PROLENE EH7222HRecalls & reported problems
Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant
Quick summary
This Ethicon Prolene suture is a nonabsorbable blue polypropylene surgical suture. Reports mostly describe the suture material separating or breaking during surgery, including needle detachment, extra needles, or foreign bodies in the packet; no patient deaths or injuries are reported. There are 40 reports, no deaths, no injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 41
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 10705031073043), received between Jun 24, 2024 and Sep 14, 2026 — classified as 41 malfunctions.
The most frequently reported specific device problem is “Material Separation” (26 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation26
- Break13
- Appropriate Device Problem Term/Code Not Available1
- Component Missing1
- Delivered as Unsterile Product1
- Device Contaminated During Manufacture or Shipping1
- Product Quality Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions39
- Insufficient Information1
- Pain1
Recent adverse event reports
The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50762654 ↗MalfunctionMay 11, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE FREQUENTLY BROKE DURING BLOOD VESSEL ANASTOMOSIS, AND THE SUTURE WAS REPLACED FOR ANASTOMOSIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25142446 ↗MalfunctionMay 11, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6)2026 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE FREQUENTLY BROKE DURING BLOOD VESSEL ANASTOMOSIS, AND THE SUTURE WAS REPLACED FOR ANASTOMOSIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25142384 ↗MalfunctionMay 4, 2026Product Quality Problem; Material Separation; Component Missing
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IDENTIFIED AN ADDITIONAL NEEDLE PRESENT TO THE SIDE OF THE PACKET 2 X NEEDLES ATTACHED TO THE THREAD 1 X NEEDLE LOOSE IN THE PACKET FIRST USE THE NEEDLE POPPED OFF WITHOUT ANY ADDITIONAL TENSION. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25073557 ↗MalfunctionApr 17, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. AT 11:38 AM, WHEN THE OPERATING ROOM TOOK THE CONSUMABLES FOR INSPECTION, IT WAS FOUND THAT THERE WAS A FOREIGN BODY IN THE SUTURE. THE THIRD OF THE ORIGINAL TWO APPEARED AND IT WAS A BROKEN HOOK AND WAS NOT USED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24916191 ↗MalfunctionMar 24, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE DOCTOR PERFORMED A VASCULAR FISTULA FOR THE PATIENT. WHEN SUTURING WAS REQUIRED, THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED AND COULD NOT BE USED NORMALLY. A NEW SET OF SUTURE WAS OPENED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24674789 ↗MalfunctionFeb 26, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN RIGHT UPPER LIMB ARTERIOVENOUS FISTULA SURGERY+VENOUS ARTERIALIZATION+ARTERIOVENOUS EXPLORATION PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE PATIENT WAS ADMITTED DUE TO "FATIGUE AND POOR APPETITE FOR 5 MONTHS, WORSENING FOR 2 MONTHS", DIAGNOSED WITH END-STAGE RENAL DISEASE. DUE TO THE CONDITION REQUIRING THE ESTABLISHMENT OF A LONG-TERM HEMODIALYSIS VASCULAR PATHWAY, SURGERY WAS PERFORMED ON THE 9TH DAY AFTER ADMISSION. THE DOCTOR USED SUTURE TO SUTURE THE PATIENT'S BLOOD VESSELS DURING THE OPERATION, BUT THE SUTURE BROKE ABOUT 5CM AWAY FROM THE NEEDLE DURING THE SUTURING PROCESS. THE SUTURE IS A DOUBLE NEEDLE SUTURE, AND THE DOCTOR CONTINUED TO SUTURE THE BLOOD VESSEL WITH THE OTHER END OF THE SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24461987 ↗MalfunctionDec 10, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. NEEDLE DETACHED FROM SUTURE WITHOUT ANY EXTRA FORCE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23770556 ↗MalfunctionDec 10, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. NEEDLE DETACHED FROM SUTURE WITHOUT ANY EXTRA FORCE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23770555 ↗MalfunctionNov 7, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN AUTOLOGOUS ARTERIOVENOUS FISTULA RECONSTRUCTION OF LEFT FOREARM PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, WHILE USING SUTURE TO LIGATE BLOOD VESSELS DURING AUTOLOGOUS ARTERIOVENOUS FISTULA RECONSTRUCTION OF LEFT FOREARM SURGERY, THE SUTURE BROKE TWICE IN A ROW. REPLACED WITH A NEW SUTURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23501265 ↗MalfunctionNov 5, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN AUTOGENOUS ARTERIOVENOUS FISTULA RECONSTRUCTION OF RIGHT FOREARM PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN THE DOCTOR USED SUTURE TO LIGATE THE BLOOD VESSEL DURING SURGERY, THE SUTURE BROKE TWICE IN A ROW. REPLACED WITH A NEW ONE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23478738 ↗MalfunctionOct 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23348770 ↗MalfunctionOct 21, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23348762 ↗MalfunctionAug 29, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22923574 ↗MalfunctionAug 20, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE HAPPENED IN THE SURGERY. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WILL RETURN 15 REPRESENTATIVE SAMPLES FOR ANALYSIS. THERE WAS NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22850275 ↗MalfunctionAug 20, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE HAPPENED IN THE SURGERY. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WILL RETURN 15 REPRESENTATIVE SAMPLES FOR ANALYSIS. THERE WAS NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22850268 ↗MalfunctionAug 20, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE HAPPENED IN THE SURGERY. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WILL RETURN 15 REPRESENTATIVE SAMPLES FOR ANALYSIS. THERE WAS NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22850261 ↗MalfunctionAug 20, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE HAPPENED IN THE SURGERY. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WILL RETURN 15 REPRESENTATIVE SAMPLES FOR ANALYSIS. THERE WAS NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22850242 ↗MalfunctionAug 18, 2025Material Separation; Appropriate Device Problem Term/Code Not Available; Material Separation; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED DURING INTRAOPERATIVE USE. THE PRODUCT WAS REMOVED PROMPTLY FROM THE SURGICAL AREA WITHOUT CAUSING ANY FOREIGN BODY RESIDUE. A NEW SUTURE WAS REPLACED AND THE OPERATION CONTINUED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22829962 ↗MalfunctionJun 13, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22206316 ↗MalfunctionMay 19, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, AT 3:30 PM, WHILE SUTURING THE FISTULA WOUND FOR THE PATIENT, THE SUTURE BROKE AND WAS REPLACED WITH A NEW ONE. THIS EVENT RESULTED IN AN EXTENSION OF THE WOUND SUTURING TIME. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 22048013 ↗MalfunctionMay 12, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22005347 ↗MalfunctionMar 17, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT A PATIENT UNDERWENT A THORACIC SURGERY PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE DEVICE PACKAGE WAS FOUND UNSEALED WHEN UNPACKING AND COUNTING. ONE END OF THE PACKAGE IS NOT SEALED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21625221 ↗MalfunctionJan 7, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2024 AND SUTURE WAS USED. THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THE SURGERY. IT CAUSED MALFUNCTIONS DURING THE ANASTOMOSIS OF THE HEART AND BLOOD VESSELS, DELAYED THE TIMING OF THE SURGERY AND CAUSING WASTE OF CONSUMABLES. STOPPED USING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21096630 ↗MalfunctionJan 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN CARDIOVASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO TIE A KNOT WITH SUTURE. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL BE 25 DEVICES RETURNED FOR ANALYSIS, INCLUDING 2 ACTUAL PRODUCTS AND 23 STERILE PRODUCTS FROM SAME BATCH. THE SURGEON REFUSED TO USE THE REMAINING 23 STERILE PRODUCTS FROM THE SAME LOT AND THEY WILL BE RETURNED FOR ANALYSIS. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21068306 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2457-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2459-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
- Z-2455-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
- Z-2450-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
- Z-2460-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
- Z-2449-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
- Z-2451-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
- Z-2461-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2454-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
…and 6 more.
Other Suture, nonabsorbable, synthetic, polypropylene devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.