Skip to content

PROLENE EH7222HRecalls & reported problems

Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant

All models of PROLENE (356)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Prolene suture is a nonabsorbable blue polypropylene surgical suture. Reports mostly describe the suture material separating or breaking during surgery, including needle detachment, extra needles, or foreign bodies in the packet; no patient deaths or injuries are reported. There are 40 reports, no deaths, no injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 10705031073043), received between Jun 24, 2024 and May 11, 2026 — classified as 40 malfunctions.

The most frequently reported specific device problem is “Material Separation” (25 reports).

21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 7 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 6 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation25
  • Break13
  • Appropriate Device Problem Term/Code Not Available1
  • Component Missing1
  • Delivered as Unsterile Product1
  • Device Contaminated During Manufacture or Shipping1
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions38
  • Insufficient Information1
  • Pain1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 11, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE FREQUENTLY BROKE DURING BLOOD VESSEL ANASTOMOSIS, AND THE SUTURE WAS REPLACED FOR ANASTOMOSIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25142446 ↗
  • MalfunctionMay 11, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6)2026 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE FREQUENTLY BROKE DURING BLOOD VESSEL ANASTOMOSIS, AND THE SUTURE WAS REPLACED FOR ANASTOMOSIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25142384 ↗
  • MalfunctionMay 4, 2026Product Quality Problem; Material Separation; Component Missing

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IDENTIFIED AN ADDITIONAL NEEDLE PRESENT TO THE SIDE OF THE PACKET 2 X NEEDLES ATTACHED TO THE THREAD 1 X NEEDLE LOOSE IN THE PACKET FIRST USE THE NEEDLE POPPED OFF WITHOUT ANY ADDITIONAL TENSION. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25073557 ↗
  • MalfunctionApr 17, 2026Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. AT 11:38 AM, WHEN THE OPERATING ROOM TOOK THE CONSUMABLES FOR INSPECTION, IT WAS FOUND THAT THERE WAS A FOREIGN BODY IN THE SUTURE. THE THIRD OF THE ORIGINAL TWO APPEARED AND IT WAS A BROKEN HOOK AND WAS NOT USED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24916191 ↗
  • MalfunctionMar 24, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE DOCTOR PERFORMED A VASCULAR FISTULA FOR THE PATIENT. WHEN SUTURING WAS REQUIRED, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED AND COULD NOT BE USED NORMALLY. A NEW SET OF SUTURE WAS OPENED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24674789 ↗
  • MalfunctionFeb 26, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN RIGHT UPPER LIMB ARTERIOVENOUS FISTULA SURGERY+VENOUS ARTERIALIZATION+ARTERIOVENOUS EXPLORATION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE PATIENT WAS ADMITTED DUE TO "FATIGUE AND POOR APPETITE FOR 5 MONTHS, WORSENING FOR 2 MONTHS", DIAGNOSED WITH END-STAGE RENAL DISEASE. DUE TO THE CONDITION REQUIRING THE ESTABLISHMENT OF A LONG-TERM HEMODIALYSIS VASCULAR PATHWAY, SURGERY WAS PERFORMED ON THE 9TH DAY AFTER ADMISSION. THE DOCTOR USED SUTURE TO SUTURE THE PATIENT'S BLOOD VESSELS DURING THE OPERATION, BUT THE SUTURE BROKE ABOUT 5CM AWAY FROM THE NEEDLE DURING THE SUTURING PROCESS. THE SUTURE IS A DOUBLE NEEDLE SUTURE, AND THE DOCTOR CONTINUED TO SUTURE THE BLOOD VESSEL WITH THE OTHER END OF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24461987 ↗
  • MalfunctionDec 10, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. NEEDLE DETACHED FROM SUTURE WITHOUT ANY EXTRA FORCE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23770556 ↗
  • MalfunctionDec 10, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. NEEDLE DETACHED FROM SUTURE WITHOUT ANY EXTRA FORCE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23770555 ↗
  • MalfunctionNov 7, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN AUTOLOGOUS ARTERIOVENOUS FISTULA RECONSTRUCTION OF LEFT FOREARM PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, WHILE USING SUTURE TO LIGATE BLOOD VESSELS DURING AUTOLOGOUS ARTERIOVENOUS FISTULA RECONSTRUCTION OF LEFT FOREARM SURGERY, THE SUTURE BROKE TWICE IN A ROW. REPLACED WITH A NEW SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23501265 ↗
  • MalfunctionNov 5, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN AUTOGENOUS ARTERIOVENOUS FISTULA RECONSTRUCTION OF RIGHT FOREARM PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN THE DOCTOR USED SUTURE TO LIGATE THE BLOOD VESSEL DURING SURGERY, THE SUTURE BROKE TWICE IN A ROW. REPLACED WITH A NEW ONE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23478738 ↗
  • MalfunctionOct 21, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23348770 ↗
  • MalfunctionOct 21, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23348762 ↗
  • MalfunctionAug 29, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22923574 ↗
  • MalfunctionAug 20, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE HAPPENED IN THE SURGERY. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WILL RETURN 15 REPRESENTATIVE SAMPLES FOR ANALYSIS. THERE WAS NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22850275 ↗
  • MalfunctionAug 20, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE HAPPENED IN THE SURGERY. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WILL RETURN 15 REPRESENTATIVE SAMPLES FOR ANALYSIS. THERE WAS NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22850268 ↗
  • MalfunctionAug 20, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE HAPPENED IN THE SURGERY. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WILL RETURN 15 REPRESENTATIVE SAMPLES FOR ANALYSIS. THERE WAS NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22850261 ↗
  • MalfunctionAug 20, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE HAPPENED IN THE SURGERY. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WILL RETURN 15 REPRESENTATIVE SAMPLES FOR ANALYSIS. THERE WAS NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22850242 ↗
  • MalfunctionAug 18, 2025Material Separation; Appropriate Device Problem Term/Code Not Available; Material Separation; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED DURING INTRAOPERATIVE USE. THE PRODUCT WAS REMOVED PROMPTLY FROM THE SURGICAL AREA WITHOUT CAUSING ANY FOREIGN BODY RESIDUE. A NEW SUTURE WAS REPLACED AND THE OPERATION CONTINUED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22829962 ↗
  • MalfunctionJun 13, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22206316 ↗
  • MalfunctionMay 19, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, AT 3:30 PM, WHILE SUTURING THE FISTULA WOUND FOR THE PATIENT, THE SUTURE BROKE AND WAS REPLACED WITH A NEW ONE. THIS EVENT RESULTED IN AN EXTENSION OF THE WOUND SUTURING TIME. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 22048013 ↗
  • MalfunctionMay 12, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22005347 ↗
  • MalfunctionMar 17, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT A PATIENT UNDERWENT A THORACIC SURGERY PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE DEVICE PACKAGE WAS FOUND UNSEALED WHEN UNPACKING AND COUNTING. ONE END OF THE PACKAGE IS NOT SEALED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21625221 ↗
  • MalfunctionJan 7, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2024 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THE SURGERY. IT CAUSED MALFUNCTIONS DURING THE ANASTOMOSIS OF THE HEART AND BLOOD VESSELS, DELAYED THE TIMING OF THE SURGERY AND CAUSING WASTE OF CONSUMABLES. STOPPED USING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21096630 ↗
  • MalfunctionJan 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN CARDIOVASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL BE 25 DEVICES RETURNED FOR ANALYSIS, INCLUDING 2 ACTUAL PRODUCTS AND 23 STERILE PRODUCTS FROM SAME BATCH. THE SURGEON REFUSED TO USE THE REMAINING 23 STERILE PRODUCTS FROM THE SAME LOT AND THEY WILL BE RETURNED FOR ANALYSIS. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21068306 ↗
  • MalfunctionJan 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN CARDIOVASCULAR PROCEDURE ON (B)(6) 2024. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL BE 25 DEVICES RETURNED FOR ANALYSIS, INCLUDING 2 ACTUAL PRODUCTS AND 23 STERILE PRODUCTS FROM SAME BATCH. THE SURGEON REFUSED TO USE THE REMAINING 23 STERILE PRODUCTS FROM THE SAME LOT AND THEY WILL BE RETURNED FOR ANALYSIS. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21068275 ↗

FDA clearance and classification

Cleared through Special 510(k) K133356 on Dec 6, 2013 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Prolene Polypropylene Nonabsorbable Suture”. The FDA review took 36 days. FDA record ↗

The FDA regulates product code GAW (suture, nonabsorbable, synthetic, polypropylene) under 21 CFR 878.5010 as a Class II device. 21 CFR 878.5010 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.

  • Z-1515-2025 · Mar 7, 2025 · Open, Classified
    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
  • Z-1096-2025 · Dec 20, 2024 · Open, Classified
    PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
  • Z-2452-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
  • Z-2460-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
  • Z-2455-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
  • Z-2449-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
  • Z-2466-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE 122CM M1.5, D10189
  • Z-2462-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
  • Z-2465-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
  • Z-2454-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
  • Z-2456-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
  • Z-2464-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
  • Z-2458-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
  • Z-2457-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
  • Z-2463-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995

…and 6 more.

Common questions about the PROLENE EH7222H

Has the PROLENE EH7222H been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705031073043), based on the FDA recall file dated Nov 27, 2026. However, 21 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the PROLENE EH7222H?

The FDA's MAUDE database lists 40 adverse event reports for this device, received between Jun 24, 2024 and May 11, 2026. In those reports, the most common specific device problems are “Material Separation” (25), “Break” (13) and “Component Missing” (1); the most common patient problem is “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PROLENE EH7222H?

None of the reports are classified as deaths, 0 as injuries and 40 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PROLENE EH7222H FDA approved?

It was cleared through Special 510(k) K133356 on Dec 6, 2013. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the PROLENE EH7222H MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.

Who makes the PROLENE EH7222H, and is it still sold?

The FDA's device database lists Ethicon, LLC as the company responsible for it (brand PROLENE, model EH7222H). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Suture, nonabsorbable, synthetic, polypropylene devices

All Suture, nonabsorbable, synthetic, polypropylene devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page