PROLENE W2777Recalls & reported problems
Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant
All models of PROLENE (358)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Prolene blue polypropylene surgical suture is a nonabsorbable stitching thread used to close tissue or join blood vessels during surgery. The 31 reports mostly describe the thread or needle breaking mid-procedure, and sometimes separating from the needle, so surgeons replaced it and continued; no deaths and no injuries were reported. Reports dropped from 23 in the prior 12 months to 5 in the most recent. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10705031126039), received between Jun 14, 2024 and Sep 30, 2026 — classified as 36 malfunctions.
The most frequently reported specific device problem is “Break” (32 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break32
- Material Separation4
Patient problems
- No Clinical Signs, Symptoms or Conditions33
- Insufficient Information3
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Break
It was reported that a patient underwent an unknown procedure on (b)(6) 2026 and suture was used. The suture was broken when the surgeon attempted to sew the tissue. A Total of six products had the issue suture breakage occur. Changed to another one to complete the surgery. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945386 ↗MalfunctionSep 30, 2026Break
It was reported that a patient underwent an unknown procedure on (b)(6) 2026 and suture was used. The suture was broken when the surgeon attempted to sew the tissue. A Total of six products had the issue suture breakage occur. Changed to another one to complete the surgery. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945360 ↗MalfunctionSep 30, 2026Break
It was reported that a patient underwent an unknown procedure on (b)(6)2026 and suture was used. The suture was broken when the surgeon attempted to sew the tissue. A Total of six products had the issue suture breakage occur. Changed to another one to complete the surgery No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945331 ↗MalfunctionSep 30, 2026Break
It was reported that a patient underwent an unknown procedure on (b)(6) 2026 and suture was used. The suture was broken when the surgeon attempted to sew the tissue. A Total of six products had the issue suture breakage occur. Changed to another one to complete the surgery. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945330 ↗MalfunctionSep 30, 2026Break
It was reported that a patient underwent an unknown procedure on (b)(6) 2026 and suture was used. The suture was broken when the surgeon attempted to sew the tissue. A Total of six products had the issue suture breakage occur. Changed to another one to complete the surgery No adverse patient consequences were reported. Additional information was requested
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945295 ↗MalfunctionSep 30, 2026Break
It was reported that a patient underwent an unknown procedure on (b)(6) 2026 and suture was used. The suture was broken when the surgeon attempted to sew the tissue. A Total of six products had the issue suture breakage occur. Changed to another one to complete the surgery No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945278 ↗MalfunctionApr 20, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24932287 ↗MalfunctionApr 14, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE WHEN SEWING IN AN UNKNOWN SURGERY. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE ACTUAL SAMPLE CAN'T BE RETURNED, 2 ACTUAL SAMPLES WERE DISCARD, BUT 20 STERILE SAMPLES FROM SAME BATCH WILL BE RETURNED FOR ANALYSIS. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24871977 ↗MalfunctionApr 14, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE WHEN SEWING IN AN UNKNOWN SURGERY. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE ACTUAL SAMPLE CAN'T BE RETURNED, 2 ACTUAL SAMPLES WERE DISCARD, BUT 20 STERILE SAMPLES FROM SAME BATCH WILL BE RETURNED FOR ANALYSIS. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24871976 ↗MalfunctionFeb 13, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE NEEDLE BROKE WHEN SEWING IN SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24337312 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015371 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015361 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015355 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015341 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015323 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015314 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015300 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015268 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015248 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015235 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015220 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015170 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015161 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF (B)(4) PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015082 ↗MalfunctionSep 10, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE DOCTOR USED SUTURE TO ANASTOMOSE BLOOD VESSELS DURING THE SURGERY. A TOTAL OF 29 PACKS OF THIS SUTURE WERE USED THROUGHOUT THE ENTIRE PROCESS, BUT 19 PACKS BROKE DURING THE SUTURING PROCESS. THE BROKEN SUTURES HAVE NOT BEEN USED AGAIN, AND NEW SUTURES HAVE BEEN REPLACED. THERE WERE NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23015077 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2455-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
- Z-2463-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
- Z-2457-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
- Z-2451-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2459-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
- Z-2454-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
- Z-2462-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
- Z-2452-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
…and 6 more.
Other Suture, nonabsorbable, synthetic, polypropylene devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page