Skip to content

Coated VICRYL VCP494HRecalls & reported problems

Ethicon, LLC · Antibacterial Undyed Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant

All models of Coated VICRYL (350)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10705031053700), received between Jul 24, 2023 and Jan 27, 2025 — classified as 6 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Material Separation” (6 reports).

2 other recalls cover devices cleared under the same FDA submission (K032420, K132580, K181652); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem6
  • Material Separation6
  • Break3
  • Dull, Blunt1

Patient problems

  • Wound Dehiscence6
  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 27, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST MASS EXCISION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE NEEDLE BROKE WHEN SEWING IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21246197 ↗
  • MalfunctionAug 13, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE TIP BROKE OFF. THE SURGEON HAD NOT GRASPED THE TIP OF THE NEEDLE. FRAGMENTS WERE GENERATED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19970997 ↗
  • MalfunctionAug 13, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE TIP BROKE OFF. THE SURGEON HAD NOT GRASPED THE TIP OF THE NEEDLE. FRAGMENTS WERE GENERATED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19970994 ↗
  • MalfunctionFeb 14, 2024Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN 2023 AND SUTURE WAS USED. THE NEEDLE WAS TOO DULL. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18703246 ↗
  • InjuryJul 24, 2023Material Separation; Adverse Event Without Identified Device or Use Problem; Material Separation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE CAME OFF THE SUTURE CAUSING A DEHISCENCE ON THE PATIENT (FACE, LIPS, MOUTH). THE SURGERY WAS DELAYED BY 30 MINUTES. IT WAS REPORTED THAT DURING THE INITIAL PROCEDURE, ¿THE PATIENT HAD TO HAVE SURGERY AGAIN¿. THIS EVENT WAS REPORTED TO HAVE OCCURRED INTRAOPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Wound Dehiscence

    MAUDE report 17385146 ↗
  • InjuryJul 24, 2023Material Separation; Adverse Event Without Identified Device or Use Problem; Material Separation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE CAME OFF THE SUTURE CAUSING A DEHISCENCE ON THE PATIENT (FACE, LIPS, MOUTH). THE SURGERY WAS DELAYED BY 30 MINUTES. IT WAS REPORTED THAT DURING THE INITIAL PROCEDURE, ¿THE PATIENT HAD TO HAVE SURGERY AGAIN¿. THIS EVENT WAS REPORTED TO HAVE OCCURRED INTRAOPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Wound Dehiscence

    MAUDE report 17385139 ↗
  • InjuryJul 24, 2023Material Separation; Adverse Event Without Identified Device or Use Problem; Material Separation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE CAME OFF THE SUTURE CAUSING A DEHISCENCE ON THE PATIENT (FACE, LIPS, MOUTH). THE SURGERY WAS DELAYED BY 30 MINUTES. IT WAS REPORTED THAT DURING THE INITIAL PROCEDURE, ¿THE PATIENT HAD TO HAVE SURGERY AGAIN¿. THIS EVENT WAS REPORTED TO HAVE OCCURRED INTRAOPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Wound Dehiscence

    MAUDE report 17385117 ↗
  • InjuryJul 24, 2023Material Separation; Adverse Event Without Identified Device or Use Problem; Material Separation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE CAME OFF THE SUTURE CAUSING A DEHISCENCE ON THE PATIENT (FACE, LIPS, MOUTH). THE SURGERY WAS DELAYED BY 30 MINUTES. IT WAS REPORTED THAT DURING THE INITIAL PROCEDURE, ¿THE PATIENT HAD TO HAVE SURGERY AGAIN¿. THIS EVENT WAS REPORTED TO HAVE OCCURRED INTRAOPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Wound Dehiscence

    MAUDE report 17385087 ↗
  • InjuryJul 24, 2023Material Separation; Adverse Event Without Identified Device or Use Problem; Material Separation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE CAME OFF THE SUTURE CAUSING A DEHISCENCE ON THE PATIENT (FACE, LIPS, MOUTH). THE SURGERY WAS DELAYED BY 30 MINUTES. IT WAS REPORTED THAT DURING THE INITIAL PROCEDURE, ¿THE PATIENT HAD TO HAVE SURGERY AGAIN¿. THIS EVENT WAS REPORTED TO HAVE OCCURRED INTRAOPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Wound Dehiscence

    MAUDE report 17385060 ↗
  • InjuryJul 24, 2023Material Separation; Adverse Event Without Identified Device or Use Problem; Material Separation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE CAME OFF THE SUTURE CAUSING A DEHISCENCE ON THE PATIENT (FACE, LIPS, MOUTH). THE SURGERY WAS DELAYED BY 30 MINUTES. IT WAS REPORTED THAT DURING THE INITIAL PROCEDURE, ¿THE PATIENT HAD TO HAVE SURGERY AGAIN¿. THIS EVENT WAS REPORTED TO HAVE OCCURRED INTRAOPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Wound Dehiscence

    MAUDE report 17385041 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K032420, K132580, K181652. One submission often covers many models, so check the product and lot details in each record.

  • Z-2145-2024 · May 10, 2024 · Open, Classified
    VICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological
  • Z-1310-2009 · Feb 2, 2009 · Terminated
    Coated VICRYL Plus Antibacterial (polyglactin 910) Suture; Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page