PROLENE EPM8755Recalls & reported problems
Ethicon, LLC · PROLENE, POLYPROPYLENE (MONOFILAMENT) STERILE SYNTHETIC NON-ABSORBABLE SURGICAL SUTURE · Class 2 · Implant
Quick summary
This Ethicon Prolene synthetic non-absorbable surgical suture is used to close tissue during surgery. The 36 reports mostly describe the suture breaking while the surgeon was stitching, or the needle pulling away from the thread, sometimes repeatedly during coronary artery bypass and other operations; a few mention blood loss. There were 36 malfunctions, no deaths and no injuries. No recalls name this device. Reports dropped from 22 in the earlier 12 months to 2 in the most recent.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10705031203709), received between Jul 10, 2020 and Feb 11, 2026 — classified as 36 malfunctions.
The most frequently reported specific device problem is “Break” (26 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break26
- Material Separation13
Patient problems
- No Clinical Signs, Symptoms or Conditions28
- Hemorrhage/Blood Loss/Bleeding7
- Insufficient Information3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 11, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. THE DOCTOR DISCARDED THE SUTURE AND REPLACED IT WITH A NEW ONE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24315777 ↗MalfunctionDec 8, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHEN SUTURING BLOOD VESSELS . BESIDES, THE SUTURE WAS BROKEN . CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23751062 ↗MalfunctionOct 2, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHEN SUTURING BLOOD VESSELS. TOTAL 3 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL RETURN 1 ACTUAL PRODUCTS AND 4 REPRESENTATIVE SAMPLES FOR ANALYSIS. ENCLOSED PLEASE FIND DEFECT PHOTO. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23199793 ↗MalfunctionOct 2, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CORONARY ARTERY BYPASS PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHEN SUTURING BLOOD VESSELS . TOTAL 3 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL RETURN 1 ACTUAL PRODUCTS AND 4 REPRESENTATIVE SAMPLES FOR ANALYSIS. ENCLOSED PLEASE FIND DEFECT PHOTO. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23199792 ↗MalfunctionOct 2, 2025Material Separation; Material Separation
PULL OFF SUTURE NEEDLE. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHEN SUTURING BLOOD VESSELS . TOTAL 3 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL RETURN 1 ACTUAL PRODUCTS AND 4 REPRESENTATIVE SAMPLES FOR ANALYSIS. ENCLOSED PLEASE FIND DEFECT PHOTO. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23199790 ↗MalfunctionJun 20, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC UNILATERAL SALPINGO OOPHORECTOMY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW WOUND. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22299551 ↗MalfunctionApr 29, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21927415 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER SIX TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21297461 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2025 AND A SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER SIX TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21297403 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER SIX TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21297364 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER SIX TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21297339 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER SIX TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21297322 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2025 AND A SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER SIX TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21297272 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER SIX TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21297256 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. DURING SURGERY, NEEDLE BREAKAGE (INVOLVED QTY:1) AND SUTURE BREAKAGE (INVOLVED QTY:2) OCCURRED DURING SEWING. THEN CT WAS USED TO LOOK FOR AND REMOVE THE BROKEN NEEDLE FROM PATIENT. THE SURGERY WAS DELAYED FOR MORE THAN 2 HOURS. THE PATIENT IS CURRENTLY STABLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21293907 ↗MalfunctionFeb 3, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. DURING SURGERY, NEEDLE BREAKAGE (INVOLVED QTY:1) AND SUTURE BREAKAGE (INVOLVED QTY:2) OCCURRED DURING SEWING. THEN CT WAS USED TO LOOK FOR AND REMOVE THE BROKEN NEEDLE FROM PATIENT. THE SURGERY WAS DELAYED FOR MORE THAN 2 HOURS. THE PATIENT IS CURRENTLY STABLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21293576 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. DURING SURGERY, NEEDLE BREAKAGE (INVOLVED QTY: (B)(4)) AND SUTURE BREAKAGE (INVOLVED QTY: (B)(4)) OCCURRED DURING SEWING. THEN CT WAS USED TO LOOK FOR AND REMOVE THE BROKEN NEEDLE FROM PATIENT. THE SURGERY WAS DELAYED FOR MORE THAN 2 HOURS. THE PATIENT IS CURRENTLY STABLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21293422 ↗MalfunctionJan 9, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE BROKE DURING THE SURGICAL SUTURING PROCESS WHICH PROLONGED THE SURGERY TIME AND INCREASED THE AMOUNT OF BLEEDING. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21118818 ↗MalfunctionJan 9, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE BROKE DURING THE SURGICAL SUTURING PROCESS WHICH PROLONGED THE SURGERY TIME AND INCREASED THE AMOUNT OF BLEEDING. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21118817 ↗MalfunctionJan 9, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE BROKE DURING THE SURGICAL SUTURING PROCESS WHICH PROLONGED THE SURGERY TIME AND INCREASED THE AMOUNT OF BLEEDING. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21118815 ↗MalfunctionJan 9, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6)2024 AND SUTURE WAS USED. THE SUTURE BROKE DURING THE SURGICAL SUTURING PROCESS WHICH PROLONGED THE SURGERY TIME AND INCREASED THE AMOUNT OF BLEEDING. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21118814 ↗MalfunctionJan 9, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE BROKE DURING THE SURGICAL SUTURING PROCESS WHICH PROLONGED THE SURGERY TIME AND INCREASED THE AMOUNT OF BLEEDING. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21118813 ↗MalfunctionJan 9, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE BROKE DURING THE SURGICAL SUTURING PROCESS WHICH PROLONGED THE SURGERY TIME AND INCREASED THE AMOUNT OF BLEEDING. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21118811 ↗MalfunctionDec 11, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED WHEN THE DOCTOR USED SUTURE TO SUTURE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THIS INCREASED THE PATIENT'S BLEEDING AND PROLONGED THE OPERATION TIME. THE DOCTOR HAS REPEATEDLY REPORTED THAT THE NEEDLE AND SUTURE WILL AUTOMATICALLY SEPARATE DURING MICRO STITCHING. THERE WERE NO POST-OPERATIVE ADVERSE PATIENT CONSEQUENCES REPORTED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20906820 ↗MalfunctionSep 26, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED ON THE BLOOD VESSELS. DURING THE PROCEDURE, THE SUTURE BROKE WITH THE SUTURING PROCESS. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20314517 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2462-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
- Z-2455-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
- Z-2459-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
- Z-2452-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
- Z-2460-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
- Z-2451-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
- Z-2463-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
- Z-2450-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
- Z-2448-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2461-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.