Skip to content

PROLENE EH7240HRecalls & reported problems

Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant

All models of PROLENE (358)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705031095618), received between Apr 16, 2019 and Mar 19, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Separation” (3 reports).

21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation3
  • Break1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 19, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE SUTURE BREAKS WHEN USED. IT IS NOT WHERE THE NEEDLE IS ATTACHED, BUT APPROX. 5-7 CM FROM THE ATTACHMENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21642242 ↗
  • MalfunctionApr 22, 2019Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE OF THE SUTURE CAME OFF. THE PROCEDURE WAS COMPLETED WITH A LIKE DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8535993 ↗
  • MalfunctionApr 22, 2019Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE OF THE SUTURE CAME OFF. THE PROCEDURE WAS COMPLETED WITH A LIKE DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8535991 ↗
  • MalfunctionApr 16, 2019Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE OF THE SUTURE CAME OFF. THE PROCEDURE WAS COMPLETED WITH A LIKE DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8518954 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.

  • Z-1515-2025 · Mar 7, 2025 · Open, Classified
    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
  • Z-1096-2025 · Dec 20, 2024 · Open, Classified
    PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
  • Z-2449-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
  • Z-2457-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
  • Z-2460-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
  • Z-2463-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
  • Z-2458-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
  • Z-2459-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
  • Z-2461-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
  • Z-2466-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE 122CM M1.5, D10189
  • Z-2451-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
  • Z-2455-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
  • Z-2452-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
  • Z-2462-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
  • Z-2448-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031

…and 6 more.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page