PROLENE EP8730HRecalls & reported problems

Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant

Quick summary

This Ethicon Prolene blue polypropylene surgical suture is a nonabsorbable stitch used to close tissue. The 36 reports mostly describe the suture material separating or breaking during surgery, with the needle coming detached from the thread mid-procedure; no deaths and no injuries were reported. There are no recalls naming this device. Reports rose from none in the prior 12 months to 7 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
36
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10705031215931), received between Jun 13, 2019 and Aug 5, 2026 — classified as 36 malfunctions.

The most frequently reported specific device problem is “Material Separation” (29 reports).

21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 5 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation29
  • Break8

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • No Consequences Or Impact To Patient10
  • Insufficient Information2

Recent adverse event reports

The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50143544 ↗
  • MalfunctionAug 5, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50143538 ↗
  • MalfunctionDec 24, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING SURGERY, THE SUTURE WAS EASY TO DETACH, AND THE ISSUE OCCURRED 5 NEEDLE-SUTURES IN A ROW. THEREFORE, ANOTHER PRODUCT WITH THE DIFFERENT LOT NUMBER WAS USED ALTERNATIVELY. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23888446 ↗
  • MalfunctionDec 24, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING SURGERY, THE SUTURE WAS EASY TO DETACH, AND THE ISSUE OCCURRED 5 NEEDLE-SUTURES IN A ROW. THEREFORE, ANOTHER PRODUCT WITH THE DIFFERENT LOT NUMBER WAS USED ALTERNATIVELY. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23888422 ↗
  • MalfunctionDec 24, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING SURGERY, THE SUTURE WAS EASY TO DETACH, AND THE ISSUE OCCURRED 5 NEEDLE-SUTURES IN A ROW. THEREFORE, ANOTHER PRODUCT WITH THE DIFFERENT LOT NUMBER WAS USED ALTERNATIVELY. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23888394 ↗
  • MalfunctionDec 24, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING SURGERY, THE SUTURE WAS EASY TO DETACH, AND THE ISSUE OCCURRED 5 NEEDLE-SUTURES IN A ROW. THEREFORE, ANOTHER PRODUCT WITH THE DIFFERENT LOT NUMBER WAS USED ALTERNATIVELY. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23888352 ↗
  • MalfunctionDec 24, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING SURGERY, THE SUTURE WAS EASY TO DETACH, AND THE ISSUE OCCURRED 5 NEEDLE-SUTURES IN A ROW. THEREFORE, ANOTHER PRODUCT WITH THE DIFFERENT LOT NUMBER WAS USED ALTERNATIVELY. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23888346 ↗
  • MalfunctionApr 20, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THERE WAS AN ISSUE WHERE THE NEEDLE CAME AWAY FROM THE SUTURE, MIDWAY THROUGH SUTURING. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16778018 ↗
  • MalfunctionApr 6, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON 03/13/2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE 8-0 PROLENE IS COMING AWAY FROM THE SUTURE, MIDWAY THROUGH SUTURING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16693811 ↗
  • MalfunctionApr 6, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE 8-0 PROLENE IS COMING AWAY FROM THE SUTURE, MIDWAY THROUGH SUTURING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16693801 ↗
  • MalfunctionApr 6, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON 03/13/2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE 8-0 PROLENE IS COMING AWAY FROM THE SUTURE, MIDWAY THROUGH SUTURING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16693796 ↗
  • MalfunctionApr 6, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE 8-0 PROLENE IS COMING AWAY FROM THE SUTURE, MIDWAY THROUGH SUTURING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16693681 ↗
  • MalfunctionAug 30, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON 08/10/2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLES HAVE BROKEN AWAY FROM THE SUTURE MATERIAL. IT WAS A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15320348 ↗
  • MalfunctionAug 30, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLES HAVE BROKEN AWAY FROM THE SUTURE MATERIAL. IT WAS A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15320327 ↗
  • MalfunctionAug 30, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLES HAVE BROKEN AWAY FROM THE SUTURE MATERIAL. IT WAS A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15320322 ↗
  • MalfunctionJul 29, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG ON (B)(6) 2022 AND SUTURE WAS USED. THE NEEDLE CAME LOOSE FROM THE SUTURE INTRA-OPERATIVELY. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THE PROCEDURE WAS PROLONGED FOR 2 MINUTES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15136460 ↗
  • MalfunctionJul 29, 2022Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG ON (B)(6) 2022 AND SUTURE WAS USED. THE NEEDLE CAME LOOSE FROM THE SUTURE INTRA-OPERATIVELY. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THE PROCEDURE WAS PROLONGED FOR 2 MINUTES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15136388 ↗
  • MalfunctionMar 24, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN USING THE SUTURE, IT IS CUT JUST AT THE JOINT BETWEEN SUTURE AND NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 13872764 ↗
  • MalfunctionMar 24, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CABG PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN USING THE SUTURE, IT IS CUT JUST AT THE JOINT BETWEEN SUTURE AND NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 13872534 ↗
  • MalfunctionMay 27, 2021Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE DETACHED FROM THE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11896036 ↗
  • MalfunctionApr 28, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A NECK SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE WHEN TYING. MULTIPLE STRANDS TRIED (AT LEAST 5). THIS CAUSED ADDITIONAL OR TIME. AMOUNT OF TIME NOT SPECIFIED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11737549 ↗
  • MalfunctionApr 28, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A NECK SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE WHEN TYING. MULTIPLE STRANDS TRIED (AT LEAST 5). THIS CAUSED ADDITIONAL OR TIME. AMOUNT OF TIME NOT SPECIFIED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11737547 ↗
  • MalfunctionApr 28, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A NECK SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE WHEN TYING. MULTIPLE STRANDS TRIED (AT LEAST 5). THIS CAUSED ADDITIONAL OR TIME. AMOUNT OF TIME NOT SPECIFIED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11737545 ↗
  • MalfunctionApr 28, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A NECK SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE WHEN TYING. MULTIPLE STRANDS TRIED (AT LEAST 5). THIS CAUSED ADDITIONAL OR TIME. AMOUNT OF TIME NOT SPECIFIED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11737543 ↗
  • MalfunctionApr 28, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A NECK SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE BROKE WHEN TYING. MULTIPLE STRANDS TRIED (AT LEAST 5). THIS CAUSED ADDITIONAL OR TIME. AMOUNT OF TIME NOT SPECIFIED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11737541 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.

  • Z-1515-2025 · Mar 7, 2025 · Open, Classified
    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
  • Z-1096-2025 · Dec 20, 2024 · Open, Classified
    PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
  • Z-2450-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
  • Z-2455-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
  • Z-2458-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
  • Z-2466-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE 122CM M1.5, D10189
  • Z-2464-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
  • Z-2452-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
  • Z-2449-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
  • Z-2465-2018 · May 7, 2018 · Terminated
    PRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
  • Z-2456-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
  • Z-2463-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
  • Z-2460-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 4"(10CM) 8-0 BLUE, D9662
  • Z-2461-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
  • Z-2459-2018 · May 7, 2018 · Terminated
    PROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067

…and 6 more.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.