Skip to content

PROLENE EP8709HRecalls & reported problems

Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant

Quick summary

This Ethicon Prolene EP8709H is a blue polypropylene surgical suture sold in a pack with two needles. Most of the 34 reports describe the thread snapping or pulling away from the needle mid-procedure, or the material separating or fraying, and in one case a needle sealed inside the foil packaging. There are no deaths and one injury, and no recalls name this device. Reports dropped sharply, from 20 in the earlier 12-month period to 2 in the most recent.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 10705031208308), received between Apr 10, 2019 and Jun 4, 2026 — classified as 1 injury, 33 malfunctions.

The most frequently reported specific device problem is “Break” (25 reports).

21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 14 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break25
  • Material Separation7
  • Defective Component2
  • Delivered as Unsterile Product1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • No Consequences Or Impact To Patient6
  • Insufficient Information3
  • Hematoma1
  • Speech Disorder1

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, FEW OF THAT SUTURES HAD ISSUE WITH NEEDLE GETTING DISCONNECTED OF THE THREAD WITH MINIMAL STRENGTH USED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25393322 ↗
  • MalfunctionJun 4, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, FEW OF THAT SUTURES HAD ISSUE WITH NEEDLE GETTING DISCONNECTED OF THE THREAD WITH MINIMAL STRENGTH USED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25393319 ↗
  • MalfunctionSep 17, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT IS CLEARLY A MANUFACTURING DEFECT. UPON OPENING THE FOIL, IT BECOMES EVIDENT THAT ONE OF THE TWO NEEDLES HAS BEEN INADVERTENTLY SEALED INSIDE THE PACKAGING. DID THE EVENT OCCUR DURING STERILE PROCESSING? YES, DID THE EVENT OCCUR DURING FIELD INSPECTION? YES, DID THE EVENT OCCUR DURING INTERNAL SERVICE ACTIVITIES SUCH AS CALIBRATION? NO, PATIENT STATUS/ OUTCOME / CONSEQUENCES: NO PATIENT INVOLVEMENT, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY: UNKNOWN, (B)(4). DEVICE PROPERTY OF: NONE, DEVICE IN POSSESSION OF: NONE. BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23076032 ↗
  • MalfunctionApr 25, 2025Break; Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21911474 ↗
  • MalfunctionMar 19, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE SUTURE BREAKS WHEN USED. IT IS NOT WHERE THE NEEDLE IS ATTACHED, BUT APPROX. 5-7 CM FROM THE ATTACHMENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21641246 ↗
  • MalfunctionMar 19, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE SUTURE BREAKS WHEN USED. IT IS NOT WHERE THE NEEDLE IS ATTACHED, BUT APPROX. 5-7 CM FROM THE ATTACHMENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21641240 ↗
  • MalfunctionFeb 20, 2025Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. VASCULAR SURGEONS ARE SAYING IS SNAPPING AND FAULTY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21424236 ↗
  • MalfunctionFeb 20, 2025Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. VASCULAR SURGEONS ARE SAYING IS SNAPPING AND FAULTY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21424000 ↗
  • MalfunctionFeb 5, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21316041 ↗
  • MalfunctionFeb 5, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21316040 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21300117 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21300110 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21300051 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21300046 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21300023 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21300015 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE IN THE MIDDLE WHILE TYING. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21299981 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE IN THE MIDDLE WHILE TYING. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21299979 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21299977 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON(B)(6)2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21299975 ↗
  • MalfunctionNov 20, 2024Material Separation; Defective Component; Material Separation; Defective Component

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES ARE DRY, AND THE NEEDLES ARE POPPING OFF. THEY SPENT 15 MINUTES SEARCHING THE FLOOR IN THE ROOM FOR ONE OF THE SUTURES THAT POPPED OFF TODAY. DR ASKED WHAT IS GOING TO BE DONE ABOUT THESE SUTURES, IT'S HAPPENING DAILY WITH MANY DIFFERENT LOT NUMBERS AND NOW EVEN 2 DIFFERENT NEEDLES. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20738007 ↗
  • MalfunctionNov 20, 2024Material Separation; Defective Component; Material Separation; Defective Component

    IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES ARE DRY, AND THE NEEDLES ARE POPPING OFF. THEY SPENT 15 MINUTES SEARCHING THE FLOOR IN THE ROOM FOR ONE OF THE SUTURES THAT POPPED OFF TODAY. DR ASKED WHAT IS GOING TO BE DONE ABOUT THESE SUTURES, IT'S HAPPENING DAILY WITH MANY DIFFERENT LOT NUMBERS AND NOW EVEN 2 DIFFERENT NEEDLES. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20736244 ↗
  • MalfunctionJul 17, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A KIDNEY SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE OFF THE POLYMER. ALMOST AS IF THIS WAS A CONTROLLED RELEASE SUTURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES FOR THE PATIENT. NO SURGICAL DELAY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19761063 ↗
  • MalfunctionJul 17, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A KIDNEY SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE OFF THE POLYMER. ALMOST AS IF THIS WAS A CONTROLLED RELEASE SUTURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES FOR THE PATIENT. NO SURGICAL DELAY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19761061 ↗
  • InjuryApr 30, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE SUTURE BROKE. THIS RESULTED IN A REOPERATION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hematoma; Insufficient Information; Speech Disorder

    MAUDE report 19215625 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.

  • Z-1515-2025 · Mar 7, 2025 · Open, Classified
    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
  • Z-1096-2025 · Dec 20, 2024 · Open, Classified
    PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
  • Z-2458-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
  • Z-2464-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
  • Z-2462-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
  • Z-2456-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
  • Z-2452-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
  • Z-2448-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
  • Z-2453-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
  • Z-2454-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
  • Z-2460-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
  • Z-2465-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
  • Z-2455-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
  • Z-2466-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE 122CM M1.5, D10189
  • Z-2461-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746

…and 6 more.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.