Skip to content

Coated VICRYL VCP500H (UDI 10705031053861)Recalls & reported problems

Ethicon, LLC · Antibacterial Undyed Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant

All models of Coated VICRYL (350)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Coated VICRYL VCP500H is a braided absorbable synthetic suture used to close wounds. The 68 reports mostly describe the thread separating from the needle mid-stitch, and a few describe the thread or needle breaking, which lengthened surgery but left no reported patient harm. No deaths or injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
70
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 10705031053861), received between Jul 22, 2021 and Aug 17, 2026 — classified as 70 malfunctions.

The most frequently reported specific device problem is “Material Separation” (62 reports).

2 other recalls cover devices cleared under the same FDA submission (K032420, K132580, K181652); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation62
  • Break8

Patient problems

  • No Clinical Signs, Symptoms or Conditions70

Recent adverse event reports

The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 17, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50275720 ↗
  • MalfunctionAug 17, 2026Material Separation

    It was reported that a patient underwent a resection procedure on (b)(6) 2026 and suture was used. The patient underwent surgery due to a facial mass. After the resection and suturing were completed, the suture and needle fell off. The doctor replaced the needle and completed the surgery. At same time, the HCP also claimed that multiple needles pull off issue case occurred recently. No adverse patient consequences were reported. Additional information was requested.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50275294 ↗
  • MalfunctionJan 16, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT SUPERFICIAL TUMOR RESECTION SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. AT 14:09, THE DOCTOR WAS SUTURING THE PATIENT'S WOUND AND PREPARING TO USE THE SUTURE. HOWEVER, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED, AND A NEW SUTURE WAS IMMEDIATELY REPLACED. THIS INCIDENT RESULTED IN AN EXTENSION OF THE SURGICAL SUTURE TIME. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24100302 ↗
  • MalfunctionNov 10, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A RESECTION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. AT 14.57, IT WAS FOUND THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHEN SUTURING THE PATIENT'S WOUND. IMMEDIATELY REPLACED WITH A NEW SUTURE AND EXPLAINED TO THE PATIENT. RESULTING IN PROLONGED TREATMENT TIME. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23512139 ↗
  • MalfunctionSep 12, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED AFTER THE FIRST STITCH. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23040092 ↗
  • MalfunctionSep 12, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED AFTER THE FIRST STITCH. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23040090 ↗
  • MalfunctionSep 12, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. THE FIRST STITCH WAS USED TO SUTURE THE SKIN, AND THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED. REPLACED IT WITH A NEW ONE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23039859 ↗
  • MalfunctionAug 22, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MOLE CUTTING SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN MOLE CUTTING SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22871740 ↗
  • MalfunctionAug 22, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A MOLE CUTTING SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN MOLE CUTTING SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22871733 ↗
  • MalfunctionMay 9, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A DENTAL PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. AFTER PERFORMING SURGERY, SURGEON USED TWO DIFFERENT TYPES OF SUTURES FOR GUM CLOSURE, HOWEVER, THEY FOUND THAT SEVERAL PIECES OF SUTURE DETACHED FROM THE SEWAGE AFTER THE 1ST PASS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21994766 ↗
  • MalfunctionMay 9, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A DENTAL PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. AFTER PERFORMING SURGERY, SURGEON USED TWO DIFFERENT TYPES OF SUTURES FOR GUM CLOSURE, HOWEVER THEY FOUND THAT SEVERAL PIECES OF SUTURE DETACHED FROM THE SEWAGE AFTER THE 1ST PASS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21994707 ↗
  • MalfunctionMay 9, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A DENTAL PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. AFTER PERFORMING SURGERY, SURGEON USED TWO DIFFERENT TYPES OF SUTURES FOR GUM CLOSURE, HOWEVER THEY FOUND THAT SEVERAL PIECES OF SUTURE DETACHED FROM THE SEWAGE AFTER THE 1ST PASS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21994682 ↗
  • MalfunctionNov 21, 2024Material Separation; Material Separation

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. WHEN SUTURING THE SKIN WITH SUTURE, IT WAS FOUND THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAD HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE SURGERY DELAYED FOR UNKNOWN TIME. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20751354 ↗
  • MalfunctionNov 21, 2024Material Separation; Material Separation

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. WHEN SUTURING THE SKIN WITH SUTURE, IT WAS FOUND THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAD HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE SURGERY DELAYED FOR UNKNOWN TIME. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20749838 ↗
  • MalfunctionAug 28, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHEN SUTURING SKIN. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20085539 ↗
  • MalfunctionMar 8, 2024Material Separation; Material Separation

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE THREAD GOT SEPARATED FROM THE NEEDLE WHILE SUTURING. NO ADVERSE PATIENT CONSEQUENCES HAVE BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18865400 ↗
  • MalfunctionOct 12, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE NEEDLES ARE POPPING OFF OF THE SUTURES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17925164 ↗
  • MalfunctionOct 12, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE NEEDLES ARE POPPING OFF OF THE SUTURES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17925056 ↗
  • MalfunctionOct 4, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE SUTURE DETACHED FROM THE NEEDLE AFTER OPENING THE PACKAGE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17870890 ↗
  • MalfunctionJun 6, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE IN (B)(6)2023 AND SUTURE WAS USED. THE SUTURE WAS BROKEN DURING INTRA-OPERATION. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17072485 ↗
  • MalfunctionApr 20, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A NEUROSURGERY PROCEDURE IN 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16782195 ↗
  • MalfunctionApr 10, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE OUT FROM THE NEEDLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16710161 ↗
  • MalfunctionApr 10, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE OUT FROM THE NEEDLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16710155 ↗
  • MalfunctionApr 10, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE OUT FROM THE NEEDLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16710154 ↗
  • MalfunctionApr 10, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE OUT FROM THE NEEDLE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16710153 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K032420, K132580, K181652. One submission often covers many models, so check the product and lot details in each record.

  • Z-2145-2024 · May 10, 2024 · Open, Classified
    VICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological
  • Z-1310-2009 · Feb 2, 2009 · Terminated
    Coated VICRYL Plus Antibacterial (polyglactin 910) Suture; Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page