Skip to content

STRATAFIX SXPP1A301Recalls & reported problems

Ethicon, LLC · Violet Antibacterial Polydioxanone, Knotless Tissue Control Device · Class 2 · Implant

All models of STRATAFIX (180)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Stratafix is a knotless absorbable suture used to close tissue during surgery. Most of its 448 reports describe the suture breaking or its parts separating during use, and some describe wounds reopening after closure, including wound dehiscence. The reports list 78 injuries, no deaths and 370 malfunctions, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
442
Deaths reported
0
Injuries reported
74
Recalls naming it
0

Summary

The FDA's MAUDE database contains 442 adverse event reports that list this exact device (DI 10705031227712), received between Nov 22, 2017 and Aug 31, 2026 — classified as 74 injuries, 368 malfunctions.

The most frequently reported specific device problem is “Break” (321 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 3 reportsJan 2024: 2 reportsFeb 2024: 10 reportsMar 2024: 3 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 4 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 1 reportsDec 2025: 4 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 4 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break321
  • Material Separation41
  • Adverse Event Without Identified Device or Use Problem40
  • Material Integrity Problem40
  • Positioning Failure8
  • Positioning Problem4
  • Defective Component3
  • Failure To Adhere Or Bond3

Patient problems

  • No Clinical Signs, Symptoms or Conditions256
  • No Consequences Or Impact To Patient114
  • Wound Dehiscence44
  • Not Applicable17
  • Hernia13
  • Unspecified Infection9
  • Abscess4
  • Fluid Discharge4

Recent adverse event reports

The 25 most recent of 442 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 31, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that a patient underwent a pancreaticoduodenectomy on (b)(6), 2026 and barbed suture was used for fascial closure. The wound length was about 25 cm. After that the site was determined to have dehisced on July 11 and reoperation was performed using different suture. It was unlikely to be attributable to the suture, though a connection could not be denied. The doctor said that the dehiscence was caused by the fragility of the patient's tissue, rather than an issue with the suture. The patient is stable now. Additional information was requested.

    Patient problems: Hernia; Vomiting; Wound Dehiscence

    MAUDE report 50115276 ↗
  • InjuryAug 3, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that a patient underwent a pancreaticoduodenectomy on (b)(6) 2026 and barbed suture was used for fascial closure. The wound length was about 25 cm. After that the site was determined to have dehisced on (b)(6) and reoperation was performed using different suture. It was unlikely to be attributable to the suture, though a connection could not be denied. The doctor said that the dehiscence was caused by the fragility of the patient's tissue, rather than an issue with the suture. The patient is stable now. Additional information was requested.

    Patient problems: Hernia; Vomiting; Wound Dehiscence

    MAUDE report 50115259 ↗
  • MalfunctionJun 23, 2026Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN ORTHOPEDIC SURGERY ON (B)(6) 2026 AND BARBED SUTURE WAS USED. THE NEEDLE WAS BROKEN AT THE 1ST STITCH WHEN THE PACKAGE WAS OPENED DURING SURGERY. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25592905 ↗
  • InjuryApr 23, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT STOMA CLOSURE PROCEDURE ON (B)(6) 2026 AND A BARBED SUTURE WAS USED ON THE FASCIA. POST-OP, LEAKAGE OCCURRED AFTER STOMA CLOSURE AND REOPERATION WAS PERFORMED. THERE WAS ADHESION BETWEEN THE WOUND (PERITONEUM) AND THE INTESTINES AND UPON DISSECTION, THE FASCIAL SYMMETRY WAS ALSO FOUND TO BE ATTACHED TO THE INTESTINES. THE PATIENT'S OVERALL CONDITION WAS NOT GOOD, AND THE INTESTINES WERE WEAK. THE ADHESIONS BETWEEN THE INTESTINE AND SYMMETRY WERE NOT ALONG THE STAPLE LINES OF THE INTESTINE, BUT RATHER ON THE UNDAMAGED WALL OF THE INTESTINE. ADHESION OCCURRED FROM THE WOUND (PERITONEUM). NO SAMPLE WILL BE RETURNED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Adhesion(s); Fluid Discharge; Insufficient Information; Unspecified Tissue Injury

