Coated VICRYL VCP663HRecalls & reported problems
Ethicon, LLC · Antibacterial Undyed Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (350)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Coated VICRYL absorbable surgical suture is used to close wounds. Reports mostly describe the suture material separating or not breaking down as expected, along with wound infections, redness, allergic reactions, and wounds reopening with the suture pushing through the skin. There are 48 reports, 24 injuries and no deaths, and no recalls name this device. Reports rose sharply in the past 12 months, from 2 to 45.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 47
- Deaths reported
- 0
- Injuries reported
- 23
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 47 adverse event reports that list this exact device (DI 10705031054691), received between May 13, 2024 and Aug 5, 2026 — classified as 23 injuries, 24 malfunctions.
The most frequently reported specific device problem is “Material Separation” (23 reports).
2 other recalls cover devices cleared under the same FDA submission (K032420, K132580, K181652); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem23
- Material Separation23
- Failure to Degrade13
- Component Misassembled1
Patient problems
- No Clinical Signs, Symptoms or Conditions24
- Unspecified Infection18
- Erythema15
- Hypersensitivity/Allergic reaction13
- Pocket Erosion8
- Wound Dehiscence8
- Impaired Healing2
- Itching Sensation2
Recent adverse event reports
The 25 most recent of 47 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50147382 ↗MalfunctionAug 5, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50147381 ↗InjuryMay 26, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PATIENT DEVELOPED AN INFECTION, WHERE CLINICAL ASSESSMENT INDICATES THAT THE SURGICAL WOUND WAS IN GOOD CONDITION, HOWEVER, THERE WAS DELAYED WOUND HEALING WHICH APPEARED TO BE ASSOCIATED WITH THE TYPE OF SUTURE MATERIAL USED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Erythema; Impaired Healing; Itching Sensation; Purulent Discharge; Unspecified Infection
MAUDE report 25284045 ↗InjuryMay 26, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PATIENT DEVELOPED AN INFECTION, WHERE CLINICAL ASSESSMENT INDICATES THAT THE SURGICAL WOUND WAS IN GOOD CONDITION, HOWEVER, THERE WAS DELAYED WOUND HEALING WHICH APPEARED TO BE ASSOCIATED WITH THE TYPE OF SUTURE MATERIAL USED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Erythema; Impaired Healing; Itching Sensation; Purulent Discharge; Unspecified Infection
MAUDE report 25283914 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED AN INCISION COMPLICATION AND REQUIRED INCISION PACKING. THE PATIENT MAY HAVE HAD A WOUND DEHISCENCE AND THE SUTURE PROTRUDING FROM THE INCISION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Pocket Erosion; Wound Dehiscence
MAUDE report 24729468 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED A WOUND INFECTION AND WAS STARTED ON ANTIBIOTICS. THE PATIENT MAY HAVE HAD A WOUND DEHISCENCE AND THE SUTURE PROTRUDING FROM THE INCISION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Pocket Erosion; Unspecified Infection; Wound Dehiscence
MAUDE report 24729459 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED AN INCISION COMPLICATION AND REQUIRED INCISION PACKING. THE PATIENT MAY HAVE HAD A WOUND DEHISCENCE AND THE SUTURE PROTRUDING FROM THE INCISION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Pocket Erosion; Wound Dehiscence
MAUDE report 24727957 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED A WOUND INFECTION AND WAS STARTED ON ANTIBIOTICS. THE PATIENT MAY HAVE HAD A WOUND DEHISCENCE AND THE SUTURE PROTRUDING FROM THE INCISION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Pocket Erosion; Unspecified Infection; Wound Dehiscence
MAUDE report 24727952 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED A WOUND INFECTION AND WAS STARTED ON ANTIBIOTICS. THE PATIENT MAY HAVE HAD A WOUND DEHISCENCE AND THE SUTURE PROTRUDING FROM THE INCISION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Pocket Erosion; Unspecified Infection; Wound Dehiscence
MAUDE report 24727237 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED AN INCISION COMPLICATION AND REQUIRED INCISION PACKING. THE PATIENT MAY HAVE HAD A WOUND DEHISCENCE AND THE SUTURE PROTRUDING FROM THE INCISION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Pocket Erosion; Wound Dehiscence
MAUDE report 24727200 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED A WOUND INFECTION AND WAS STARTED ON ANTIBIOTICS. THE PATIENT MAY HAVE HAD A WOUND DEHISCENCE AND THE SUTURE PROTRUDING FROM THE INCISION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Pocket Erosion; Unspecified Infection; Wound Dehiscence
MAUDE report 24727191 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE PATIENT EXPERIENCED AN INCISION COMPLICATION AND REQUIRED INCISION PACKING. THE PATIENT MAY HAVE HAD A WOUND DEHISCENCE AND THE SUTURE PROTRUDING FROM THE INCISION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Pocket Erosion; Wound Dehiscence
MAUDE report 24727097 ↗MalfunctionFeb 25, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT TOTAL 3 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WILL RETURN 3 ACTUAL PRODUCTS FOR ANALYSIS. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24442678 ↗MalfunctionFeb 25, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT TOTAL 3 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WILL RETURN 3 ACTUAL PRODUCTS FOR ANALYSIS. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24442676 ↗MalfunctionFeb 25, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT TOTAL 3 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WILL RETURN 3 ACTUAL PRODUCTS FOR ANALYSIS. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24442674 ↗MalfunctionFeb 23, 2026Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. NEEDLE PULL OFF ISSUE IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24413418 ↗MalfunctionFeb 23, 2026Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. NEEDLE PULL OFF ISSUE IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24413414 ↗MalfunctionFeb 23, 2026Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. NEEDLE PULL OFF ISSUE IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24413413 ↗MalfunctionFeb 23, 2026Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. NEEDLE PULL OFF ISSUE IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24413412 ↗MalfunctionFeb 23, 2026Material Separation
IT WAS REPORTED A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. NEEDLE PULL OFF ISSUE IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24413340 ↗MalfunctionFeb 2, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. A TOTAL OF TWO PRODUCTS HAD THE NEEDLE PULL OFF ISSUE OCCUR. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24241277 ↗MalfunctionFeb 2, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6)2026 AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. A TOTAL OF TWO PRODUCTS HAD THE NEEDLE PULL OFF ISSUE OCCUR. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24241159 ↗MalfunctionFeb 2, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. TOTAL 2 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THE NEEDLE DID NOT FELL INTO THE PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24240016 ↗MalfunctionFeb 2, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY . TOTAL 2 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THE NEEDLE DID NOT FELL INTO THE PATIENT . NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24240009 ↗MalfunctionJan 30, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. TOTAL OF 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24228750 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K032420, K132580, K181652. One submission often covers many models, so check the product and lot details in each record.
- Z-2145-2024 · May 10, 2024 · Open, ClassifiedVICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological
- Z-1310-2009 · Feb 2, 2009 · TerminatedCoated VICRYL Plus Antibacterial (polyglactin 910) Suture; Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page