The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705031058866), received between Oct 26, 2021 and May 30, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K960653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Integrity Problem2
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 30, 2023Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A OB-GYN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN OB-GYN SURGERY, AFTER OPENING THE PACKAGE AND BEFORE USE, IT WAS FOUND THAT THE SUTURE HAD BEEN DETACHED FROM THE NEEDLE. IT WAS NOT A CONTROL RELEASE NEEDLE. VISUAL INSPECTION EVALUATION WAS CONDUCTED ON THE RETURNED DEVICE. THE RETURNED SAMPLE REVEALED THAT IT WAS RECEIVED ONE LABELED WINDING FORMER WITH SOME SUTURE PIECES IN THE DEGRADATION PROCESS AND ONE DETACHED NEEDLE THAT PERTAINS TO THE PRODUCT CODE Y340H. AS PER VISUAL INSPECTION OF THE SAMPLE, IT WAS OBSERVED THAT THE STRAND HAS BEGUN WITH A DEGRADATION PROCESS CAUSED BY EXPOSURE TO THE ENVIRONMENT. THIS CONDITION CONTRIBUTES TO THE NEEDLE SEPARATING FROM THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17021186 ↗
  • MalfunctionJun 1, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, WHEN THE PACKAGE WAS OPENED, THE SUTURE WAS FOUND TO HAVE BEEN TORN INTO SEVERAL MILLIMETERS BEFORE USE. UPON EVALUATION OF THE RETURNED SAMPLE, THE FOIL WAS VISUALLY INSPECTED, AND EXCESSIVE WRINKLES AND HOLES IN THE CAVITY WERE OBSERVED. THIS CONDITION CONTRIBUTED TO THE DEGRADATION OF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14570841 ↗
  • MalfunctionOct 26, 2021Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS FOUND THAT THE SUTURE HAD BEEN DETACHED FROM THE NEEDLE BEFORE USE. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12698454 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K960653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2148-2024 · May 10, 2024 · Open, Classified
    MONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated PRODUCT CODE: Y315H Y345
  • Z-1311-2009 · Feb 2, 2009 · Terminated
    MONOCRYL (poliglecaprone 25 ) Suture, Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and/or ligation.

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page