    MAUDE report 24966956 ↗
  • MalfunctionMar 31, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING AN UNKNOWN ORTHOPEDIC SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE EASILY WHEN THE PRODUCT WAS USED. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742658 ↗
  • MalfunctionMar 31, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN ORTHOPEDIC SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE EASILY AFTER THE PRODUCT WAS USED. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742176 ↗
  • MalfunctionMar 20, 2026Positioning Failure; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A FASCIAL SUTURING IN ROBOT-ASSISTED LAPAROSCOPIC PROSTATECTOMY PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING SUTURING CHECK AFTER FASCIAL SUTURING IN ROBOT-ASSISTED LAPAROSCOPIC PROSTATECTOMY, THE SUTURE DETACHED. THEREFORE, STRATAFIX WAS REMOVED AND THE FASCIA WAS SUTURED WITH PDP. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Wound Dehiscence

    MAUDE report 24651294 ↗
  • MalfunctionMar 12, 2026Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL HIP ARTHROPLASTY PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WERE USED. DURING SURGERY, THE SUTURE WAS BROKEN. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24582395 ↗
  • InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT EMERGENCY LIFE-SAVING SURGERY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED ON THE FASCIA. THE PATIENT EXPERIENCED AN INFECTION. AS TREATMENT AFTER THE INFECTION, REOPERATION WAS PERFORMED. THERE ARE CURRENTLY NO PROBLEMS WITH THE PATIENT¿S CONDITION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Unspecified Infection

    MAUDE report 24379368 ↗
  • MalfunctionDec 30, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, THE SUTURE WAS BROKEN DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23927416 ↗
  • MalfunctionDec 19, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN OPEN CHOLECYSTECTOMY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE SUTURE AND THE NEEDLE WERE DETACHED. THE PRODUCT WAS USED ON THE FASCIA. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23852692 ↗
  • InjuryDec 4, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY (TKA) ON AN (B)(6) 2025 AND A BARBED SUTURE WAS USED ON THE FASCIA. FIVE DAYS AFTER SURGERY, WHEN THE PATIENT WAS CHANGED POSITION IN THE HOSPITAL ROOM, THE KNEE FELT UNCOMFORTABLE AND HAD IT EXAMINED, IT WAS FOUND THAT THE SUTURE WAS BROKEN AT THE WARD, SO RE-OPERATION WAS PERFORMED ON (B)(6) 2025. DURING THE REOPERATION, THE WOUND WAS INCISED AND THE FASCIA AND DERMIS WERE SUTURED AGAIN TO COMPLETE THE PROCEDURE. THE PATIENT'S CONDITION IS CURRENTLY STABLE AND PROGRESSING WITHOUT ANY PARTICULAR PROBLEMS. THE DOCTOR OPINES THAT THE CAUSE WAS NOT THE SUTURES, BUT THAT THE FASCIA WAS SUTURED IN EXTENSION RATHER THAN FLEXED. NO SAMPLE WILL BE RETURNED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Wound Dehiscence

    MAUDE report 23717645 ↗
  • InjuryDec 4, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY (TKA) ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED ON THE FASCIA. AFTER THE SURGERY, THE SUTURE WAS BROKEN AT THE WARD. THE SUTURE BREAKAGE OCCURRED 2 TO 3 DAYS AFTER SURGERY. IT WAS REPORTED THAT REOPERATION WAS REQUIRED, CAUSED BY AN INFECTION. UPON RE-OPERATION, WHEN MAKING AN INCISION IN THE WOUND, THE FASCIA WAS ALSO FOUND TO HAVE SEPARATED. DURING THE REOPERATION, THE WOUND WAS INCISED AND THE FASCIA AND DERMIS WERE SUTURED AGAIN TO COMPLETE THE PROCEDURE. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT. THE DOCTOR OPINES THAT THE CAUSE IS NOT THE SUTURE ITSELF, BUT RATHER THE PROCEDURE ITSELF (BECAUSE THE KNEE WAS SUTURED IN AN EXTENDED POSITION, NOT IN A FLEXED POSITION). NO SAMPLE WILL BE RETURNED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Unspecified Infection; Wound Dehiscence

