PROLENE EPH7477HRecalls & reported problems
Ethicon, LLC · PROLENE (Polypropylene) Suture Blue Monofilament, Hemo-Seal Needle Suture, Sterile, nonabsorbable surgical suture · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Prolene blue polypropylene surgical suture with a Hemo-Seal needle is used to close tissue during surgery. The 30 reports mostly describe the thread breaking during the operation, coming apart from the needle, fraying or bending, and needles popping loose, sometimes forcing staff to use extra sutures to finish. There are no deaths and no injuries reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 10705031233454), received between Aug 22, 2019 and Apr 17, 2025 — classified as 30 malfunctions.
The most frequently reported specific device problem is “Break” (22 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break22
- Material Separation10
- Material Frayed6
- Material Twisted/Bent2
Patient problems
- No Clinical Signs, Symptoms or Conditions22
- No Consequences Or Impact To Patient9
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT IT IS A RECURRENT ISSUE: THE NEEDLE LOOSENED DURING USE. THIS IS A REPETITIVE INCIDENT IN THE SERVICE, AS THIS PROBLEM OCCURS ALMOST EVERY TIME THERE IS AN INTERVENTION. THIS DETACHMENT ISSUE SEEMS TO BE SPECIFIC TO THIS REFERENCE, WHICH REPLACES THE ONE USUALLY USED IN THE SERVICE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21859565 ↗MalfunctionApr 17, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. AT THE CCV BLOCK, THE NEEDLE LOOSENED DURING USE. A SECOND THREAD WAS USED TO COMPLETE THE GESTURE. THERE WERE NO CLINICAL CONSEQUENCES FOR THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21859529 ↗MalfunctionDec 17, 2024Break; Material Separation; Material Twisted/Bent; Break; Material Separation; Material Twisted/Bent
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP IS REPORTING THAT THE SUTURE IS BREAKING AT TIMES ALONG WITH THE NEEDLES POPPING OFF AND BENDING WHEN PASSING THROUGH. THEY HAD TO KEEP USING MORE UNTIL IT WORKED TO COMPLETE THE CASE. NEEDLES DID NOT FALL INTO THE PATIENT AND THERE WAS NO PATIENT HARM REPORTED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20951496 ↗MalfunctionDec 17, 2024Break; Material Separation; Material Twisted/Bent; Break; Material Separation; Material Twisted/Bent
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REP IS REPORTING THAT THE SUTURE IS BREAKING AT TIMES ALONG WITH THE NEEDLES POPPING OFF AND BENDING WHEN PASSING THROUGH. THEY HAD TO KEEP USING MORE UNTIL IT WORKED TO COMPLETE THE CASE. NEEDLES DID NOT FALL INTO THE PATIENT AND THERE WAS NO PATIENT HARM REPORTED. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20951483 ↗MalfunctionOct 31, 2024Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, SUTURE BREAKAGE OCCURRED A LOT OF TIMES AND THE SUTURE DETACHED FROM THE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20582867 ↗MalfunctionOct 31, 2024Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, SUTURE BREAKAGE OCCURRED A LOT OF TIMES AND THE SUTURE DETACHED FROM THE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20582824 ↗MalfunctionOct 31, 2024Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, SUTURE BREAKAGE OCCURRED A LOT OF TIMES AND THE SUTURE DETACHED FROM THE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20582801 ↗MalfunctionOct 31, 2024Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, SUTURE BREAKAGE OCCURRED A LOT OF TIMES AND THE SUTURE DETACHED FROM THE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20582770 ↗MalfunctionOct 31, 2024Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, SUTURE BREAKAGE OCCURRED A LOT OF TIMES AND THE SUTURE DETACHED FROM THE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20582766 ↗MalfunctionOct 31, 2024Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, SUTURE BREAKAGE OCCURRED A LOT OF TIMES AND THE SUTURE DETACHED FROM THE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20582737 ↗MalfunctionJul 12, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19730391 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715753 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715664 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715656 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715486 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715263 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715247 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715134 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715115 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715114 ↗MalfunctionJul 10, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT HAS BEEN NOTICED THAT THE THREADS BREAK WITH EACH USE JUST UNDER THE NEEDLE OR IN THE MIDDLE OF THE THREAD. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715113 ↗MalfunctionOct 28, 2020Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2020, AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10748470 ↗MalfunctionAug 22, 2019Material Frayed; Material Frayed
IT WAS REPORTED THAT THE PATIENT UNDERWENT CAROTID ARTERY REPAIR ON AN UNKNOWN DATE AND SUTURE WAS USED. THE SUTURE FRAYED WHEN PASSING THROUGH THE VASCULAR PROSTHESIS. THE SUTURE WAS CHANGED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8923681 ↗MalfunctionAug 22, 2019Material Frayed; Material Frayed
IT WAS REPORTED THAT THE PATIENT UNDERWENT CAROTID ARTERY REPAIR ON AN UNKNOWN DATE AND SUTURE WAS USED. THE SUTURE FRAYED WHEN PASSING THROUGH THE VASCULAR PROSTHESIS. THE SUTURE WAS CHANGED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8923680 ↗MalfunctionAug 22, 2019Material Frayed; Material Frayed
IT WAS REPORTED THAT THE PATIENT UNDERWENT CAROTID ARTERY REPAIR ON AN UNKNOWN DATE AND SUTURE WAS USED. THE SUTURE FRAYED WHEN PASSING THROUGH THE VASCULAR PROSTHESIS. THE SUTURE WAS CHANGED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8923679 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2455-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
- Z-2456-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2462-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
- Z-2452-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
- Z-2450-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36"(90CM) 4-0 BLUE, D5988
- Z-2465-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE MONO POLY - SH-1 Taperpoint 1/2C, D10027
- Z-2454-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X18"5/0 DA RB-2 +PLG, D7879
- Z-2459-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2457-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 6-0 BLUE, D8688
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page