This Ethicon Prolene EP8709H is a blue polypropylene surgical suture sold in a pack with two needles. Most of the 34 reports describe the thread snapping or pulling away from the needle mid-procedure, or the material separating or fraying, and in one case a needle sealed inside the foil packaging. There are no deaths and one injury, and no recalls name this device. Reports dropped sharply, from 20 in the earlier 12-month period to 2 in the most recent.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 10705031208308), received between Apr 10, 2019 and Jun 4, 2026 — classified as 1 injury, 33 malfunctions.
The most frequently reported specific device problem is “Break” (25 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break25
- Material Separation7
- Defective Component2
- Delivered as Unsterile Product1
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions27
- No Consequences Or Impact To Patient6
- Insufficient Information3
- Hematoma1
- Speech Disorder1
Recent adverse event reports
The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 4, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, FEW OF THAT SUTURES HAD ISSUE WITH NEEDLE GETTING DISCONNECTED OF THE THREAD WITH MINIMAL STRENGTH USED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25393322 ↗MalfunctionJun 4, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, FEW OF THAT SUTURES HAD ISSUE WITH NEEDLE GETTING DISCONNECTED OF THE THREAD WITH MINIMAL STRENGTH USED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25393319 ↗MalfunctionSep 17, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT IS CLEARLY A MANUFACTURING DEFECT. UPON OPENING THE FOIL, IT BECOMES EVIDENT THAT ONE OF THE TWO NEEDLES HAS BEEN INADVERTENTLY SEALED INSIDE THE PACKAGING. DID THE EVENT OCCUR DURING STERILE PROCESSING? YES, DID THE EVENT OCCUR DURING FIELD INSPECTION? YES, DID THE EVENT OCCUR DURING INTERNAL SERVICE ACTIVITIES SUCH AS CALIBRATION? NO, PATIENT STATUS/ OUTCOME / CONSEQUENCES: NO PATIENT INVOLVEMENT, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY: UNKNOWN, (B)(4). DEVICE PROPERTY OF: NONE, DEVICE IN POSSESSION OF: NONE. BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23076032 ↗MalfunctionApr 25, 2025Break; Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21911474 ↗MalfunctionMar 19, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE SUTURE BREAKS WHEN USED. IT IS NOT WHERE THE NEEDLE IS ATTACHED, BUT APPROX. 5-7 CM FROM THE ATTACHMENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21641246 ↗MalfunctionMar 19, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE SUTURE BREAKS WHEN USED. IT IS NOT WHERE THE NEEDLE IS ATTACHED, BUT APPROX. 5-7 CM FROM THE ATTACHMENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21641240 ↗MalfunctionFeb 20, 2025Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. VASCULAR SURGEONS ARE SAYING IS SNAPPING AND FAULTY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21424236 ↗MalfunctionFeb 20, 2025Break; Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. VASCULAR SURGEONS ARE SAYING IS SNAPPING AND FAULTY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21424000 ↗MalfunctionFeb 5, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21316041 ↗MalfunctionFeb 5, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21316040 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21300117 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21300110 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21300051 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21300046 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21300023 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21300015 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE IN THE MIDDLE WHILE TYING. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21299981 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE IN THE MIDDLE WHILE TYING. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21299979 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21299977 ↗MalfunctionFeb 3, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A CAROTID ARTERY PROCEDURE ON(B)(6)2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE RIGHT AT THE SWAGE. ANOTHER LIKE DEVICE FROM THE SAME LOT WAS USED TO CONTINUE WITH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21299975 ↗MalfunctionNov 20, 2024Material Separation; Defective Component; Material Separation; Defective Component
IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES ARE DRY, AND THE NEEDLES ARE POPPING OFF. THEY SPENT 15 MINUTES SEARCHING THE FLOOR IN THE ROOM FOR ONE OF THE SUTURES THAT POPPED OFF TODAY. DR ASKED WHAT IS GOING TO BE DONE ABOUT THESE SUTURES, IT'S HAPPENING DAILY WITH MANY DIFFERENT LOT NUMBERS AND NOW EVEN 2 DIFFERENT NEEDLES. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20738007 ↗MalfunctionNov 20, 2024Material Separation; Defective Component; Material Separation; Defective Component
IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURES ARE DRY, AND THE NEEDLES ARE POPPING OFF. THEY SPENT 15 MINUTES SEARCHING THE FLOOR IN THE ROOM FOR ONE OF THE SUTURES THAT POPPED OFF TODAY. DR ASKED WHAT IS GOING TO BE DONE ABOUT THESE SUTURES, IT'S HAPPENING DAILY WITH MANY DIFFERENT LOT NUMBERS AND NOW EVEN 2 DIFFERENT NEEDLES. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20736244 ↗MalfunctionJul 17, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A KIDNEY SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE OFF THE POLYMER. ALMOST AS IF THIS WAS A CONTROLLED RELEASE SUTURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES FOR THE PATIENT. NO SURGICAL DELAY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19761063 ↗MalfunctionJul 17, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A KIDNEY SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE OFF THE POLYMER. ALMOST AS IF THIS WAS A CONTROLLED RELEASE SUTURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES FOR THE PATIENT. NO SURGICAL DELAY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19761061 ↗InjuryApr 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, THE SUTURE BROKE. THIS RESULTED IN A REOPERATION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Hematoma; Insufficient Information; Speech Disorder
MAUDE report 19215625 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
- Z-2462-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
- Z-2456-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
- Z-2452-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
- Z-2448-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE - BV1755 Taper Point 3/8C, D10031
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2454-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
- Z-2460-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2455-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
- Z-2461-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.