ACCLARENT BALLOON INFLATION DEVICE BID30Recalls & reported problems

Acclarent, Inc. · ACCLARENT INFLATION DEVICE · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10705031063570), received between Mar 22, 2024 and May 15, 2025 — classified as 1 injury, 15 malfunctions.

The most frequently reported specific device problem is “Contamination of Device Ingredient or Reagent” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 12 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination of Device Ingredient or Reagent15
  • Insufficient Device Problem Information1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • Perforation of Vessels1
  • Stroke/CVA1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2025Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, WHEN THE PACKAGING OF AN ACCLARENT BALLOON INFLATION DEVICE (BID30, LOT UNKNOWN) WAS OPENED, IT HAD A RESIDUE ON IT (AS SEEN IN THE PHOTOS). NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22029199 ↗
  • MalfunctionDec 26, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A HYBRID BALLOON SINUPLASTY PROCEDURE (BSP), WHEN THE ACCLARENT BALLOON INFLATION DEVICE (BID30 / 96375267) WAS REMOVED FROM ITS PACKAGING, A WHITE SUBSTANCE WAS NOTICED ON THE HANDLE OF THE DEVICE. THE REPORTER STATED THAT ¿THE WHITE SUBSTANCE MAY HAVE BEEN PAPER FROM THE LABEL RUBBING OFF.¿ THE ACCLARENT BALLOON INFLATION DEVICE WAS REPLACED WITH ANOTHER FROM THE SAME LOT (96375267), BUT THE SAME WHITE SUBSTANCE WAS NOTICED ON THE HANDLE. THE DEVICE WAS REPLACED A THIRD TIME, AND THE THIRD DEVICE HAD NO WHITE SUBSTANCE ON IT. THE PROCEDURE WAS RESUMED. THERE WAS NO NEGATIVE PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21022653 ↗
  • MalfunctionDec 26, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING A HYBRID BALLOON SINUPLASTY PROCEDURE (BSP), WHEN THE ACCLARENT BALLOON INFLATION DEVICE (BID30 / 96375267) WAS REMOVED FROM ITS PACKAGING, A WHITE SUBSTANCE WAS NOTICED ON THE HANDLE OF THE DEVICE. THE REPORTER STATED THAT ¿THE WHITE SUBSTANCE MAY HAVE BEEN PAPER FROM THE LABEL RUBBING OFF.¿ THE ACCLARENT BALLOON INFLATION DEVICE WAS REPLACED WITH ANOTHER FROM THE SAME LOT (96375267), BUT THE SAME WHITE SUBSTANCE WAS NOTICED ON THE HANDLE. THE DEVICE WAS REPLACED A THIRD TIME, AND THE THIRD DEVICE HAD NO WHITE SUBSTANCE ON IT. THE PROCEDURE WAS RESUMED. THERE WAS NO NEGATIVE PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21022652 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, 96375262) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19229031 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, 96375262) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19229026 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, 96375262) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228998 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, 96375262) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228980 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, 96375262) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228970 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, 96375262) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228954 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, 96375262) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228949 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, 96375262) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228938 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, (B)(6)) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228926 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES(B)(6) (SEID, (B)(6) ) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228896 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, (B)(6)) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON (B)(6)2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228890 ↗
  • MalfunctionMay 2, 2024Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, INSIDE THE STERILE PACKAGE OF 15 ACCLARENT SE INFLATION DEVICES (SEID, 96375262) WHITE RESIDUE WAS PRESENT. ADDITIONAL EVENT INFORMATION RECEIVED ON 25-APR-2025 FURTHER CLARIFIED THAT THE ¿WHITE RESIDUE¿ IS BEST DESCRIBED AS ¿ALMOST MOISTURE IN THE PACKAGE¿. THERE WERE NO PATIENT CONSEQUENCES AS THE DEVICES WERE NOT CLINICALLY USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228844 ↗
  • InjuryMar 22, 2024Insufficient Device Problem Information; Patient Device Interaction Problem; Insufficient Device Problem Information; Patient Device Interaction Problem

    IT WAS REPORTED THAT ON (B)(6) 2022, DURING A BALLOON SINUPLASTY (BSP) PROCEDURE, THE PATIENT SUSTAINED AN INJURY TO HER CAROTID ARTERY THAT RESULTED IN HER SUFFERING A CATASTROPHIC STROKE. THE REPORTED DEVICES USED DURING THE PROCEDURE WERE: A RELIEVE TRACT BALLOON (RT1640A / 211202A-PC), TWO 6MM X 16MM RELIEVA SPINPLUS NAVIGATION BALLOON SINUPLASTY SYSTEM (RSP0616MFSN), ONE FROM LOT 220419A-PC AND ONE FROM LOT 220504A-PC, A DISPOSABLE TRUDI PROBE 0-DEGREE, 5IN (TDP0005 / 210201B-PC), AN ACCLARENT BALLOON INFLATION DEVICE (BID30 / 96375250), AND A TRUDI NAV CABLE (TDNC001 / 211206C-PC) AND THE TRUDI NAVIGATION SYSTEM. IT IS REPORTED THAT THE PATIENT BELIEVED THAT THE ACCLARENT NAVIGATION BALLOON DILATION SYSTEM AND ITS COMPONENT PARTS, INCLUDING ANY AI (ARTIFICIAL INTELLIGENCE SOFTWARE) MAY HAVE MALFUNCTIONED AND FAILED TO PERFORM AS INTENDED DURING THE PROCEDURE, AND THAT AS A RESULT, THE PATIENT¿S CAROTID ARTERY WAS INJURED.

    Patient problems: Perforation of Vessels; Stroke/CVA

    MAUDE report 18962842 ↗

Other INSTRUMENT, ENT MANUAL SURGICAL devices

All INSTRUMENT, ENT MANUAL SURGICAL devices →