Coated VICRYL VCP935HRecalls & reported problems
Ethicon, LLC · Antibacterial Undyed Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (350)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10705031056060), received between Apr 29, 2020 and Jun 9, 2026 — classified as 3 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Material Separation” (12 reports).
2 other recalls cover devices cleared under the same FDA submission (K032420, K132580, K181652); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation12
- Adverse Event Without Identified Device or Use Problem3
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- Erythema3
- Impaired Healing3
- Inflammation3
- Necrosis3
- Skin Inflammation/ Irritation3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2026, AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25444114 ↗MalfunctionJul 25, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE ARE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22620919 ↗MalfunctionJul 25, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT C-SECTION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22610696 ↗MalfunctionJun 30, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A PERINEAL LATERAL EXCISION REPAIR SURGERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22359316 ↗MalfunctionJun 23, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A PERINEAL TEAR REPAIR PROCEDURE ((B)(6) 2025 AND A SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22306242 ↗MalfunctionJun 11, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHEN SUTURING SKIN. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22188517 ↗MalfunctionNov 18, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20708812 ↗MalfunctionNov 11, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND A SUTURE DEVICE WAS USED. WHEN MEDICAL STAFF WERE USING THE SUTURE DURING SURGERY, THE SUTURE BROKE FROM THE END OF THE NEEDLE. AS THE SUTURE WAS INITIALLY USED FOR A SHORT PERIOD OF TIME, THE MEDICAL STAFF IMMEDIATELY WITHDREW THE SUTURE AND REPLACED IT WITH A NEW ONE TO CONTINUE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20659158 ↗MalfunctionNov 11, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2024 AND A SUTURE DEVICE WAS USED. WHEN MEDICAL STAFF USED THE SUTURE DURING SURGERY, THE SUTURE BROKE FROM THE END OF THE NEEDLE. AS THE WOUND WAS HALF CLOSED, THE MEDICAL STAFF IMMEDIATELY USED A DISPOSABLE LEATHER NEEDLE TO CONNECT THE SUTURE AND CONTINUE SUTURING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20658604 ↗MalfunctionAug 22, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE DR. FOUND OUT THE SUTURE THREAD DETACHED FROM THE NEEDLE WHEN PASSING THROUGH TISSUE. HE OPENED ANOTHER ONE AND SAME INCIDENT HAPPENED AS WELL. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20034295 ↗MalfunctionAug 22, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE DR. FOUND OUT THE SUTURE THREAD DETACHED FROM THE NEEDLE WHEN PASSING THROUGH TISSUE. HE OPENED ANOTHER ONE AND SAME INCIDENT HAPPENED AS WELL. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20031451 ↗InjuryJan 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED ON THE SKIN. POST-OP, TWO DAYS AFTER THE SURGERY, THE WOUND WAS FOUND WITH ERYTHEMA AND INFLAMMATION DURING DRESSING CHANGE. ABOUT 7 DAYS AFTER SURGERY, DARK SKIN TONE AND EPIDERMAL NECROSIS WERE FOUND AROUND THE INCISION SITE. THE PATIENT WAS GIVEN A DRESSING CHANGE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Erythema; Impaired Healing; Inflammation; Necrosis; Skin Inflammation/ Irritation
MAUDE report 18581474 ↗InjuryJan 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON (B)(6) 2023 AND SUTURE WAS USED ON THE SKIN. POST-OP, TWO DAYS AFTER THE SURGERY, THE WOUND WAS FOUND WITH ERYTHEMA AND INFLAMMATION DURING DRESSING CHANGE. ABOUT 7 DAYS AFTER SURGERY, DARK SKIN TONE AND EPIDERMAL NECROSIS WERE FOUND AROUND THE INCISION SITE. THE PATIENT WAS GIVEN A DRESSING CHANGE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Erythema; Impaired Healing; Inflammation; Necrosis; Skin Inflammation/ Irritation
MAUDE report 18581472 ↗InjuryJan 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON (B)(6) 2023 AND SUTURE WAS USED ON THE SKIN. POST-OP, TWO DAYS AFTER THE SURGERY, THE WOUND WAS FOUND WITH ERYTHEMA AND INFLAMMATION DURING DRESSING CHANGE. ABOUT 7 DAYS AFTER SURGERY, DARK SKIN TONE AND EPIDERMAL NECROSIS WERE FOUND AROUND THE INCISION SITE. THE PATIENT WAS GIVEN A DRESSING CHANGE. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Erythema; Impaired Healing; Inflammation; Necrosis; Skin Inflammation/ Irritation
MAUDE report 18581468 ↗MalfunctionJan 17, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON (B)(6) 2023 AND SUTURE WAS USED. WHEN THE DOCTOR WAS SUTURING, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18527641 ↗MalfunctionDec 9, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNSPECIFIED WOUND CLOSURE PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHILE SEWING THE WOUND IN SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15950204 ↗MalfunctionApr 29, 2020Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE SEPARATED FROM THE SUTURE MID PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10011383 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K032420, K132580, K181652. One submission often covers many models, so check the product and lot details in each record.
- Z-2145-2024 · May 10, 2024 · Open, ClassifiedVICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological
- Z-1310-2009 · Feb 2, 2009 · TerminatedCoated VICRYL Plus Antibacterial (polyglactin 910) Suture; Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.