MONOCRYL Y463GRecalls & reported problems

Ethicon, LLC · Violet Monofilament (Poliglecaprone 25), Absorbable Surgical Suture · Class 2 · Implant

Quick summary

This Ethicon Monocryl absorbable surgical suture is a violet monofilament thread used to close tissue during surgery. The 50 reports mostly describe the suture breaking during use or while opening the box, and material separating into pieces, with no patient harm mentioned in any of them. There are no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 10705031059252), received between Dec 12, 2017 and Feb 24, 2026 — classified as 50 malfunctions.

The most frequently reported specific device problem is “Material Separation” (23 reports).

2 other recalls cover devices cleared under the same FDA submission (K960653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 12 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 4 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 4 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation23
  • Break21
  • Detachment of Device or Device Component2
  • Incorrect Device Or Component Shipped2
  • Detachment Of Device Component1
  • Device Contaminated During Manufacture or Shipping1
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions40
  • No Consequences Or Impact To Patient10

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY HEALTHCARE PROFESSIONAL THAT THE ACCOUNT HAS AN ISSUE WITH THE SUTURE. THE FACILITY HAD ORDERED NINETEEN BOXES AND THE ISSUE IS WITH FOUR (4) BOXES BREAKING DURING MISCLOSURE. THERE WAS NO REPORT OF HARM TO ANYONE. DEVICE AVAILABILITY IS UNKNOWN. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24431475 ↗
  • MalfunctionFeb 24, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY HEALTHCARE PROFESSIONAL THAT THE ACCOUNT HAS AN ISSUE WITH THE SUTURE. THE FACILITY HAD ORDERED NINETEEN BOXES AND THE ISSUE IS WITH FOUR (4) BOXES BREAKING DURING MISCLOSURE. THERE WAS NO REPORT OF HARM TO ANYONE. DEVICE AVAILABILITY IS UNKNOWN. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24431474 ↗
  • MalfunctionFeb 24, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY HEALTHCARE PROFESSIONAL THAT THE ACCOUNT HAS AN ISSUE WITH THE SUTURE. THE FACILITY HAD ORDERED NINETEEN BOXES AND THE ISSUE IS WITH FOUR (4) BOXES BREAKING DURING MISCLOSURE. THERE WAS NO REPORT OF HARM TO ANYONE. DEVICE AVAILABILITY IS UNKNOWN. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24431466 ↗
  • MalfunctionFeb 24, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY HEALTHCARE PROFESSIONAL THAT THE ACCOUNT HAS AN ISSUE WITH THE SUTURE. THE FACILITY HAD ORDERED (B)(4) BOXES AND THE ISSUE IS WITH FOUR (4) BOXES BREAKING DURING MISCLOSURE. THERE WAS NO REPORT OF HARM TO ANYONE. DEVICE AVAILABILITY IS UNKNOWN. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24431458 ↗
  • MalfunctionJan 16, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE SUTURE WAS BROKEN EASILY DURING USE. WHEN PUT OUT THREE OR FOUR OF THE SUTURES, THEY WERE IMMEDIATELY BROKEN. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21172615 ↗
  • MalfunctionJan 16, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE SUTURE WAS BROKEN EASILY DURING USE. WHEN PUT OUT THREE OR FOUR OF THE SUTURES, THEY WERE IMMEDIATELY BROKEN. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21172592 ↗
  • MalfunctionJan 16, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE SUTURE WAS BROKEN EASILY DURING USE. WHEN PUT OUT THREE OR FOUR OF THE SUTURES, THEY WERE IMMEDIATELY BROKEN. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21172531 ↗
  • MalfunctionJan 16, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE SUTURE WAS BROKEN EASILY DURING USE. WHEN PUT OUT THREE OR FOUR OF THE SUTURES, THEY WERE IMMEDIATELY BROKEN. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21172508 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483514 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483511 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483507 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483458 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483457 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483449 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483309 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483307 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483305 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483301 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483299 ↗
  • MalfunctionOct 18, 2024Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN ORAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL, THAT THEY HAD SEVERAL OF THE NEEDLES COME OFF THE SUTURE FROM 1 BOX OF 5/0 MONOCRYL. THEY USED A DIFFERENT BOX TODAY AND DID NOT HAVE ANY ISSUES SO FAR. UNKNOWN IF THE DEVICE WAS RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20483298 ↗
  • MalfunctionApr 3, 2024Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT PERIODONTAL SURGERY ON AN UNKNOWN DATE IN 2024, AND SUTURE WAS USED. THE SUTURE BROKE ON WHEN TYING THE KNOT AND/OR TIGHTENING THE SUTURE ON THE TISSUE TO BRING TISSUE EDGES TOGETHER. A NEW PACKAGE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19035404 ↗
  • MalfunctionApr 2, 2024Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT PERIODONTAL SURGERY ON AN UNKNOWN DATE IN 2024, AND SUTURE WAS USED. THE SUTURE BROKE ON WHEN TYING THE KNOT AND/OR TIGHTENING THE SUTURE ON THE TISSUE TO BRING TISSUE EDGES TOGETHER. A NEW PACKAGE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19026017 ↗
  • MalfunctionApr 2, 2024Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT PERIODONTAL SURGERY ON AN UNKNOWN DATE IN 2024, AND SUTURE WAS USED. THE SUTURE BROKE ON WHEN TYING THE KNOT AND/OR TIGHTENING THE SUTURE ON THE TISSUE TO BRING TISSUE EDGES TOGETHER. A NEW PACKAGE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19021944 ↗
  • MalfunctionApr 2, 2024Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT PERIODONTAL SURGERY ON AN UNKNOWN DATE IN 2024, AND SUTURE WAS USED. THE SUTURE BROKE ON WHEN TYING THE KNOT AND/OR TIGHTENING THE SUTURE ON THE TISSUE TO BRING TISSUE EDGES TOGETHER. A NEW PACKAGE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19021733 ↗
  • MalfunctionApr 2, 2024Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT PERIODONTAL SURGERY ON AN UNKNOWN DATE IN 2024, AND SUTURE WAS USED. THE SUTURE BROKE ON WHEN TYING THE KNOT AND/OR TIGHTENING THE SUTURE ON THE TISSUE TO BRING TISSUE EDGES TOGETHER. A NEW PACKAGE WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19021545 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K960653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2148-2024 · May 10, 2024 · Open, Classified
    MONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated PRODUCT CODE: Y315H Y345
  • Z-1311-2009 · Feb 2, 2009 · Terminated
    MONOCRYL (poliglecaprone 25 ) Suture, Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and/or ligation.

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →