RELIEVA FLEX GC090RFRecalls & reported problems
Acclarent, Inc. · RELIEVA FLEX SINUS GUIDE CATHETER M-90 · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705031063457), received between May 22, 2018 and May 22, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment Of Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment Of Device Component2
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2018Detachment Of Device Component; Detachment Of Device Component
ACCLARENT WAS INFORMED ON (B)(6) 2018, OF AN EVENT WHICH OCCURRED DURING THE PRIMARY HYBRID BALLOON SINUPLASTY PROCEDURE WITH TURBINATE REDUCTION AND PARTIAL ETHMOIDECTOMY ON (B)(6) 2018, IN WHICH A RELIEVA® FLEX SINUS GUIDE CATHETER M-90 (GC090RF / 170124C-PC) WAS USED. THE HEALTHCARE PROFESSIONAL REPORTED THAT THE SPACE BETWEEN THE ETHMOID BULLA AND THE UNCINATE PROCESS WAS TIGHT, WHICH MADE IT DIFFICULT TO ACCESS THE SINUS. THE BLUE TIP ON THE FLEX GUIDE CATHETER BECAME DETACHED AND FELL OFF WITHIN THE MIDDLE MEATUS AFTER THE MAXILLARY SINUS WAS DILATED. THE TIP WAS RETRIEVED AND DISCARDED. THE GUIDE CATHETER HAD BEEN USED THREE TIMES BEFORE THE BLUE TIP BECAME SEPARATED. THE PROCEDURE WAS CONTINUED WITHOUT THE USE OF THE TIP; THE PROCEDURE WAS COMPLETED WITHOUT FURTHER ISSUE AND WITHOUT ANY PATIENT CONSEQUENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7532061 ↗MalfunctionMay 22, 2018Detachment Of Device Component; Detachment Of Device Component
ACCLARENT WAS INFORMED ON (B)(6) 2018, OF AN EVENT WHICH OCCURRED DURING THE PRIMARY HYBRID BALLOON SINUPLASTY PROCEDURE WITH TURBINATE REDUCTION AND PARTIAL ETHMOIDECTOMY ON (B)(6) 2018, IN WHICH A RELIEVA® FLEX SINUS GUIDE CATHETER M-90 (GC090RF / 170124C-PC) WAS USED. THE HEALTHCARE PROFESSIONAL REPORTED THAT THE SPACE BETWEEN THE ETHMOID BULLA AND THE UNCINATE PROCESS WAS TIGHT, WHICH MADE IT DIFFICULT TO ACCESS THE SINUS. THE BLUE TIP ON THE FLEX GUIDE CATHETER BECAME DETACHED AND FELL OFF AFTER THE MAXILLARY SINUS WAS DILATED. THE TIP FELL OFF OUTSIDE OF THE PATIENT¿S ANATOMY; IT WAS DISCARDED. THE GUIDE CATHETER HAD BEEN USED THREE TIMES BEFORE THE BLUE TIP BECAME SEPARATED. THE PROCEDURE WAS CONTINUED WITHOUT THE USE OF THE TIP; THE PROCEDURE WAS COMPLETED WITHOUT FURTHER ISSUE AND WITHOUT ANY PATIENT CONSEQUENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7531931 ↗
Other CANNULA, SINUS devices
- RELIEVA FLEX GC070CRFAcclarent, Inc. · CANNULA, SINUS1 report
- RELIEVA FLEX GC110RFAcclarent, Inc. · CANNULA, SINUS1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page