PROLENE W8761GRecalls & reported problems
Ethicon, LLC · Blue Monofilament (Polypropylene), Nonabsorbable Surgical Suture · Class 2 · Implant
Quick summary
This Ethicon Prolene W8761G is a blue polypropylene nonabsorbable surgical suture used to close tissue. The 33 reports mostly describe the thread snapping during surgery, sometimes even with no tension applied, plus material separating, a split needle, and objects found inside the packaging. Reports mention 4 injuries and no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 35
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 10705031242265), received between Mar 23, 2020 and Sep 3, 2026 — classified as 4 injuries, 31 malfunctions.
The most frequently reported specific device problem is “Break” (25 reports).
21 other recalls cover devices cleared under the same FDA submission (K133356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break25
- Material Separation7
- Device Contaminated During Manufacture or Shipping3
- Adverse Event Without Identified Device or Use Problem1
- Defective Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions25
- Insufficient Information7
- No Consequences Or Impact To Patient2
- Not Applicable2
- Impaired Healing1
Recent adverse event reports
The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50590409 ↗MalfunctionSep 3, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50590408 ↗MalfunctionJul 9, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25764566 ↗MalfunctionFeb 25, 2026Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE SUTURE BROKE DESPITE NO TENSION BEING APPLIED. NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24441690 ↗MalfunctionFeb 25, 2026Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE SUTURE BROKE DESPITE NO TENSION BEING APPLIED. NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24441689 ↗MalfunctionFeb 25, 2026Break
IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE SUTURE BROKE DESPITE NO TENSION BEING APPLIED. NO PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24441688 ↗MalfunctionJan 6, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC ASSISTED RADICAL RESECTION FOR COLON CANCER PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE SURGERY, THE DOCTOR USED BARBED SUTURE SXMP1B427, TO REINFORCE THE ANASTOMOTIC SITE, BUT THE SUTURE BROKE. TWO BARBED SUTURES SXPP1B410 WERE USED FOR SUBCUTANEOUS SUTURING, BUT THE BARB NON-ENGAGEMENT IN TISSUE AND THE SUTURES COULD BE EXTRACTED FROM THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23974437 ↗MalfunctionJan 6, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LOWER SEGMENT CESAREAN SECTION UNDER LUMBAR HARD JOINT PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE INTRADERMAL SUTURE, IT WAS FOUND THAT THE NEEDLE HAD SPLIT AND COULD NOT BE SUTURED NORMALLY. PRELIMINARY ANALYSIS SHOWS THAT THE PRODUCT QUALITY IS POOR AND THERE IS A PROBLEM OF LAX CONTROL. TIMELY REMOVAL, REPLACEMENT WITH A NEW NEEDLE, AND CONTINUED SUTURING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23974135 ↗MalfunctionDec 18, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING CLINICAL USE, UNKNOWN OBJECT WAS FOUND INSIDE THE PACKAGING, AND IT WAS DISCONTINUED AND REPLACED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23843501 ↗MalfunctionNov 4, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 4 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL RETURN 121 REPRESENTATIVE SAMPLES FOR ANALYSIS. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23467073 ↗MalfunctionNov 4, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 4 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL RETURN 121 REPRESENTATIVE SAMPLES FOR ANALYSIS. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23467068 ↗MalfunctionNov 4, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 4 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL RETURN 121 REPRESENTATIVE SAMPLES FOR ANALYSIS. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23467067 ↗MalfunctionNov 4, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 4 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. THERE WILL RETURN 121 REPRESENTATIVE SAMPLES FOR ANALYSIS. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23467066 ↗MalfunctionNov 4, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY . TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23464493 ↗MalfunctionNov 4, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY . TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23464490 ↗MalfunctionNov 4, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23464489 ↗MalfunctionNov 4, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6)2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN SURGERY. TOTAL 4 PRODUCTS OCCURRED NEEDLE PULL OFF ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23464487 ↗InjuryJul 11, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE HOSPITAL SAID THE DEVICE WAS BREAKING FROM THE NEEDLE-THREAD CONNECTION POINT CONSTANTLY. ONE PATIENT HAS PASSED AWAY. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 22484323 ↗MalfunctionJul 11, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE HOSPITAL SAID THE DEVICE WAS BREAKING FROM THE NEEDLE-THREAD CONNECTION POINT CONSTANTLY. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22484322 ↗MalfunctionMay 22, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, SUTURE SLIPPED OUT OFF NEEDLE. THERE WERE NO ADVERSE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22073711 ↗MalfunctionApr 29, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN AORTIC VALVE REPAIR PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, WHILE STITCHING THE SUTURELINE OF THE AORTIC GRAFT WITH A PROLENE 4-0 20MM, ONE OF THE NEEDLES BROKE OFF, WITH PART OF THE NEEDLE STILL WITH THE SUTURE. INFORMED THE WHOLE TEAM. CONSULTANT SURGEON SCRUBBED IN AND TRIED TO LOOK FOR THE NEEDLE. FOLLOWED PROTOCOL AND INFORMED THE THEATRE COORDINATOR. X-RAY WAS REQUESTED AND DONE. NEEDLE WAS FOUND ON THE X-RAY IMAGES. MORNING CARDIAC COORDINATOR INFORMED AND AWARE AT 8AM. NEEDLE WAS REMOVED BY THE CONSULTANT SURGEON. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21925837 ↗MalfunctionJan 17, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENDON REPAIR PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THERE WAS LACK OF ADEQUATE STRENGTH WITH THREAD TENSION. THERE WAS BREAKING OF SURGICAL SUTURE DURING THE PROCEDURE AND KNOT TYING. IT WAS REPORTED THAT THERE WAS A SURGICAL DELAY. A DIFFERENT SUTURE WAS USED TO COMPLETE THE PROCEDURE. IT WAS REPORTED THAT THIS EVENT CAUSED A SERIOUS INJURY AND ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED, HOWEVER, SPECIFICS ARE UNKNOWN AT THIS TIME. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 21180048 ↗MalfunctionJan 17, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENDON REPAIR PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THERE WAS LACK OF ADEQUATE STRENGTH WITH THREAD TENSION. THERE WAS BREAKING OF SURGICAL SUTURE DURING THE PROCEDURE AND KNOT TYING. IT WAS REPORTED THAT THERE WAS A SURGICAL DELAY. A DIFFERENT SUTURE WAS USED TO COMPLETE THE PROCEDURE. IT WAS REPORTED THAT THIS EVENT CAUSED A SERIOUS INJURY AND ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED, HOWEVER, SPECIFICS ARE UNKNOWN AT THIS TIME. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 21180042 ↗MalfunctionJan 17, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENDON REPAIR PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THERE WAS LACK OF ADEQUATE STRENGTH WITH THREAD TENSION. THERE WAS BREAKING OF SURGICAL SUTURE DURING THE PROCEDURE AND KNOT TYING. IT WAS REPORTED THAT THERE WAS A SURGICAL DELAY. A DIFFERENT SUTURE WAS USED TO COMPLETE THE PROCEDURE. IT WAS REPORTED THAT THIS EVENT CAUSED A SERIOUS INJURY AND ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED, HOWEVER, SPECIFICS ARE UNKNOWN AT THIS TIME. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 21180041 ↗MalfunctionJan 17, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENDON REPAIR PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THERE WAS LACK OF ADEQUATE STRENGTH WITH THREAD TENSION. THERE WAS BREAKING OF SURGICAL SUTURE DURING THE PROCEDURE AND KNOT TYING. IT WAS REPORTED THAT THERE WAS A SURGICAL DELAY. A DIFFERENT SUTURE WAS USED TO COMPLETE THE PROCEDURE. IT WAS REPORTED THAT THIS EVENT CAUSED A SERIOUS INJURY AND ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED, HOWEVER, SPECIFICS ARE UNKNOWN AT THIS TIME. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 21180040 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133356. One submission often covers many models, so check the product and lot details in each record.
- Z-1515-2025 · Mar 7, 2025 · Open, ClassifiedPROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
- Z-1096-2025 · Dec 20, 2024 · Open, ClassifiedPROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in ca
- Z-2464-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 18" 7-0 D/A BV-1, DC631
- Z-2462-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
- Z-2452-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748
- Z-2463-2018 · May 7, 2018 · TerminatedPROLENE SUTURE MONO BLUE BV1755 Taperpoint 3/8C, D9995
- Z-2456-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 6-0 BLUE, D8647
- Z-2458-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 24"(60CM) 7-0 BLUE, D8881
- Z-2455-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 30"(75CM) 8-0 BLUE, D8292
- Z-2459-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D9067
- Z-2453-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 36"(90CM) 4-0 BLUE, D7768
- Z-2451-2018 · May 7, 2018 · TerminatedPROLENE SUTURE 4-36" (90CM) 3-0 BLUE, D6416
- Z-2461-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 48" 4-0 D/A SH-1, D9746
- Z-2466-2018 · May 7, 2018 · TerminatedPRONOVA SUTURE BLUE 122CM M1.5, D10189
- Z-2449-2018 · May 7, 2018 · TerminatedPROLENE SUTURE BLUE 4X90CM M1.5 USP 4/0 DA RB-1 +P, D5981
…and 6 more.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.