STRATAFIX SXMP1B449Recalls & reported problems

Ethicon, LLC · Knotless Tissue Control Device, Dyed Unidirectional (Antibacterial) Synthetic Absorbable Device · Class 2 · Implant

All models of STRATAFIX (181)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10705031470002), received between Oct 5, 2023 and Feb 24, 2026 — classified as 4 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Positioning Failure2
  • Adverse Event Without Identified Device or Use Problem1
  • Material Frayed1

Patient problems

  • Failure to Anastomose4
  • No Clinical Signs, Symptoms or Conditions3
  • Impaired Healing1
  • Insufficient Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Material Frayed

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PARTIAL NEPHRECTOMY PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, THE SALES REPRESENTATIVE REPORTED THAT DURING A PARTIAL NEPHRECTOMY PROCEDURE, THE SURGEON OBSERVED A SECTION OF THE SUTURE MATERIAL THAT APPEARED FRAYED, AS IF PART OF THE SUTURE WAS ¿SPITTING,¿ WITH NO PATIENT CONSEQUENCES NOTED; THE DEVICE IS NOT AVAILABLE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24434842 ↗
  • MalfunctionOct 6, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A HERNIA REPAIR PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23227789 ↗
  • MalfunctionMay 16, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A HEPATOBILIARY SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT THE NEEDLE BROKE WHEN SEWING IN SURGERY AT THE SECOND STITCH. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THE BROKEN NEEDLE WAS RETRIEVED SHORTLY. THE PROCEDURE WAS COMPLETED AND NO PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22036428 ↗
  • MalfunctionMay 14, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE SUTURING OPERATION, THE TAIL END OF THE NEEDLE BROKE AND COULD NOT BE USED. IT WAS IMMEDIATELY REMOVED AND CHECKED FOR INTEGRITY, AND A NEW NEEDLE WAS REPLACED TO CONTINUE SUTURING. THIS EVENT HAS AFFECTED THE PATIENT'S SURGICAL PROGRESS. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 22017812 ↗
  • InjuryApr 7, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PROSTATECTOMY ON (B)(6) 2025 AND A BARBED SUTURE WAS USED. POST-OP, THE SUTURE CAME APART CAUSING A LEAK. THE PATIENT WAS READMITTED. THE PATIENT WAS OVER SEWED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Failure to Anastomose

    MAUDE report 21784252 ↗
  • InjuryApr 1, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC PROSTATECTOMY ON AN UNKNOWN DATE IN 2025 AND BARBED SUTURE WAS TRAILED. THE PATIENT EXPERIENCED A BLADDER LEAK FROM THE URETEROVESICAL ANASTOMOSIS. A LEAK TEST WAS CONDUCTED INTRA-OPERATIVELY, AND THE ANASTOMOSIS LOOKED GOOD AT THE TIME. DURING PATIENT FOLLOW-UP, A LEAK WAS DISCOVERED AND DETERMINED TO BE FROM THE BLADDER. SURGEON OPINES THAT THE STRATAFIX SLIPPED AND CAUSED THE ANASTOMOSIS TO FAIL. SURGEON STATES HE HAS NEVER HAD A BLADDER LEAK IN OVER 1000 CASES, AND THE ONLY CHANGED VARIABLE IN THIS SPECIFIC PATIENT WAS THE BARBED SUTURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Failure to Anastomose

    MAUDE report 21740924 ↗
  • InjuryApr 1, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC PROSTATECTOMY ON AN UNKNOWN DATE IN 2025 AND BARBED SUTURE WAS TRIALED. THE PATIENT EXPERIENCED A BLADDER LEAK FROM THE URETEROVESICAL ANASTOMOSIS. A LEAK TEST WAS CONDUCTED INTRA-OPERATIVELY, AND THE ANASTOMOSIS LOOKED GOOD AT THE TIME. DURING PATIENT FOLLOW-UP, A LEAK WAS DISCOVERED AND DETERMINED TO BE FROM THE BLADDER. SURGEON OPINES THAT THE STRATAFIX SLIPPED AND CAUSED THE ANASTOMOSIS TO FAIL. SURGEON STATES HE HAS NEVER HAD A BLADDER LEAK IN OVER 1000 CASES, AND THE ONLY CHANGED VARIABLE IN THIS SPECIFIC PATIENT WAS THE BARBED SUTURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Failure to Anastomose

    MAUDE report 21740896 ↗
  • InjuryOct 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A RADICAL ROBOTIC PROSTATECTOMY ON AN UNKNOWN DATE AND A BARBED SUTURE WAS USED ON THE ANASTOMOSIS. IT WAS STATED THAT ALL LOOKED GOOD IN THE PROCEDURE. POST-OP, THE PATIENT¿S ANASTOMOSIS DID NOT HEAL WELL AND DID NOT CLOSE. THE PATIENT IS BEING SENT TO ANOTHER FACILITY TO GET A RECONSTRUCTION. THE CURRENT STATUS OF THE PATIENT WAS REPORTED AS WAITING FOR A RECONSTRUCTION SURGERY. THE PATIENT ALSO HAS OTHER ISSUES UNRELATED TO THE ANASTOMOSIS. THE SURGEON WAS NOT SURE IF IT WAS FROM THE BARBED SUTURE OR IT WAS BECAUSE THE PATIENT HAS OTHER ISSUES. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Failure to Anastomose; Impaired Healing

    MAUDE report 17877630 ↗

Other Suture, absorbable, synthetic, polyglycolic acid devices

All Suture, absorbable, synthetic, polyglycolic acid devices →