ENDOLOOP EJ10GRecalls & reported problems
Ethicon Inc. · VICRYL VLT 0 45CM ENDOLOOP · Class 2 · Implant
All models of ENDOLOOP (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705031206465), received between Mar 20, 2018 and Oct 27, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Component Misassembled” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Component Misassembled1
- Delivered as Unsterile Product1
- Hole In Material1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 27, 2023Component Misassembled; Component Misassembled
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2023 AND SUTURE WAS USED. BEFORE USE ON THE PATIENT, IT WAS REPORTED BY THE DISTRIBUTOR THAT THE BOX OF COATED ENDOLOOP CONTAINS A PROLENE SUTURE INSTEAD OF LABELED VICRYL SUTURE. STERILE SAMPLES RETURNING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18024753 ↗InjuryJul 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE DEVICE WAS LEFT INSIDE THE PATIENTS LEG. THE DEVICE WAS REMOVED AT A DIFFERENT ACCOUNT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 17408482 ↗MalfunctionMar 20, 2018Hole In Material; Delivered as Unsterile Product
SURGEON REQUESTED THE ENDOLOOP DURING A LAP CHOLE. THE PACKAGING IS THE SAME FOIL MATERIAL AS ANOTHER ETHICON PRODUCT STAFFS HAVE FACED ISSUES WITH. THE RN OPENED PACKAGE TO THE TECH, BUT INSTRUCTED THE TECH NOT TO PLACE ITEM ONTO THE SURGICAL FIELD AS IT MAY HAVE PIN-PRICK TYPE HOLES IN THE FOIL. AS EXPECTED, 8+ SMALL HOLES WERE FOUND IN THE PRODUCT, INDICATING A BREACH IN STERILITY THE TECH REMOVED THE ITEM AND HIS GLOVE TO RN TO PLACE INTO GARBAGE. THE TECH DONNED A NEW PAIR OF STERILE GLOVES ASEPTICALLY AND REPEATED PROCESS WITH SECOND ENDOLOOP PACKAGING INTACT. THE PACKAGING WAS SAVED FOR CLINICAL ENGINEERING, AND IMAGES OF THE PIN PRICK HOLES WERE ACQUIRED USING A MICROSCOPE LIGHT TO ILLUMINATE WHERE THE HOLES WERE LOCATED IN THE FOIL. PER SITE REPORTER: A COMPLAINT FILE HAS BEEN OPENED AND THE PACKAGE WILL BE RETURNED FOR FAILURE ANALYSIS.
Patient problems: No Patient Involvement
MAUDE report 7352168 ↗
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page