Coated VICRYL VCP603HRecalls & reported problems
Ethicon, LLC · Antibacterial Violet Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (350)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is a coated Vicryl absorbable suture used in surgery. Reports mostly describe needles or sutures breaking during procedures, sometimes leaving fragments in the patient. There are 78 reports naming this device, no deaths and 5 injuries. One open recall started in May 2024 addresses a packaging hole that could compromise sterility and lead to infection.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 82
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 82 adverse event reports that list this exact device (DI 10705031054394), received between Nov 5, 2019 and Sep 30, 2026 — classified as 5 injuries, 77 malfunctions.
The most frequently reported specific device problem is “Break” (52 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K032420, K132580, K181652); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 10, 2024Class IIOpen, Classified
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
Z-2145-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Break52
- Material Separation23
- Adverse Event Without Identified Device or Use Problem2
- Component Missing1
- Defective Component1
- Detachment of Device or Device Component1
- Device Contaminated During Manufacture or Shipping1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions63
- No Consequences Or Impact To Patient9
- Insufficient Information7
- Foreign Body In Patient2
- Hypersensitivity/Allergic reaction2
- Wound Dehiscence2
- No Code Available1
- Pain1
Recent adverse event reports
The 25 most recent of 82 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Break
It was reported a patient underwent an unknown procedure on (b)(6) 2026 and suture was used. When opening the package, the wire was broken, during the procedure before use.Upon receipt and analysis, a short insertion was noted.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946712 ↗MalfunctionSep 28, 2026Break
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used. It was reported that the needle broken on the suture during the surgery. There is no Adverse Impact on patient/user reported. Device is not available for return. Additional information was requested.
Patient problems: Insufficient Information
MAUDE report 50917545 ↗MalfunctionSep 1, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50548953 ↗MalfunctionAug 18, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50308076 ↗MalfunctionJun 18, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT ON AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN SUTURING. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25549572 ↗MalfunctionMay 1, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC APPENDECTOMY PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE OFF OF THE SUTURE WHILE SURGEON WAS SUTURING IN THE FASCIA. NO PATIENT CONSEQUENCES. DEVICE IS AVAILABLE FOR RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25057379 ↗MalfunctionApr 9, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN LAPAROSCOPIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING CLOSING FASCIA ON LAPAROSCOPIC SITE, A NEEDLE BROKE. DEVICE IS AVAILABLE FOR RETURN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24829498 ↗MalfunctionApr 7, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A PT ROBOTIC PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY THE SALES REP THAT DURING A PT ROBOTIC PROCEDURE, THE SUTURE BROKE INTO TWO PIECES. NO PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24798815 ↗MalfunctionMar 25, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A HYSTERECTOMY ON (B)(6) 2026 AND SUTURE WAS USED. THE NEEDLE BROKE. THE NEEDLE DID FALL INTO THE PATIENT. IT TOOK A LOT OF TIME TO FIND THE NEEDLE IN THE PATIENT. THE INCISION HAD TO BE EXTENDED. THEY USED ULTRASOUND AND X-RAY TO LOCATE IT. THE NEEDLE WAS RETRIEVED DURING THE PROCEDURE. NO ADDITIONAL TISSUE DAMAGE OCCURRED. THERE WERE NO PATIENT ADVERSE EVENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24690874 ↗MalfunctionMar 18, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A GYN ROBOTIC PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT DURING THE GYN ROBOTIC PROCEDURE, THE NEEDLE GOT BROKEN INTO TWO PIECES. DEVICES ARE RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24628260 ↗MalfunctionMar 13, 2026
DURING CLOSING FASCIA ON LAPAROSCOPIC SITE, A NEEDLE (O VICRYL UR-6) BROKE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24589480 ↗MalfunctionJan 16, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. NEEDLE BROKE ON THE SECOND PASS THROUGH TISSUE FOR A HASSON TROCAR PLACEMENT. NEEDLE BROKE A QUARTER LENGTH OF THE NEEDLE DOWN FROM THE SWAGE. NEVER HAPPENED BEFORE. THERE WERE NO SIGNIFICANT DELAYS. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24093240 ↗MalfunctionDec 22, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL RETURN 3 ACTUAL PRODUCTS AND 102 REPRESENTATIVE SAMPLES FOR ANALYSIS. ENCLOSED PLEASE FIND DEFECT PHOTO. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23864533 ↗MalfunctionDec 22, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL RETURN 3 ACTUAL PRODUCTS AND 102 REPRESENTATIVE SAMPLES FOR ANALYSIS. ENCLOSED PLEASE FIND DEFECT PHOTO. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23864521 ↗MalfunctionDec 22, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 3 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL RETURN 3 ACTUAL PRODUCTS AND 102 REPRESENTATIVE SAMPLES FOR ANALYSIS. ENCLOSED PLEASE FIND DEFECT PHOTO. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23864519 ↗MalfunctionDec 19, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT LAPAROSCOPIC REPAIR OF EXTRAPERITONEAL HERNIA ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE SURGERY, THE DOCTOR SUTURED THE OUTER LAYER OF THE PERITONEUM AND FOUND THAT THE NEEDLE TIP WAS MISSING WHEN REMOVING THE NEEDLE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23852325 ↗MalfunctionDec 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. AT THE END OF THE PROCEDURE THE SUTURE THE NEEDLE IS BROKEN, SO A NEW ONE IS REQUESTED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23788739 ↗MalfunctionNov 4, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DISTRIBUTOR PER ACCOUNT THAT THE CST REPORTING ITEM DEFECTIVE. TIP OF THE SUTURE NEEDLE BROKE OFF IN THE PATIENT. PRODUCT IS NOT AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 23465280 ↗InjuryOct 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A MINIMALLY INVASIVE LUMBAR HERNIA SURGERY ON (B)(6) 2025 AND SUTURE WAS USED TO CLOSE THE SURGICAL WOUND FOR FASCIAL CLOSURE. ON (B)(6) 2025, AFTER BEING DISCHARGED, THE PATIENT TOLD THE DOCTOR THAT THE WOUND WAS OOZING AND DEHISCING. THE DOCTOR IMMEDIATELY PRESCRIBED MEDICATION TO PREVENT INFECTION AND BEGAN WOUND MANAGEMENT. REVIEWING THE INCIDENT WITH THE DOCTOR, HE STATED THAT THE PATIENT'S CASE WAS NOT RELATED TO A SURGICAL SITE INFECTION, AS THERE WAS RE-EPITHELIALIZATION AT THE WOUND EDGES AND THERE WEREN´T ANY OTHER SSI ALARM DATA. RATHER, IT WAS AN ALLERGIC REACTION TO THE SUTURE. ONCE THE STITCH OF SUTURE WAS REMOVED, THE INFLAMMATORY REACTION SUBSIDED, AND A SECONDARY WOUND CLOSURE WAS POSSIBLE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hypersensitivity/Allergic reaction; Insufficient Information; Purulent Discharge; Wound Dehiscence
MAUDE report 23361370 ↗InjurySep 9, 2025Positioning Failure; Positioning Failure
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPY ON (B)(6) 2025 AND SUTURE WAS USED FOR UMBILICAL ORIFICE CLOSURE. SURGICAL RESUMPTION (B)(6) 2025 FOR EVISCERATION; DURING RESUMPTION FINDING THAT THE THREADS AND APONEUROSIS ARE INTACT, BUT THE KNOTS HAVE LOOSENED (SLIPPING), DUE TO THE COATED CHARACTER OF THE THREAD. PATIENT HAD EVISCERATION WITH EMERGENCY RECOVERY. THE OLD REGULAR SUTURE WAS USED DURING THE RE-OP. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Wound Dehiscence
MAUDE report 23003441 ↗MalfunctionAug 11, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND A BARBED SUTURE WAS USED. DURING SURGERY, THE NEEDLE WAS BROKEN AT THE SWAGE PART WHEN CLOSING THE TROCAR WOUND DURING USE. ONE NEEDLE WAS BROKEN DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22773983 ↗MalfunctionJun 13, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A GENERAL SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE SURGERY, THE SUTURE DETACHED EASILY FROM THE NEEDLE. THIS HAPPENED TWICE THIS WEEK WITH THIS LOT NUMBER, AND THEY COULD HAVE LOST THE NEEDLE IN THE PATIENT'S ABDOMEN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22203183 ↗MalfunctionJun 13, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A GENERAL SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE SURGERY, THE SUTURE DETACHED EASILY FROM THE NEEDLE. THIS HAPPENED TWICE THIS WEEK WITH THIS LOT NUMBER, AND THEY COULD HAVE LOST THE NEEDLE IN THE PATIENT'S ABDOMEN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22203181 ↗MalfunctionMay 29, 2025Component Missing; Component Missing
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, INCORRECT QUANTITY-MISSING PACKAGES. IT WAS REPORTED THAT THERE WAS ONE PACKAGE OF PRODUCT LESS THAN EXPECTED IN THE BOX WHEN JUST OPENED . THERE SHOULD BE THIRTY-SIX PACKAGE OF PRODUCTS IN THE BOX, BUT ACTUALLY THERE WERE ONLY THIRTY-FIVE PACKAGE OF PRODUCTS. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. UPON THE VISUAL INSPECTION OF THE RECEIVED BOX, IT WAS FOUND TO CONTAIN THIRTY-FIVE FOIL PACKETS INSTEAD OF THIRTY-SIX. AS PER PRODUCT REQUIREMENTS, THE BOX SHOULD CONTAIN THIRTY-SIX PACKAGES. THE TAB DISPENSER WAS REMOVED FROM THE BOX, AND ONE SIDE OF THE BOX WAS CRUSHED. IN ADDITION, IT WAS NOTED THAT THE FOIL PACKETS WERE RELABELED. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22108299 ↗MalfunctionApr 17, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A HARTMAN REVERSAL PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE NEEDLE BROKE OFF OF THE SUTURE. CASE WAS COMPLETED WITH A LIKE DEVICE. NEEDLE WAS FOUND AND REMOVED FROM PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21858071 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K032420, K132580, K181652. One submission often covers many models, so check the product and lot details in each record.
- Z-1310-2009 · Feb 2, 2009 · TerminatedCoated VICRYL Plus Antibacterial (polyglactin 910) Suture; Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page