STRATAFIX SXMP1B119Recalls & reported problems
Ethicon, LLC · Knotless Tissue Control Device, Undyed Unidirectional (Antibacterial) Synthetic Absorbable Device · Class 2 · Implant
Quick summary
This Ethicon STRATAFIX knotless barbed suture is an absorbable synthetic surgical thread. The 31 reports mostly describe the suture breaking or separating during surgery, the barbs not holding tissue properly, and in a few cases the suture not dissolving as expected, with some stitches working their way back out through the wound. Seven reports describe injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 48
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 48 adverse event reports that list this exact device (DI 10705031236745), received between Jan 2, 2019 and Sep 18, 2026 — classified as 9 injuries, 39 malfunctions.
The most frequently reported specific device problem is “Break” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break24
- Material Separation9
- Adverse Event Without Identified Device or Use Problem6
- Positioning Failure6
- Failure to Degrade2
- Device Markings/Labelling Problem1
- Insufficient Device Problem Information1
- Tear, Rip or Hole in Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions32
- No Consequences Or Impact To Patient7
- Pocket Erosion3
- Wound Dehiscence3
- Erythema2
- Not Applicable2
- Purulent Discharge2
- Foreign Body Reaction1
Recent adverse event reports
The 25 most recent of 48 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2026Break
It was reported that a patient underwent a Skin Closure procedure on an unknown date and barbed suture was used. During the procedure this barbed suture code has been breaking as they close the skin. They have been closing with this suture for a long time and never had issues. No delay or patient harm was reported. They noticed that it has been the last couple of cases that they have used the product on. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820906 ↗MalfunctionSep 18, 2026Break
It was reported that a patient underwent a Skin Closure procedure on an unknown date and barbed suture was used. During the procedure this barbed suture code has been breaking as they close the skin. They have been closing with this suture for a long time and never had issues. No delay or patient harm was reported. They noticed that it has been the last couple of cases that they have used the product on. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820758 ↗MalfunctionSep 18, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820751 ↗MalfunctionSep 18, 2026Break
It was reported that a patient underwent a Skin Closure procedure on an unknown date and barbed suture was used. During the procedure this barbed suture code has been breaking as they close the skin. They have been closing with this suture for a long time and never had issues. No delay or patient harm was reported. They noticed that it has been the last couple of cases that they have used the product on. No adverse patient consequences were reported. Additional information was requested
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820745 ↗MalfunctionSep 18, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820738 ↗MalfunctionSep 18, 2026Break
It was reported that a patient underwent a Skin Closure procedure on an unknown date and barbed suture was used. During the procedure this barbed suture code has been breaking as they close the skin. They have been closing with this suture for a long time and never had issues. No delay or patient harm was reported. They noticed that it has been the last couple of cases that they have used the product on. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820576 ↗MalfunctionSep 18, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820477 ↗MalfunctionSep 18, 2026Break
It was reported that a patient underwent a Skin Closure procedure on an unknown date and barbed suture was used. During the procedure this barbed suture code has been breaking as they close the skin. They have been closing with this suture for a long time and never had issues. No delay or patient harm was reported. They noticed that it has been the last couple of cases that they have used the product on. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820424 ↗MalfunctionSep 18, 2026Break
It was reported that a patient underwent a Skin Closure procedure on an unknown date and barbed suture was used. During the procedure this barbed suture code has been breaking as they close the skin. They have been closing with this suture for a long time and never had issues. No delay or patient harm was reported. They noticed that it has been the last couple of cases that they have used the product on. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820418 ↗MalfunctionSep 18, 2026Break
It was reported that a patient underwent a Skin Closure procedure on an unknown date and barbed suture was used. During the procedure this barbed suture code has been breaking as they close the skin. They have been closing with this suture for a long time and never had issues. No delay or patient harm was reported. They noticed that it has been the last couple of cases that they have used the product on. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820396 ↗MalfunctionSep 18, 2026Break
It was reported that a patient underwent a Skin Closure procedure on (b)(6) 2026 and barbed suture was used. During the procedure this barbed suture code has been breaking as they close the skin. They have been closing with this suture for a long time and never had issues. No delay or patient harm was reported. They noticed that it has been the last couple of cases that they have used the product on. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820261 ↗MalfunctionSep 18, 2026Break
It was reported that a patient underwent a Skin Closure procedure on (b)(6)2026 and barbed suture was used. During the procedure this barbed suture code has been breaking as they close the skin. They have been closing with this suture for a long time and never had issues. No delay or patient harm was reported. They noticed that it has been the last couple of cases that they have used the product on. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820243 ↗MalfunctionSep 18, 2026Break
It was reported that a patient underwent a Skin Closure procedure on (b)(6)2026 and barbed suture was used. During the procedure this barbed suture code has been breaking as they close the skin. They have been closing with this suture for a long time and never had issues. No delay or patient harm was reported. They noticed that it has been the last couple of cases that they have used the product on. No adverse patient consequences were reported. Additional information was requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820140 ↗MalfunctionSep 18, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820081 ↗MalfunctionSep 18, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820055 ↗InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a patient underwent a knee surgery on an unknown date and barbed suture was used. The patient encountered pus collection. Additional information was requested.
Patient problems: Purulent Discharge
MAUDE report 50009368 ↗InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a patient underwent a hand surgery on an unknown date and barbed suture was used. The patient encountered pus collection and was treated with I&D. Additional information was requested.
Patient problems: Purulent Discharge
MAUDE report 50009367 ↗MalfunctionMay 12, 2026Device Markings/Labelling Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. THE BARBED SUTURE PACKAGES DO NOT LOOK THE SAME. THEY'RE MISSING INFORMATION AND THE CUSTOMER IS WONDERING WHY SOME SYMBOLS ARE NOT ON THE OUTER AND INNER PACKAGING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25148527 ↗MalfunctionMay 7, 2026Positioning Failure
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE CLIENT REPORTS AN OBSERVATION RELATED TO STRATAFIX¿ 4-0 SUTURE ETHICON BRAND, CORRESPONDING TO THE FOLLOWING LOTS: (B)(4) UNITS LOT: 106BG7 AND (B)(4) UNIT LOT: 108MBR. AS INDICATED, IN TWO OF THE UNITS, A SCARCELY PERCEPTIBLE PRESENCE OF THE BALEEN STRUCTURE OF THE MATERIAL WAS VISUALLY EVIDENCED, AN OBSERVATION MADE AT THE TIME OF HANDLING THE PRODUCT. THE SUTURES INVOLVED WERE USED DURING THE SURGICAL PROCEDURE, WITH NO ADVERSE EVENT, CLINICAL COMPLICATION, OR IMPACT ON PATIENT SAFETY REPORTED. NO ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED ASSOCIATED WITH THIS OBSERVATION. THE REPORT IS MADE AT THE REQUEST OF THE USER, FOR THE PURPOSES OF EVALUATION AND CORRESPONDING MONITORING. ALSO REQUESTS THAT SAMPLES FROM THE SAME BATCH BE SENT TO BE OPENED IN THE PRESENCE OF THE DOCTOR. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25111350 ↗MalfunctionMay 7, 2026Positioning Failure
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE CLIENT REPORTS AN OBSERVATION RELATED TO STRATAFIX¿ 4-0 SUTURE ETHICON BRAND, CORRESPONDING TO THE FOLLOWING LOTS: 2 UNITS LOT 106BG7 AND 1 UNIT LOT 108MBR. AS INDICATED, IN TWO OF THE UNITS, A SCARCELY PERCEPTIBLE PRESENCE OF THE BALEEN STRUCTURE OF THE MATERIAL WAS VISUALLY EVIDENCED, AN OBSERVATION MADE AT THE TIME OF HANDLING THE PRODUCT. THE SUTURES INVOLVED WERE USED DURING THE SURGICAL PROCEDURE, WITH NO ADVERSE EVENT, CLINICAL COMPLICATION, OR IMPACT ON PATIENT SAFETY REPORTED. NO ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED ASSOCIATED WITH THIS OBSERVATION. THE REPORT IS MADE AT THE REQUEST OF THE USER, FOR THE PURPOSES OF EVALUATION AND CORRESPONDING MONITORING. ALSO REQUESTS THAT SAMPLES FROM THE SAME BATCH BE SENT TO BE OPENED IN THE PRESENCE OF THE DOCTOR. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25111347 ↗MalfunctionMay 7, 2026Positioning Failure
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE CLIENT REPORTS AN OBSERVATION RELATED TO STRATAFIX¿ 4-0 SUTURE ETHICON BRAND, CORRESPONDING TO THE FOLLOWING LOTS: (B)(4) UNITS LOT 106BG7 AND (B)(4) UNIT LOT 108MBR. AS INDICATED, IN TWO OF THE UNITS, A SCARCELY PERCEPTIBLE PRESENCE OF THE BALEEN STRUCTURE OF THE MATERIAL WAS VISUALLY EVIDENCED, AN OBSERVATION MADE AT THE TIME OF HANDLING THE PRODUCT. THE SUTURES INVOLVED WERE USED DURING THE SURGICAL PROCEDURE, WITH NO ADVERSE EVENT, CLINICAL COMPLICATION, OR IMPACT ON PATIENT SAFETY REPORTED. NO ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED ASSOCIATED WITH THIS OBSERVATION. THE REPORT IS MADE AT THE REQUEST OF THE USER, FOR THE PURPOSES OF EVALUATION AND CORRESPONDING MONITORING. ALSO REQUESTS THAT SAMPLES FROM THE SAME BATCH BE SENT TO BE OPENED IN THE PRESENCE OF THE DOCTOR. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25111341 ↗InjuryApr 2, 2026Failure to Degrade
IT WAS REPORTED THAT A PATIENT UNDERWENT A HIP PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. IN THE PAST WEEK THE PATIENT HAD SUTURE THAT SPIT OUT FROM THE HIP WOUND. SURGERY WAS 6-8 WEEKS AGO. WHEN THE HIP WAS REOPENED, THE SUTURE DID NOT DEGRADE OR ABSORB AT ALL. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Erythema; Impaired Healing; Pocket Erosion; Swelling/ Edema; Wound Dehiscence
MAUDE report 24769035 ↗InjuryApr 2, 2026Failure to Degrade
IT WAS REPORTED THAT A PATIENT UNDERWENT A HIP PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. IN THE PAST WEEK THE PATIENT HAD SUTURE THAT SPIT OUT FROM THE HIP WOUND. SURGERY WAS 6-8 WEEKS AGO. WHEN THE HIP WAS REOPENED, THE SUTURE DID NOT DEGRADE OR ABSORB AT ALL. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Pocket Erosion
MAUDE report 24768942 ↗MalfunctionAug 28, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A PLASTIC SURGERY ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING A PLASTIC SURGERY SUTURES WERE BREAKING. NO FURTHER INFORMATION PROVIDED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22914318 ↗MalfunctionAug 27, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A PLASTIC SURGERY ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT THE SUTURES WERE BREAKING. NO FURTHER INFORMATION PROVIDED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. DEVICE IS NOT AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22906015 ↗
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.