    MAUDE report 23717642 ↗
  • MalfunctionNov 20, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, ONLY THE NEEDLE TIP WAS BROKEN DURING FASCIORRHAPHY. IT WAS NOT A CONTROL RELEASE NEEDLE. THE NEEDLE TIP FELL TO THE SURGICAL FIELD. THE FALLEN NEEDLE WAS SEARCHED USING X-RAYS, AND IT COULD BE REMOVED FROM INSIDE THE BODY. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23605745 ↗
  • MalfunctionOct 30, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN OB-GYN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23425599 ↗
  • MalfunctionOct 27, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, WHEN USING STRATAFIX SYMMETRIC FOR WOUND CLOSURE, THE NEEDLE-SUTURE WAS DETACHED FROM THE JOINT PART AFTER THE 1ST TO 2ND STITCH. IT WAS HANDLED WITH OPENING AN ADDITIONAL PACKAGE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23395478 ↗
  • MalfunctionOct 15, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL HIP ARTHROPLASTY PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING A TOTAL HIP ARTHROPLASTY: THA, THE PRODUCT INVOLVED WAS USED AS USUAL ON THE FASCIA, BUT THE NEEDLE DETACHED EVEN THOUGH THERE WAS NO TENSION APPLIED. THE PRODUCT WAS BEING USED NORMALLY, BUT THE NEEDLE CAME OFF EVEN THOUGH THERE WAS NO TENSION APPLIED. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23280195 ↗
  • InjuryAug 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CHOLECYSTECTOMY ON (B)(6) 2025 AND BARBED SUTURE WAS USED. THE PATIENT WAS IN A VERY BAD CONDITION WHEN HE WAS TRANSPORTED, BUT SURGERY WAS PERFORMED IMMEDIATELY AT THE REQUEST OF HIS FAMILY. THE PRODUCT WAS USED ON THE FASCIA. THE PATIENT'S VITAL SIGN WAS BAD DURING THE SURGERY, AND THE OPERATION HAD TO BE COMPLETED QUICKLY. TWO BARBED SUTURES WERE USED TO CLOSE THE WOUND. THE MIDLINE INCISION ABOUT 15CM WAS STITCHED FROM THE UPPER AND LOWER ABDOMEN, THE MIDDLE PART WAS OVERLAPPED AND BACKSTITCH WAS PERFORMED. THE EPIDERMIS WAS CLOSED WITH A STAPLER. THE WOUND WAS DEHISCED AND THE ABDOMINAL CAVITY BECAME VISIBLE WHEN A STAPLER WAS REMOVED ON (B)(6). REOPERATION WAS PERFORMED AND THE FASCIA WAS SUTURED WITH NYLON SUTURE. THE PATIENT'S CURRENT CONDITION IS STABLE. THE WOUND WAS DEHISCED FROM THE UPPER TO LOWER ABDOMEN, AND THE SUTURE WAS STILL IN THE TISSUE. THE SUTURE WAS NOT BROKEN, BUT THAT THE TISSUES COULDN'T WITHSTAND. THE DOCTOR COMMENTED THAT THE CAUSE WAS NOT DUE TO THE SUTURE, BUT RATHER TO THE PATIENT'S CONDITION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Wound Dehiscence

    MAUDE report 22862326 ↗
  • InjuryAug 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CHOLECYSTECTOMY ON (B)(6)2025 AND BARBED SUTURE WAS USED. THE PATIENT WAS IN A VERY BAD CONDITION WHEN HE WAS TRANSPORTED, BUT SURGERY WAS PERFORMED IMMEDIATELY AT THE REQUEST OF HIS FAMILY. THE PRODUCT WAS USED ON THE FASCIA. THE PATIENT'S VITAL SIGN WAS BAD DURING THE SURGERY, AND THE OPERATION HAD TO BE COMPLETED QUICKLY. TWO BARBED SUTURES WERE USED TO CLOSE THE WOUND. THE MIDLINE INCISION ABOUT 15CM WAS STITCHED FROM THE UPPER AND LOWER ABDOMEN, THE MIDDLE PART WAS OVERLAPPED AND BACKSTITCH WAS PERFORMED. THE EPIDERMIS WAS CLOSED WITH A STAPLER. THE WOUND WAS DEHISCED AND THE ABDOMINAL CAVITY BECAME VISIBLE WHEN A STAPLER WAS REMOVED ON AUGUST 4TH. RE OPERATION WAS PERFORMED AND THE FASCIA WAS SUTURED WITH NYLON SUTURE. THE PATIENT'S CURRENT CONDITION IS STABLE. THE WOUND WAS DEHISCED FROM THE UPPER TO LOWER ABDOMEN, AND THE SUTURE WAS STILL IN THE TISSUE. THE SUTURE WAS NOT BROKEN, BUT THAT THE TISSUES COULDN'T WITHSTAND. THE DOCTOR COMMENTED THAT THE CAUSE WAS NOT DUE TO THE SUTURE, BUT RATHER TO THE PATIENT'S CONDITION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Wound Dehiscence

