MONOCRYL Y604HRecalls & reported problems

Ethicon, LLC · Violet Monofilament (Poliglecaprone 25), Absorbable Surgical Suture · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10705031059474), received between Jul 27, 2020 and Nov 20, 2025 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

2 other recalls cover devices cleared under the same FDA submission (K960653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Material Separation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient2
  • Failure to Anastomose1
  • Insufficient Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC OOPHORECTOMY FOR OVARIAN LESIONS ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE SURGERY, WHEN THE SURGEON USED SUTURE TO SUTURE THE SKIN, THE SUTURE WAS BROKEN. AFTER REPLACING THE SUTURE WITH A NEW ONE, THE SUTURING OPERATION CONTINUED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23605901 ↗
  • MalfunctionSep 3, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PROSTATE CANCER SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED WHEN THE DOCTOR USED SUTURE TO SUTURE AND ANASTOMOSE THE URETHRA DURING SURGERY, THE SUTURE BROKE WHEN TIGHTENED, RESULTING IN A PROLONGED ANASTOMOSIS TIME AND POSING A SAFETY HAZARD TO THE ANASTOMOSIS EFFECT. REPLACE THE SUTURE WITH A NEW ONE AND CONTINUE WITH SURGICAL SUTURING, PROCEEDING NORMALLY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Failure to Anastomose; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20129342 ↗
  • MalfunctionOct 26, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) /2023 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN THE DOCTOR WAS USING SUTURE FOR THE PATIENT, THE DOCTOR FOUND THAT THE NEEDLE AND SUTURE WERE BROKEN. THE DOCTOR IMMEDIATELY TOOK THEM OFF THE TABLE AND KEPT THEM, AND REMOVED A NEW ONE, DELAYING THE PROCESS BY ABOUT 5 MINUTES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 18016613 ↗
  • MalfunctionJun 21, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, AT 12:10, WHEN THE SURGERY WAS COMPLETED AND THE SKIN WAS READY TO BE SUTURED, SUTURE WERE PUT ON THE STAGE, THE NURSE FOUND THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAD HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17173020 ↗
  • MalfunctionJan 25, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ENDOSCOPIC PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAD HAPPENED IN THE NEWLY DISPENSED SUTURE WHEN IT WAS OPENED TO PREPARE FOR UROLOGICAL ENDOSCOPIC SURGERY. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE SURGERY. UPON THE EVALUATION OF THE RETURNED SAMPLE, THE VISUAL INSPECTION CONCLUDED THAT THE COMBINATION OF THE NEEDLE/SUTURE WAS FOUND DETACHED SINCE THE INSERTION END OF THE SUTURE DID NOT MEET THE REQUIREMENT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13343651 ↗
  • MalfunctionDec 21, 2020Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT LAPAROSCOPIC APPENDECTOMY ON (B)(6) 2020 AND SUTURE WAS USED. THE SUTURE BROKE WHEN SEWING THE INCISION AT THE TIME OF KNOTTING IN THE SURGERY. CHANGED TO ANOTHER DEVICE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11051796 ↗
  • MalfunctionJul 27, 2020Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC RADICAL PROSTATECTOMY PROCEDURE ON (B)(6) 2020; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE WHILE KNOTTING. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE SURGERY WAS DELAYED NEARLY ONE HOUR. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10326226 ↗
  • MalfunctionJul 27, 2020Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC RADICAL PROSTATECTOMY PROCEDURE ON (B)(6) 2020; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE WHILE KNOTTING. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE SURGERY WAS DELAYED NEARLY ONE HOUR. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10326223 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K960653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2148-2024 · May 10, 2024 · Open, Classified
    MONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated PRODUCT CODE: Y315H Y345
  • Z-1311-2009 · Feb 2, 2009 · Terminated
    MONOCRYL (poliglecaprone 25 ) Suture, Ethicon Inc., A Johnson & Johnson Company, Somerville, NJ Sutures are intended for general soft tissue approximation and/or ligation.