Coated VICRYL VCPP41DRecalls & reported problems
Ethicon, LLC · Antibacterial Undyed Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
Quick summary
This Ethicon coated Vicryl absorbable suture is a synthetic, dissolvable surgical thread. Among 25 reports, most describe the suture separating from its needle or otherwise coming apart; some mention damaged packaging or an unsterile product, while others describe wound reopening, poor healing, or an abscess. The reports include no deaths, 4 injuries, and 21 malfunctions. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 10705031056848), received between Jul 19, 2021 and Sep 14, 2026 — classified as 4 injuries, 22 malfunctions.
The most frequently reported specific device problem is “Material Separation” (16 reports).
2 other recalls cover devices cleared under the same FDA submission (K032420, K132580, K181652); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation16
- Adverse Event Without Identified Device or Use Problem4
- Delivered as Unsterile Product3
- Break2
- Device Markings/Labelling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions22
- Granuloma2
- Impaired Healing2
- Insufficient Information2
- Wound Dehiscence2
- Abscess1
- Unspecified Infection1
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2026Device Markings/Labelling Problem
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used. It was reported by distributor that historically, the product has been labeled as an undyed Control Release needle suture. Two received boxes have the same product description and reference number as per current inventory; however, there are notable differences in the packaging. The new boxes contain Chinese labeling, and the suture description does not specifically indicate that the suture is undyed. Other info appears to be the same. Device availability unknown. There were no patient consequences reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50753095 ↗MalfunctionJan 8, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNSPECIFIED PROCEDURE, THE NEEDLES WERE POPPING OFF THE SUTURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24004381 ↗MalfunctionJan 8, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNSPECIFIED PROCEDURE, THE SUTURES WERE DETACHING FROM THE NEEDLE. NO PATIENT CONSEQUENCES WERE NOTED, AND THE DEVICE IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24004333 ↗MalfunctionJan 8, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNSPECIFIED PROCEDURE, THE NEEDLES WERE POPPING OFF THE SUTURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24004135 ↗MalfunctionJan 8, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNSPECIFIED PROCEDURE, THE SUTURES WERE DETACHING FROM THE NEEDLE. NO PATIENT CONSEQUENCES WERE NOTED, AND THE DEVICE IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24004070 ↗MalfunctionMay 9, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING AN UNKNOWN PROCEDURE; SUTURE TIP MISSING. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21994784 ↗InjuryMay 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ANTERIOR HIP PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. ABOUT 4 WEEKS AFTER THE SURGERY, BETWEEN DECEMBER AND MARCH, THE PATIENT DEVELOPED A SUTURE ABSCESS AND/OR THE WOUND OPENED UP. THE WOUND WOULD DEHISCE AND WE WOULD HAVE TO OPEN UP THE AREA WASH IT OUT IN THE OR AND RE-CLOSE. THE HCP DOES NOT KNOW IF THIS IS AN ISSUE WITH THE LOT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Abscess; Wound Dehiscence
MAUDE report 21944327 ↗InjuryMay 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ANTERIOR HIP PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. ABOUT 4 WEEKS AFTER THE SURGERY, BETWEEN (B)(6), THE HAD A WOUND DEHISCENCE AND THE PATIENT'S HIP GOT A DEEP INFECTION AND WE HAD TO EXPLANT ALL HER HARDWARE. SHE IS CURRENTLY ON ANTIBIOTICS AND WILL GET REPLANTED LATER. THE HCP DOES NOT KNOW IF THIS IS AN ISSUE WITH THE LOT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Unspecified Infection; Wound Dehiscence
MAUDE report 21944325 ↗MalfunctionJul 8, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT ONE OF THE NURSES BROUGHT SOME SUTURE THEY FOUND THAT HAD HOLES IN THE PACKAGES. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19693108 ↗MalfunctionJul 8, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT ONE OF THE NURSES BROUGHT SOME SUTURE THEY FOUND THAT HAD HOLES IN THE PACKAGES. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19692752 ↗MalfunctionJul 8, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT ONE OF THE NURSES BROUGHT SOME SUTURE THEY FOUND THAT HAD HOLES IN THE PACKAGES. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19692398 ↗MalfunctionOct 16, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TAH/BSO PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE KEEPS DETACHING TOO EASILY DURING CASE. NO PATIENT HARM. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17941332 ↗MalfunctionOct 16, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TAH/BSO PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE KEEPS DETACHING TOO EASILY DURING CASE. NO PATIENT HARM. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17941331 ↗InjuryAug 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL ABDOMINAL HYSTERECTOMY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PATIENT PRESENTED WITH GRANULAR LESIONS ON THE VAGINAL CUFF. THE PATIENT WAS TREATED WITH A SILVER NITRATE DRESSING ON THE VAGINAL CUFF AND IT TOOK 10 WEEKS TO HEAL. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Granuloma; Impaired Healing; Insufficient Information
MAUDE report 17511478 ↗InjuryAug 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL ABDOMINAL HYSTERECTOMY ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PATIENT PRESENTED WITH GRANULAR LESIONS ON THE VAGINAL CUFF. THE PATIENT WAS TREATED WITH A SILVER NITRATE DRESSING ON THE VAGINAL CUFF AND IT TOOK 10 WEEKS TO HEAL. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Granuloma; Impaired Healing; Insufficient Information
MAUDE report 17511472 ↗MalfunctionMar 15, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE GOT DETACHED FROM THE NEEDLE. UNKNOWN AS TO HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13766003 ↗MalfunctionAug 25, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF EIGHT STRANDS WHILE SUTURING THE SUBCUTANEOUS FAT LAYER. ANOTHER BOX OF THE SAME PRODUCT CODE WAS OPENED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12368478 ↗MalfunctionAug 25, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF EIGHT STRANDS WHILE SUTURING THE SUBCUTANEOUS FAT LAYER. ANOTHER BOX OF THE SAME PRODUCT CODE WAS OPENED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12368473 ↗MalfunctionAug 25, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF EIGHT STRANDS WHILE SUTURING THE SUBCUTANEOUS FAT LAYER. ANOTHER BOX OF THE SAME PRODUCT CODE WAS OPENED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12368472 ↗MalfunctionAug 25, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF EIGHT STRANDS WHILE SUTURING THE SUBCUTANEOUS FAT LAYER. ANOTHER BOX OF THE SAME PRODUCT CODE WAS OPENED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12368470 ↗MalfunctionAug 25, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF EIGHT STRANDS WHILE SUTURING THE SUBCUTANEOUS FAT LAYER. ANOTHER BOX OF THE SAME PRODUCT CODE WAS OPENED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12368468 ↗MalfunctionAug 25, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(4) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF EIGHT STRANDS WHILE SUTURING THE SUBCUTANEOUS FAT LAYER. ANOTHER BOX OF THE SAME PRODUCT CODE WAS OPENED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12368466 ↗MalfunctionAug 25, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF EIGHT STRANDS WHILE SUTURING THE SUBCUTANEOUS FAT LAYER. ANOTHER BOX OF THE SAME PRODUCT CODE WAS OPENED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12368464 ↗MalfunctionAug 25, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE POPPED OFF EIGHT STRANDS WHILE SUTURING THE SUBCUTANEOUS FAT LAYER. ANOTHER BOX OF THE SAME PRODUCT CODE WAS OPENED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12368461 ↗MalfunctionJul 27, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE POP OFF CAME OFF TOO EASILY. THE SUTURE AND THE NEEDLE BECOME SEPARATED. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12229668 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K032420, K132580, K181652. One submission often covers many models, so check the product and lot details in each record.
- Z-2145-2024 · May 10, 2024 · Open, ClassifiedVICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological
- Z-1310-2009 · Feb 2, 2009 · TerminatedCoated VICRYL Plus Antibacterial (polyglactin 910) Suture; Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.