Skip to content

STRATAFIX SXMP2B416Recalls & reported problems

Ethicon, LLC · Stratafix (Spiral MONOCRYL Plus), Bidirectional · Class 2 · Implant

All models of STRATAFIX (181)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705031465114), received between Jul 25, 2025 and Jul 25, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Positioning Failure” (1 report).

1 other recall cover devices cleared under the same FDA submission (K192580); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 25, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOT ASSISTED PROSTATECTOMY PROCEDURE IN (B)(6) 2025 AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT DURING A ROBOT ASSISTED PROSTATECTOMY, THE SURGEON FELT THAT THE SUTURE STRAND DID NOT ADEQUATELY HOLD THE TISSUE APPROXIMATED FOR THE UVA. THEY WERE UNABLE TO SEE THE BARBS UNDER THE MAGNIFICATION OF THE SCOPE. THE STRAND IN QUESTION IS NOT AVAILABLE FOR RETURN, BUT A STRAND FROM THE SAME BOX IS AVAILABLE FOR ANALYSIS. SURGICAL DELAY WAS UNKNOWN. NO PATIENT WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22621108 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K192580. One submission often covers many models, so check the product and lot details in each record.

  • Z-1533-2024 · Mar 8, 2024 · Open, Classified
    STRATAFIX Spiral MONOCRYL Plus Bidirectional Knotless Tissue Control Device: STRATAFIX SPIRAL MONOCRYL PLUS BI UD 12+12IN(30+30CM) USP3-0(M2) D/A PS PRM- Indicated for use in soft

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page