STRATAFIX SXMP1B441Recalls & reported problems

Ethicon, LLC · Knotless Tissue Control Device, Dyed Unidirectional (Antibacterial) Synthetic Absorbable Device · Class 2 · Implant

All models of STRATAFIX (181)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705031470057), received between Feb 26, 2026 and Feb 26, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Separation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 4 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation4
  • Positioning Failure4

Patient problems

  • Foreign Body In Patient2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2026Positioning Failure; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT DURING ROBOTIC HERNIA CASES FOR THE SURGEONS, WHEN TRYING TO INSERT SUTURE SPIRAL DOWN 8MM INTUITIVE TROCAR PORT THE NEEDLE DETACHED AT THE SWAGE SITE AND FELL INTO THE PATIENT. THE NEEDLE WAS PROMPTLY LOCATED AND REMOVED FROM THE PATIENT IN MOST SITUATIONS. IN ONE INSTANCE, THE NEEDLE WAS LOST IN THE PATIENT AND REQUIRED X-RAYS TO LOCATE THE LOST NEEDLE. THERE WAS ALSO COMPLAINTS MENTIONED IN OUR DISCUSSION REGARDING THE SUTURE "SLIPPING" WHEN TRYING TO CLOSE THE PERITONEUM. THE SLIPPING WAS TYPICALLY RESOLVED THROUGH ADDITIONAL PULLING TO ENGAGE THE BARBS FURTHER, HOWEVER THE SURGEONS DID HAVE CONCERN REGARDING LONG TERM HOLD. THE DEVICES WERE NOT RETURNING. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24461270 ↗
  • MalfunctionFeb 26, 2026Positioning Failure; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT DURING ROBOTIC HERNIA CASES FOR THE SURGEONS, WHEN TRYING TO INSERT SUTURE SPIRAL DOWN 8MM INTUITIVE TROCAR PORT THE NEEDLE DETACHED AT THE SWAGE SITE AND FELL INTO THE PATIENT. THE NEEDLE WAS PROMPTLY LOCATED AND REMOVED FROM THE PATIENT IN MOST SITUATIONS. IN ONE INSTANCE, THE NEEDLE WAS LOST IN THE PATIENT AND REQUIRED X-RAYS TO LOCATE THE LOST NEEDLE. THERE WAS ALSO COMPLAINTS MENTIONED IN OUR DISCUSSION REGARDING THE SUTURE "SLIPPING" WHEN TRYING TO CLOSE THE PERITONEUM. THE SLIPPING WAS TYPICALLY RESOLVED THROUGH ADDITIONAL PULLING TO ENGAGE THE BARBS FURTHER; HOWEVER, THE SURGEONS DID HAVE CONCERN REGARDING LONG TERM HOLD. THE DEVICES WERE NOT RETURNING. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24461267 ↗
  • MalfunctionFeb 26, 2026Positioning Failure; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC HERNIA PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT DURING ROBOTIC HERNIA CASES FOR THE SURGEONS, WHEN TRYING TO INSERT SUTURE SPIRAL DOWN 8MM INTUITIVE TROCAR PORT THE NEEDLE DETACHED AT THE SWAGE SITE AND FELL INTO THE PATIENT. THE NEEDLE WAS PROMPTLY LOCATED AND REMOVED FROM THE PATIENT IN MOST SITUATIONS. IN ONE INSTANCE, THE NEEDLE WAS LOST IN THE PATIENT AND REQUIRED X-RAYS TO LOCATE THE LOST NEEDLE. THERE WAS ALSO COMPLAINTS MENTIONED IN OUR DISCUSSION REGARDING THE SUTURE "SLIPPING" WHEN TRYING TO CLOSE THE PERITONEUM. THE SLIPPING WAS TYPICALLY RESOLVED THROUGH ADDITIONAL PULLING TO ENGAGE THE BARBS FURTHER; HOWEVER, THE SURGEONS DID HAVE CONCERN REGARDING LONG TERM HOLD. THE DEVICES WERE NOT RETURNING. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 24457417 ↗
  • MalfunctionFeb 26, 2026Positioning Failure; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ROBOTIC HERNIA PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. IT WAS REPORTED THAT DURING ROBOTIC HERNIA CASES FOR THE SURGEONS, WHEN TRYING TO INSERT SUTURE SPIRAL DOWN 8 MM INTUITIVE TROCAR PORT THE NEEDLE DETACHED AT THE SWAGE SITE AND FELL INTO THE PATIENT. THE NEEDLE WAS PROMPTLY LOCATED AND REMOVED FROM THE PATIENT IN MOST SITUATIONS. IN ONE INSTANCE, THE NEEDLE WAS LOST IN THE PATIENT AND REQUIRED X-RAYS TO LOCATE THE LOST NEEDLE. THERE WAS ALSO COMPLAINTS MENTIONED IN OUR DISCUSSION REGARDING THE SUTURE "SLIPPING" WHEN TRYING TO CLOSE THE PERITONEUM. THE SLIPPING WAS TYPICALLY RESOLVED THROUGH ADDITIONAL PULLING TO ENGAGE THE BARBS FURTHER. HOWEVER, THE SURGEONS DID HAVE CONCERN REGARDING LONG TERM HOLD. THE DEVICES WERE NOT RETURNING. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 24457415 ↗