 2026Failure to Degrade

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS RESULTS IN INFLAMMATION AND REJECTION OF THE WIRE. SUTURE USED ON PTG/PTH/UVC LCA. SUBCUTANEOUS IN SEPARATE POINTS. RESORPTION PROBLEM WITH REPEATED INFLAMMATION: ABOUT 5 TO 10% OF PATIENTS. THE SURGEON HAS NOT CHANGED ANYTHING IN HIS PRACTICE EXCEPT THE SWITCH TO SUTURES MORE. THIS DOES NOT HAPPEN SYSTEMATICALLY BUT IT NOW CONCERNS SEVERAL CASES. IT HAS HAPPENED SEVERAL TIMES SINCE 3 OR 4 MONTHS SO IT IS DIFFICULT TO ESTABLISH WHICH BOX THE THREAD BELONGED TO. ESPECIALLY SINCE IT IS A THREAD THAT IS REGULARLY CONSUMED ON THE ESTABLISHMENT BY SEVERAL SURGEONS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation; Pocket Erosion

    MAUDE report 24484754 ↗
  • MalfunctionMar 2, 2026Failure to Degrade

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS RESULTS IN INFLAMMATION AND REJECTION OF THE WIRE. SUTURE USED ON PTG/PTH/UVC LCA. SUBCUTANEOUS IN SEPARATE POINTS. RESORPTION PROBLEM WITH REPEATED INFLAMMATION: ABOUT 5 TO 10% OF PATIENTS. THE SURGEON HAS NOT CHANGED ANYTHING IN HIS PRACTICE EXCEPT THE SWITCH TO SUTURES MORE. THIS DOES NOT HAPPEN SYSTEMATICALLY BUT IT NOW CONCERNS SEVERAL CASES. IT HAS HAPPENED SEVERAL TIMES SINCE 3 OR 4 MONTHS SO IT IS DIFFICULT TO ESTABLISH WHICH BOX THE THREAD BELONGED TO. ESPECIALLY SINCE IT IS A THREAD THAT IS REGULARLY CONSUMED ON THE ESTABLISHMENT BY SEVERAL SURGEONS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation; Pocket Erosion

    MAUDE report 24484640 ↗
  • MalfunctionMar 2, 2026Failure to Degrade

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THIS RESULTS IN INFLAMMATION AND REJECTION OF THE WIRE. SUTURE USED ON PTG/PTH/UVC LCA. SUBCUTANEOUS IN SEPARATE POINTS. RESORPTION PROBLEM WITH REPEATED INFLAMMATION: ABOUT 5 TO 10% OF PATIENTS. THE SURGEON HAS NOT CHANGED ANYTHING IN HIS PRACTICE EXCEPT THE SWITCH TO SUTURES MORE. THIS DOES NOT HAPPEN SYSTEMATICALLY BUT IT NOW CONCERNS SEVERAL CASES. IT HAS HAPPENED SEVERAL TIMES SINCE 3 OR 4 MONTHS SO IT IS DIFFICULT TO ESTABLISH WHICH BOX THE THREAD BELONGED TO. ESPECIALLY SINCE IT IS A THREAD THAT IS REGULARLY CONSUMED ON THE ESTABLISHMENT BY SEVERAL SURGEONS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation; Pocket Erosion

    MAUDE report 24484542 ↗
  • MalfunctionJan 29, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED, DURING THE PROCEDURE, IT WAS FOUND THAT THE SUTURE WAS ALREADY BROKEN OR THE SUTURE WAS ALREADY DETACHED FROM THE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18594603 ↗
  • MalfunctionAug 29, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE DR WAS ATTEMPTING TO CLOSE A PORT SITE WITH VICRYL SUTURE VCP423, AND HAD TAKEN ONE BITE THROUGH THE TISSUE, WHEN THE NEEDLE BROKE OFF THE SUTURE. THE NEEDLE HAD TO BE RETRIEVED, AND THE SUTURE WAS REMOVED AND SET ASIDE. THIS IS THE FIRST TIME THE SURGEON HAS EXPERIENCED THIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17646646 ↗
  • MalfunctionJan 6, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE SUTURE BREAKS EASILY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16115893 ↗
  • InjuryAug 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PLASTIC SURGERY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. TWO TO THREE WEEKS POST OP, THE PATIENTS PRESENTED WITH SUTURES EXTRUDING FROM THE BODY AND OPEN WOUNDS. THE SUTURES WERE REMOVED AND WOUND MANAGEMENT INCLUDING ANTIBIOTICS WAS INSTIGATED. THE PATIENT WAS HAVING SUTURE SENSITIVITY AND EXTRUSION OF THE SUTURES. ADDITIONAL INFORMATION WAS REQUESTED. NO FURTHER INFORMATION AVAILABLE AT THIS TIME.

    Patient problems: Inflammation; Pocket Erosion; Scar Tissue; Wound Dehiscence

    MAUDE report 15301173 ↗
  • InjuryJul 19, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PLASTIC SURGERY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. TWO TO THREE WEEKS POST OP, THE PATIENTS PRESENTED WITH SUTURES EXTRUDING FROM THE BODY AND OPEN WOUNDS. THE SUTURES WERE REMOVED AND WOUND MANAGEMENT INCLUDING ANTIBIOTICS WAS INSTIGATED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation; Pocket Erosion; Scar Tissue; Wound Dehiscence

    MAUDE report 15057195 ↗
  • MalfunctionJan 16, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST IMPLANT PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. THE SUTURE LOOSED FROM THE NEEDLE DURING THE PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9599487 ↗
  • MalfunctionJan 16, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST IMPLANT PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. THE SUTURE LOOSED FROM THE NEEDLE DURING THE PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9599486 ↗
  • MalfunctionJan 16, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST IMPLANT PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. THE SUTURE LOOSED FROM THE NEEDLE DURING THE PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9599485 ↗
  • MalfunctionJan 16, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST IMPLANT PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. THE SUTURE LOOSED FROM THE NEEDLE DURING THE PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9599484 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K032420, K132580, K181652. One submission often covers many models, so check the product and lot details in each record.

  • Z-2145-2024 · May 10, 2024 · Open, Classified
    VICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological
  • Z-1310-2009 · Feb 2, 2009 · Terminated
    Coated VICRYL Plus Antibacterial (polyglactin 910) Suture; Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates