Skip to content

Stratafix Spiral SXMD1B403Recalls & reported problems

Surgical Specialties México, S. de R.L. de C.V. · RB-1 2-0 MONODERM 20CM · Class 2 · Implant

All models of Stratafix Spiral (24)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705031228054), received between Mar 17, 2022 and Aug 11, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 24, 2022Class IIOpen, Classified

    Affected product lots may contain the incorrect needle type/size and suture length due to component mix-up.

    Z-1053-2022 · Root cause: Employee error · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Wound Dehiscence1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 11, 2025Break; Break

    IT WAS REPORTED ON (B)(6) 2025 THAT POST-OPERATIVELY THE SURGICAL INCISION RUPTURED AND CAUSED BLEEDING. AN ADDITIONAL OPERATION WAS COMPLETED AND THE PATIENT IS REPORTEDLY STABLE. NO OTHER INJURIES REPORTED.

    Patient problems: Wound Dehiscence

    MAUDE report 22776426 ↗
  • MalfunctionMar 17, 2022Nonstandard Device

    WRONG NEEDLE TYPE. IT WAS REPORTED THAT THE NEEDLE TYPE WAS FOUND TO BE INCONSISTENT WITH THE ACTUAL SPECIFICATION. IT WAS A CUTTING NEEDLE RATHER THAN A TAPER POINT NEEDLE WHEN THE PACKAGE WAS OPENED FOR AN UNKNOWN SURGERY. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT OF PATIENT INJURY. NOT USED IN THE PROCEDURE. ANOTHER DEVICE WAS READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13801309 ↗

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices