Coated VICRYL VCP416H (UDI 10705031078147)Recalls & reported problems
Ethicon Inc. · Antibacterial Undyed Braided (Polyglactin 910), Synthetic Absorbable Suture · Class 2 · Implant
All models of Coated VICRYL (114)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10705031078147), received between Dec 4, 2019 and Mar 24, 2026 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Break” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break9
- Material Separation2
- Device Markings/Labelling Problem1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Insufficient Information2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 24, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE NEEDLE IS BREAKING. THERE WERE THREE OF THESE IN THE SAME PROCEDURE. NO ADVERSE IMPACT PATIENT/USER. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24681381 ↗MalfunctionMar 24, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE NEEDLE IS BREAKING. THERE WERE THREE OF THESE IN THE SAME PROCEDURE. NO ADVERSE IMPACT PATIENT/USER. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24675475 ↗MalfunctionMar 24, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE NEEDLE IS BREAKING. THERE WERE THREE OF THESE IN THE SAME PROCEDURE. NO ADVERSE IMPACT PATIENT/USER. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24675474 ↗MalfunctionMar 17, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A PEDIATRIC PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY THE PEDIATRIC SURGERY AND ENDOSCOPY MANAGER THAT TWO CONSECUTIVE NEEDLE TIPS OF THE SUTURE BROKE OFF DURING A CASE. NURSE STATES THAT NEEDLE TIP BROKE OFF WHEN GRABBED AFTER PULLING THROUGH TISSUE IN AN ORTHO PROCEDURE ON THE SUBCUTANEOUS LAYER. BROKEN TIP HAD TO BE RETRIEVED FROM PATIENT. THERE WAS A 5MIN OF SURGICAL DELAY REPORTED. BOTH DEVICES WERE RETURNING. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 24618004 ↗MalfunctionMar 17, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A PEDIATRIC PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. IT WAS REPORTED BY THE PEDIATRIC SURGERY AND ENDOSCOPY MANAGER THAT TWO CONSECUTIVE NEEDLE TIPS OF THE SUTURE BROKE OFF DURING A CASE. NURSE STATES THAT NEEDLE TIP BROKE OFF WHEN GRABBED AFTER PULLING THROUGH TISSUE IN AN ORTHO PROCEDURE ON THE SUBCUTANEOUS LAYER. BROKEN TIP HAD TO BE RETRIEVED FROM PATIENT. THERE WAS A 5MIN OF SURGICAL DELAY REPORTED. BOTH DEVICES WERE RETURNING. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 24617994 ↗MalfunctionMar 9, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN OPEN ABDOMINAL SURGERY ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, THE TIP OF THE NEEDLE BROKE OFF. AND IT WAS FOUND DURING X-RAY. THE NEEDLE PIECE WAS RETRIEVED DURING THE PROCEDURE AND RADIOLOGY WAS CALLED IN TO PERFORM AN X-RAY AND NO ADDITIONAL NEEDLE PIECES WERE FOUND INSIDE THE BODY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24548418 ↗MalfunctionFeb 24, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. THE DEVICE WAS RECEIVED WITH NO COMPLAINT INFORMATION. ATTEMPTS WERE MADE TO RECONCILE THE DEVICE WITH NO SUCCESS RESULTING IN THE CREATION OF A NEW COMPLAINT. UPON EVALUATION OF THE RETURNED DEVICE, A FRACTURE WAS OBSERVED IN THE BODY OF THE NEEDLE. THE NEEDLE WAS RECEIVED IN TWO PIECES, ONLY ONE OF THE MATING FRACTURE SURFACES WAS EXAMINED FOR THIS EVALUATION. A SCANNING ELECTRON MICROSCOPE (SEM) WAS USED TO EXAMINE THE FRACTURE SURFACES AND SURROUNDING AREA OF THE NEEDLE. THE FRACTURE SURFACES WERE EXAMINED IN MULTIPLE LOCATIONS TO DETERMINE THE FRACTURE MODE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24430071 ↗MalfunctionJan 5, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED BY DISTRIBUTOR PER ACCOUNT THAT DURING CLOSURE, THE PROVIDER NOTED A SUTURE NEEDLE TIP HAD BROKEN. THE PROCEDURE WAS PAUSED, INTRAOPERATIVE X-RAY CONFIRMED THE FRAGMENT LOCATION, AND IT WAS SUCCESSFULLY RETRIEVED. A SECOND X-RAY VERIFIED NO REMAINING FRAGMENTS. THE PROCEDURE WAS COMPLETED WITHOUT FURTHER COMPLICATION. THERE IS NO INJURY OR HARM TO THE PATIENT REPORTED. DEVICE IS NOT AVAILABLE FOR RETURN. NO ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23965607 ↗MalfunctionDec 18, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY HEALTHCARE PROFESSIONAL THAT THEY ORDERED 2 MORE BOXES OF THE VCP416H 3-0 VICRYL SH AND THE CAME IN DYED AGAIN. THIS AN UNDYED SUTURE BUT THE LABEL STATES DYED. PRODUCTS ARE AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23837045 ↗MalfunctionSep 15, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A EXCISION OF LESION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING EXCISION OF LESION PROCEDURE THEY NOTICED THAT THE NEEDLE TIP BROKE OFF ON THE NEEDLE THEY FOUND IT ON THE MAYO STAND NOT ON THE PATIENT. CASE WAS COMPLETED BY OPENING A NEW ONE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23057508 ↗MalfunctionApr 23, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT WHILE SUTURING THE PATIENT'S SKIN, THE SURGEON STATED THAT THE VERY TIP OF THE NEEDLE BROKE OFF THE MAIN NEEDLE. NO INJURY, REACHED THE INDIVIDUAL BUT DID NOT CAUSE HARM. THE SAMPLE WAS DISCARDED IN THE SHARPS CONTAINER AFTER USE. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21894378 ↗MalfunctionDec 4, 2019Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2019 AND THE SUTURE WAS USED. IT WAS ALSO REPORTED THAT THE NEEDLE WAS MISSING FROM SUTURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9411247 ↗MalfunctionDec 4, 2019Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2019 AND THE SUTURE WAS USED. IT WAS ALSO REPORTED THAT THE NEEDLE WAS MISSING FROM SUTURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9411144 ↗
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page