    MAUDE report 22862315 ↗
  • InjuryJul 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN OVARIAN CANCER RESECTION ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED TO SIMULTANEOUSLY SUTURE THE FASCIA AND PERITONEUM. IT WAS STATED THAT THE TAB PLACEMENT METHOD AND SUTURING COMPLETION METHOD WERE PERFORMED PROPERLY. POST-OP, THE PATIENT EXPERIENCED AN INTESTINAL OBSTRUCTION AND AN ILEUS. A RE-OPERATION WAS PERFORMED. THERE WERE NO PROBLEMS WITH THE PATIENT'S SUBSEQUENT PROCESS AFTER SURGERY. THE PHYSICIANS OPINED THAT THE ANCHOR OF THE DEVICE MAY HAVE BEEN CAUGHT IN THE INTESTINE, CAUSING AN ILEUS. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Obstruction/Occlusion

    MAUDE report 22642758 ↗
  • MalfunctionMay 13, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE ANCHOR WAS BROKEN WHEN TRIED TO USE IT DURING WOUND CLOSURE. (IT WAS FOUND THAT THERE HAD BEEN LIKE A KINK.) IT WAS NOT A CONTROL RELEASE NEEDLE. THIS EVENT WAS FOUND BEFORE USE. THE PRODUCT WAS USED ON THE FASCIA. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. PLEASE TAKE PHOTOS FOR JAPANESE CUSTOMER. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22013239 ↗
  • InjuryMay 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT MIDLINE ABDOMINAL INCISION SURGERY ON (B)(6) 2025 AND BARBED SUTURE WAS USED. THE PATIENT HAD AN ABDOMINAL WALL SCAR HERNIA. THE DOCTOR THINKS THAT THE SUTURE WAS NOT BROKEN, BUT EITHER THE LEFT OR RIGHT SIDE WAS TORN BECAUSE THE STITCHES PROLAPSED THE SAME LENGTH AS THE INCISION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hernia

    MAUDE report 21972251 ↗
  • MalfunctionMay 1, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ARTIFICIAL HEAD SURGERY ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN ARTIFICIAL HEAD SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE. THE PRODUCT WAS USED ON THE FASCIA. THERE WAS NO PROBLEM TO THE PATIENT. IT WAS NOT A CONTROL RELEASE NEEDLE. THIS EVENT WAS FOUND BEFORE USE. ANOTHER PACKAGE OF THE PRODUCT WAS USED TO COMPLETE THE CASE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21944003 ↗
  • MalfunctionApr 4, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING AN UNKNOWN SURGERY, THE SUTURE WAS BROKEN DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21773318 ↗
  • InjuryMar 14, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN OOPHORECTOMY ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED FOR WOUND CLOSURE WITHOUT ANY PROBLEMS. THE REDNESS DID NOT STOP FOR ABOUT 5 DAYS AFTER THE SURGERY, AND WHEN EXAMINED BY X-RAY, IT WAS IDENTIFIED THAT THE PATIENT HAD CONTRACTED SURGICAL SITE INFECTION, AND THEN IT HAD BECOME AN ABDOMINAL WALL INCISIONAL HERNIA. RE-OPERATION (REOPEN THE ABDOMEN) WAS PERFORMED. THE PATIENT HAS BEEN DISCHARGED FROM THE HOSPITAL. ACCORDING TO THE DOCTOR, THERE MIGHT BE A PROBLEM WITH THE BARBED SUTURE AS IT MELTED QUICKLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Erythema; Hernia; Unspecified Infection; Wound Dehiscence

    MAUDE report 21610400 ↗

FDA clearance and classification

The FDA regulates product code NEW (suture, surgical, absorbable, polydioxanone) under 21 CFR 878.4840 as a Class II device. 21 CFR 878.4840 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the STRATAFIX SXPP1A301

Has the STRATAFIX SXPP1A301 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705031227712), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the STRATAFIX SXPP1A301?

The FDA's MAUDE database lists 442 adverse event reports for this device, received between Nov 22, 2017 and Aug 31, 2026. In those reports, the most common specific device problems are “Break” (321), “Material Separation” (41) and “Material Integrity Problem” (40); the most common patient problems are “Wound Dehiscence” (44), “Not Applicable” (17) and “Hernia” (13). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the STRATAFIX SXPP1A301?

None of the reports are classified as deaths, 74 as injuries and 368 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the STRATAFIX SXPP1A301 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the STRATAFIX SXPP1A301, and is it still sold?

The FDA's device database lists Ethicon, LLC as the company responsible for it (brand STRATAFIX, model SXPP1A301). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices

All SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page